Jeff Snyder1, Kelly C McPhee2, Alan H Wilman1. 1. Department of Biomedical Engineering, University of Alberta, Edmonton, Canada. 2. CancerCare Manitoba, Winnipeg, Canada.
Abstract
PURPOSE: Three-dimensional fast spin-echo (FSE) sequences commonly use very long echo trains (>64 echoes) and severely reduced refocusing angles. They are increasingly used in brain exams due to high, isotropic resolution and reasonable scan time when using long trains and short interecho spacing. In this study, T2 quantification in 3D FSE is investigated to achieve increased resolution when comparing with established 2D (proton-density dual-echo and multi-echo spin-echo) methods. METHODS: The FSE sequence design was explored to use long echo trains while minimizing T2 fitting error and maintaining typical proton density and T2 -weighted contrasts. Constant and variable flip angle trains were investigated using extended phase graph and Bloch equation simulations. Optimized parameters were analyzed in phantom experiments and validated in vivo in comparison to 2D methods for eight regions of interest in brain, including deep gray-matter structures and white-matter tracts. RESULTS: Phantom and healthy in vivo brain T2 measurements showed that optimized variable echo-train 3D FSE performs similarly to previous 2D methods, while achieving three-fold-higher slice resolution, evident visually in the 3D T2 maps. Optimization resulted in better T2 fitting and compared well with standard multi-echo spin echo (within the 8-ms confidence limits defined based on Bland-Altman analysis). CONCLUSION: T2 mapping using 3D FSE with long echo trains and variable refocusing angles provides T2 accuracy in agreement with 2D methods with additional high-resolution benefits, allowing isotropic views while avoiding incidental magnetization transfer effects. Consequently, optimized 3D sequences should be considered when choosing T2 mapping methods for high anatomic detail.
PURPOSE: Three-dimensional fast spin-echo (FSE) sequences commonly use very long echo trains (>64 echoes) and severely reduced refocusing angles. They are increasingly used in brain exams due to high, isotropic resolution and reasonable scan time when using long trains and short interecho spacing. In this study, T2 quantification in 3D FSE is investigated to achieve increased resolution when comparing with established 2D (proton-density dual-echo and multi-echo spin-echo) methods. METHODS: The FSE sequence design was explored to use long echo trains while minimizing T2 fitting error and maintaining typical proton density and T2 -weighted contrasts. Constant and variable flip angle trains were investigated using extended phase graph and Bloch equation simulations. Optimized parameters were analyzed in phantom experiments and validated in vivo in comparison to 2D methods for eight regions of interest in brain, including deep gray-matter structures and white-matter tracts. RESULTS: Phantom and healthy in vivo brain T2 measurements showed that optimized variable echo-train 3D FSE performs similarly to previous 2D methods, while achieving three-fold-higher slice resolution, evident visually in the 3D T2 maps. Optimization resulted in better T2 fitting and compared well with standard multi-echo spin echo (within the 8-ms confidence limits defined based on Bland-Altman analysis). CONCLUSION: T2 mapping using 3D FSE with long echo trains and variable refocusing angles provides T2 accuracy in agreement with 2D methods with additional high-resolution benefits, allowing isotropic views while avoiding incidental magnetization transfer effects. Consequently, optimized 3D sequences should be considered when choosing T2 mapping methods for high anatomic detail.
Three‐dimensional spin echo (SE)–based methods are gaining a significant foothold in weighted imaging and can provide increased resolution often with only negligible effects to total imaging time.
,
,
In terms of 3D turbo spin echo or fast spin echo (FSE), optimized long echo trains
,
,
,
,
,
,
,
in combination with short interecho spacings and low refocusing angles has resulted in specialized sequences delivering T
1, T
2, or proton‐density (PD) contrast while minimizing acquisition time (TA) and specific absorption rate. These methods have different names depending on manufacturer, such as SPACE (Sampling Perfection with Application optimized Contrasts using different flip angle Evolution [Siemens]), CUBE (General Electric), and VISTA (Volume Isotropic Turbo spin echo Acquisition [Philips]).Although increased use of 3D‐weighted imaging is emerging, it still lacks the robust, clinical quantification options widely developed for the 2D‐FSE analogue. The two prominent 2D methods include acquiring several echoes along the signal decay curve (multi‐echo spin‐echo [MESE]) or dual‐echo (PD‐T2) approaches. Although MESE is considered the gold standard for T
2 mapping, it is time‐consuming and generally not clinically available. As shown in McPhee and Wilman,
dual‐echo FSE T
2 mapping from PD and T
2‐weighted contrasts can be accurate if the spin response and flip angles are properly modeled. With adequate SNR, PD‐T2 methods have shown to produce similar results to MESE for a specified range of T
2 values, while substantially shortening imaging time due to acquisition of only two effective TEs. Modeling the PD‐T2 approximation to the decay curve requires inclusion of the measured flip angles (i.e., the deviation of flip angle from nominal) through map generation
and detailed knowledge of the pulse sequence. Improving resolution is possible due to decreased TA over MESE, although 2D sequences generally have lower resolution in the slice direction than in‐plane, which may result in partial voluming and misclassification of T
2s. The 2D PD‐T2 methods can also be confounded with incidental magnetization‐transfer effects.
These drawbacks are considerable, although 2D T
2 mapping is well established and used in clinical studies.Building on the 2D approach to T
2 fitting, 3D methods have become viable due to sequence modifications, allowing shortening of the TA by using long FSE trains with low flip angles and short interecho spacing, obtaining multiple subresolution images at several TEs, or the addition of gradient‐echo (GE) modules to speed data acquisition. Subresolution images, such as those acquired in the stack of stars technique,
,
use long echo trains, optimized flip angles, and subspace reconstruction methods to extract T
2 information for each echo, thereby decreasing the total TA from a dual‐echo or multi‐echo scheme, with the drawback of producing a single weighted image for a specific TE. Combination GE/SE techniques speed up the acquisition by additional GE modules following SE trains
or interleaving of SE and GE acquisitions, such as typically used in myelin‐water mapping methods.
T
2‐preparation pulses can also be used in standard 3D‐SE sequences to produce different magnetization pathways,
resulting in T
2 mapping ability but limited to single image output. Furthermore, the development of specialized MR fingerprinting sequences
,
,
,
has pushed quantification boundaries while seeking to maintain resolution with clinical images within relatively short TAs.Given the multiple techniques available, T
2 quantification from 3D FSE appears promising, although specialized research sequences unavailable in the clinic are often used. Constraining techniques to clinical sequences, the comparison of 3D FSE (often used for high‐resolution anatomical images such as fluid‐attenuated inversion recovery and T
2‐weighted and T
1‐weighted contrast) with established 2D methods is required to prove clinical utility and accuracy. Three‐dimensional FSE differs from 2D primarily in the use of extremely low and variable refocusing flip angles and long echo trains, which has not been fully investigated for T
2 fitting as compared with gold‐standard 2D methods.In this work, we explore T
2 quantification from long echo train 3D‐FSE sequences using PD‐weighted and T
2‐weighted acquisitions with optimized echo train parameters, while keeping resulting standard‐weighted PD and T
2 images. This allows a direct comparison between the established 2D PD‐T2 method, as both rely on properly fitting the signal acquired from imaging data at different time points to a set of predicted values (a simulated MR sequence dictionary).
An easy‐to‐visualize example—similar to 2D—of a constant flip angle train is used as the initial starting point, and more complicated examples using variable flip angle trains are later examined in simulation, phantom, and healthy volunteer brain experiments.
METHODS
To quantify the performance of 3D‐FSE sequences for T
2 mapping, simulations are necessary to provide dictionary values for both decay curve matching and improved fitting via parameter optimization. Although the optimization and evaluation approach is presented for the generic case, the solution space is systematically constrained to two types of sequences, specifically the constant flip angle train mimicking 2D acquisitions and a common T
2‐weighted variable flip angle train. Once the sequence has been analyzed, phantom and volunteer experiments were conducted to verify the numerical outcomes.
Choosing a 3D‐FSE sequence
The first step is to choose a desired 3D‐FSE sequence to determine T
2 mapping performance. Common sequences include constant flip angle trains or scanner‐defined variable flip angle trains targeting specific image weightings (PD‐weighted, T
2‐weighted, or T
1‐weighted). Knowledge of the echo train length (ETL), echo spacing, TR, and flip angle array is required for sequence evaluation.
Pulse sequence simulations
The a priori information required for quantification, the decay curve dictionary, can be as simple as assuming a mono‐exponential decay and fitting the acquired image points, although studies have shown that this approach results in heavily biased T
2 values. Accurate T
2 quantification requires true sequence simulation using extended phase graph (EPG
for 3D) or Bloch equation techniques (2D and 3D) to account for indirect and stimulated echoes when flip angles deviate from 180°.
,
Furthermore, in 2D experiments, slice selection and crusher gradient application requires simulation of spins in space.
,
,
Additionally, while simulations can predict the response at each echo for a fully relaxed system, steady‐state solutions should be explored when using TRs shorter than complete relaxation described by typical in vivo longitudinal relaxation rates.Fortunately, 3D FSE often uses nonselective refocusing; therefore, slice modeling is not required, resulting in significant overhead time savings. Instead, single voxels can be modeled as receiving a constant rather than a slice profile distribution. The simplest and most intuitive FSE sequence uses a reduced (from 180°) constant refocusing flip angle throughout the echo train, with a series of intermediate pulses following the 90° excitation pulse to move the response into the pseudo‐steady state (PSS).
,
More generic sequences manipulate the flip angle evolution in terms of plateaus and ramps to provide increased signal sustainability at a certain level, reduction of point‐spread function artifact, or selection of specific mixed contrasts through traversal of the k‐space center at a specific TE with optimized signal.
As discussed in Weigel and Hennig,
the flip angle evolution for FSE can be described by several sections, consisting of a transition from excitation to the PSS, a flip angle ramp, and a plateau followed by a transition to a second PSS. During the ramp, flip angles are increased for succeeding echoes to generate a relatively high signal plateau for k‐space center acquisition (for examples, see Mugler
).For each echo train investigated, simulations of decay curves were produced using in‐house MATLAB code, modeling steady‐state nonselective EPG and Bloch equation solutions for various combinations of T
1, T
2, and , with other sequence parameters fixed (ETL, echo spacing, TR, and flip angle array). The decay curve dictionary is therefore a 4D matrix with dimensions of T
1, , T
2, and echo number (or TE). After the dictionary is constructed, fitting involves matching image points (PD, T
2, and ) to simulated decay curves, with the assumption that T
1 does not vary significantly (see Section 3).
Echo‐time selection
The fitting approach requires selection of TEs for two images, to model the simulated decay curve and provide T
2 quantification. Curve fitting will be successful if there is (1) sufficient SNR for both echoes to alleviate T
2 errors from noise,
and (2) adequate variance in relative signal intensity between PD and T
2 across the T
2 range of interest. Although actual image weighting may differ based on TE selection, S
1 and S
2 will refer to the signal of the PD and T
2 images, respectively. These two requirements can be explored using calculated signal parameters. First, the difference between the two echo signals, S = S
1 − S
2, can be used as a measure of available contrast. A large range of S values across the T
2 range of interest is desirable to allow maximum distinction between T
2 isochromats. In other words, S should be as large as possible for each investigated T
2 value, and S should also be different for each T
2 value in the range of interest to prevent multiple fitting solutions. As S will depend on sequence parameters, parameter maps provide insight into the variation of S and performance across the T
2 range (e.g., S as a function of T
1 and T
2). To aid TE selection for the two images for an individual sequence, a normalized difference parameter, S
nd, can be defined as
ensuring that each difference is weighted by the signal of the first echo. For a typical decay, S
2 is less than S
1 (e.g., during a near‐exponential decay) and therefore 0 ≤ S
nd ≤ 1. In the case when a later echo (S
2) has greater amplitude than S
1, a negative value for S
nd occurs (signal ramp). During a perfect plateau in the signal, S
1 = S
2 and S
nd = 0. S
nd is a means for measuring the difference produced by the sequence for individual T
2s to be mapped, and due to the normalization, this parameter can be compared between different sequences. To include both parameters (S and S
nd) in measuring the T
2 mapping performance of a sequence, a third measure can be constructed as follows:The S parameter essentially weights the differences in T
2 mapping values (S
nd) by the available contrast (S) based on the difference of the two points.As S depends on specific selections of TEs for evaluation, a method is needed to determine optimal TEs for the PD and T
2 images. An iterative scheme can be used by looping through all echoes in the train and computing S
nd for each pair of echoes. By fixing T
1 and and constraining T
2 to in vivo values, a matrix of modified S
nd values (oT2) can be computed with dimensions ETL × ETL, and each element described by
where S is the signal for the ith or jth echo, and T
2 max and T
2 min represent the maximum and minimum T
2s to be investigated. Consequently, each oT2 value describes the signal variation between extreme values in the T
2 range (i.e., the spread between T
2 min and T
2 max), with the largest oT2 value corresponding to the echo combination producing the maximum signal variation between T
2s.As an example to visualize the signal parameters, consider the plot in Figure 1E illustrating S
nd for a range of T
1 and T
2 values for the constant 32‐echo case, with echo selections of 1 and 16. In healthy brain samples in vivo at 3 T, expected values of T
2 and T
1 fall in the range of 40–90 ms and 800–1400 ms,
respectively. A single point in the oT2 matrix (i.e., the available mapping space for T
2 values at echoes 1 and 16) can be estimated using S
nd and taking the difference of S
nd (, T
1, 0.1 s) and S
nd (, T
1, 0.03 s), where and T
1 are the constant values used for subtraction in Equation (3) and the T
2 range has been broadened to allow for uncertainty. Note that in sequences with strong T
1 dependence, this approximation will not hold, as different T
1s will produce different results. To complete oT2 generation, the difference calculation is repeated for each combination of echo pairs ranging from 1 to ETL. Optimal TEs are then chosen based on the maximum value of oT2. After TE selection is determined for a given sequence, it is important to consider the signal parameters: the total signal difference between echoes reflecting SNR (S) and the normalized difference showing T
2 decay spread and T
1 dependence (S
nd), which are reflected in S. Additionally, images should be visually inspected to ensure that the desired contrast is produced for the PD‐weighted and T
2‐weighted images to maintain clinical usability. Once S has been computed, different sequences can be compared at optimal TEs.
FIGURE 1
Echo signal, signal difference (S), and normalized difference (S
nd) as a function of T
1 and T
2 with = 1.0 for a 32‐echo constant flip angle train. (A) Comparison of flip angles and responses (S120 and S180) between the constant 120° (circles) and 180° (squares) trains for T
2 = 100 ms and T
1 = 1 s (flip angles are normalized to 180°). Signal amplitudes are shown for echo 1 (TEeff = 6 ms) (B) and echo 16 (TEeff = 96 ms) (C). The signal has been normalized to the maximum achievable value of 1.0 at TE = 0. (D) S parameter, and (E) Normalized differences (S
nd). Note that the nearly horizontal contours in S
nd reflect the T
1 invariance (TR = 1.4 s).
Echo signal, signal difference (S), and normalized difference (S
nd) as a function of T
1 and T
2 with = 1.0 for a 32‐echo constant flip angle train. (A) Comparison of flip angles and responses (S120 and S180) between the constant 120° (circles) and 180° (squares) trains for T
2 = 100 ms and T
1 = 1 s (flip angles are normalized to 180°). Signal amplitudes are shown for echo 1 (TEeff = 6 ms) (B) and echo 16 (TEeff = 96 ms) (C). The signal has been normalized to the maximum achievable value of 1.0 at TE = 0. (D) S parameter, and (E) Normalized differences (S
nd). Note that the nearly horizontal contours in S
nd reflect the T
1 invariance (TR = 1.4 s).
Steps for sequence evaluation
In summary, optimization of the sequence includes the following steps:Choose a flip angle evolution either manually or based on clinical implementations of 3D FSE sequences with a specified flip angle array, ETL, TR, and echo spacing.Simulate the sequence (in T
1, T
2, and space) using Bloch equation or EPG simulations to produce a dictionary of signal‐decay values.Compute oT2 and determine the optimal TEs (using oT2 and visually inspecting contrast) for the PD (S
1) and T
2 (S
2) images.Calculate S
d, S
nd, and the combined parameter, S, using the optimal TEs from oT2 and validate the constant T
1 approximation (using 3D plots of S
nd in T
2 and T
1 space).Determine S response across T
2s of interest (using 3D plots of S in T
2 and T
1 space).Repeat steps 1–5 for different sequences to evaluate different flip angle evolutions on T
2 mapping and compare S.Selection optimization will be discussed in examples given in the following sections.
Constant echo train
Simulations were conducted to model echo trains with ETLs of 16 (the 3D analogue of the 2D PD‐T2 FSE sequence) and 32. Contrast is typically reflective of PD‐weighted and T
2‐weighted images, with degree dependent on the selected echoes, similarly to 2D PD‐T2 methods. Therefore, the first echo in the train was chosen for S
1, and a middle echo was used for S
2. Constant cases used flip angle trains with PSS initialization through a one‐ahead approach described in Hennig et al.
The signal difference, S, was calculated across T
1, T
2 and , and while S
nd can still be used to help understand the constant case, its utility is greater in more complex trains.
Generic echo trains
A 96 ETL with variable refocusing angles using the PSS‐ramp‐plateau‐ramp flip angle approach described earlier was investigated. Generally, this shape is used for T
2‐weighted contrast in 3D‐FSE images and sustains relatively high signal at later TEs. (For example, the T
2‐weighted variable flip angle train for our Siemens Prisma produces a plateau for the T
2 ~ 100 ms species.) The signal plateau was varied between responses for T
2 = 50 (P50), 100 (P100), and 200 (P200) ms species to investigate the T
2 spread based on S. Echo‐time selection was performed through oT2 optimization.
analysis
As is also a variable used in the fitting, S
nd can be computed for expected ranges (S
nd []), given the optimized TEs for each sequence to evaluate the variation of T
2 fitting response in terms of and ensure that a large signal difference is present to provide unique mapping solutions. For T
2 values of 20–100 ms, ranging from 0.7 to 1.25, and a constant T
1 = 1 s, signal parameters S, S
nd, and S were computed. Additionally, experiments were performed on a doped spherical phantom to measure P200 T
2 mapping performance in varying environments, detailed in the next section.
Phantom and healthy brain experiments
All data were acquired using a Siemens Prisma (Erlangen, Germany) 3 T scanner with an 80 mT/m gradient set. A Siemens 64‐channel head and neck array was used for signal reception. The phantom consisted of six 50 mL tubes filled with water and doped with magnesium chloride to simulate different T
2 environments ranging from 40 to 82 ms (with T
1 ranging from 720 to 1950 ms). Sequences tested included 2D MESE, 2D PD‐T2, C32 (constant flip angle 32‐echo train), C16, P200, P100, and P50 (acquisition parameters detailed subsequently). Experiments using an additional spherical phantom (20 cm diameter, doped with 1.25 g NiSO4/L) were conducted to measure the T
2 mapping performance of P200 in an environment mimicking the variation in the human head.The volunteer study consisted of 3 healthy subjects giving informed consent with ages of 23, 30, and 42 years. Parameters for 3D‐FSE quantification included two independent experiments with optimized TEs (P200: 54 and 294 ms; P100: 78 and 318 ms; P50: 60 and 198 ms; C32: 6 and 96 ms), TR = 1.4 s, an isotropic resolution of 1 mm3 (matrix size of 256 × 256 × 208), parallel imaging (GRAPPA) with an acceleration factor of 2 in the phase‐encode dimension, echo spacing of 6 ms, and a TA depending on ETL (96 for P200, P100 and P50; and 32 for C32). A ‐mapping sequence using the Bloch‐Siegert
method was included to supply flip angles to the quantification, with parameters of TE = 2.24 ms, TR = 4.6 ms, flip angle = 5°, voxel size = 1.1 × 1.1 × 3.0 mm3, matrix size = 192 × 192 × 36, and TA = 33 s. Furthermore, an MP‐RAGE sequence (1 mm3 isotropic, TE = 2.27 ms, TI = 1800 ms) was used as reference for registering all images using FSL FLIRT.
,
,
Finally, compared with previous 2D methods, standard 2D PD‐T2 (16 ETL, TE1 = 10 ms, TE2 = 90 ms, 10‐ms echo spacing, TR = 7000 ms, flip angle [constant] = 165°, 1.0 × 1.0 mm2 in‐plane resolution with 3‐mm slices [35 total], TA = 3:58 min) and 2D MESE images (32 ETL, 10‐ms echo spacing, TR = 3000 ms, flip angle = 180°, 1.0 × 1.0 mm2 in‐plane resolution, 5‐mm slice thickness [6 total slices], TA = 6:09 min) were acquired.
In vivo T
2 analysis
Several regions were segmented manually on matching axial slices of the T
2 maps for each method (2D PD‐T2, MESE, and 3D FSE). Included regions of interest were the frontal white matter (64 total pixels), inferior longitudinal fasciculus (20), cortical gray matter (21), corticospinal tract/internal capsule (16), putamen (56), globus pallidus (48), thalamus (48), and red nucleus (25). Comparison of methods was made using Bland‐Altman analysis.
,
Differences of each method compared with MESE for all regions and subjects were computed and tested for normality using a one‐sided Kolmogorov‐Smirnov test. An acceptable range of tolerance between two measurements was defined as the T
2 mean SD of each tissue measurement for all MESE regions in all subjects. As measured in Section 3, the mean MESE SD was equal to 4 ms; consequently, by this definition, two methods agree within limits if the confidence interval (CI) range is less than or equal to 8 ms.
RESULTS
Constant echo train simulations
Simulations for the constant 32 echo train with flip angle of 120°, echo selections of 1 (S
1, effective TE [TEeff] of 6 ms) and 16 (S
2, [TEeff] = 96 ms), and nominal of 1.0 are shown in Figure 1B,C, with signal as a function of T
1 and T
2. The signal response at T
2 = 100 ms and T
1 = 1 s is illustrated in Figure 1A (as compared with a constant 180° train). These echoes were chosen to mimic the 2D PD‐T2 contrasts (typically using echoes 1 and 9 in a 16‐echo sequence, corresponding to TEeff of 10 ms and 90 ms, respectively). Maximum echo signal variance (S
nd) across in vivo T
2s was calculated to be 0.58 with an average of 0.51 (measured between 30 and 100 ms at a constant T
1 of 1 s). The nearly constant T
1 variation illustrated by S
nd (Figure 1E) indicates mapping using the constant 32‐echo sequence with 120° flip angle is practically T
1 invariant. Consequently, the assumption of a constant T
1 in T
2 quantification to reduce the number of variables is valid. Due to the T
1 invariance, further reduced TR cases (32 echoes with TR of 0.5 s and 16 echoes with minimal TR of 0.2 s) were explored but did not provide sufficient SNR in phantom and in vivo experiments.
Generic train simulations
The response and flip angles for a 3D‐FSE sequence commonly used for T
2‐weighted images with signal plateau at T
2 = 100 ms (P100) are shown in Figure 2. Figure 2A illustrates the signal response of three T
2 isochromats (50, 100, and 200 ms) to the P100 sequence and shows the flip angle evolution used to produce the signal. Further analysis of the spread of different T
2 decays for a T
2 range of 20 to 200 ms is shown in Figure 2B (constant T
1), and T
1 decays (range of 0.6 to 1.6 s, constant T
2) in Figure 2C. Because the spread of response based on individual T
2 isochromats in Figure 2B is greatest following the plateau (echo number 53, TEeff = 318 ms), it follows intuitively that this point would make an ideal second echo candidate for mapping T
2s in combination with an earlier echo in the train. Compared with the diverse variation of T
2 decays in Figure 2B, the amount of T
1 variation is noticeably constrained (Figure 2C), indicating a small variation in signal due to differences in T
1.
FIGURE 2
(A) Response at T
2 = 200 ms (blue), T
2 = 100 ms (orange), and T
2 = 50 ms (black) to the 96‐echo flip angle evolution optimized for 100 ms (P100) prescribed in purple (secondary y‐axis). and T
1 were held constant for this illustration with values of 1.0 and 1.0 s, respectively. Note the long ramp with sustained T
2 = 100 ms signal. The variation of the response is shown with respect to T
2 (B) and T
1 (C), where each line shows a different T
2 (in the range of 20–200 ms) or T
1 (with range of 0.6 to 1.6 s) species. Echo spacing is 6 ms; therefore, 60 ms denotes 10 echoes
(A) Response at T
2 = 200 ms (blue), T
2 = 100 ms (orange), and T
2 = 50 ms (black) to the 96‐echo flip angle evolution optimized for 100 ms (P100) prescribed in purple (secondary y‐axis). and T
1 were held constant for this illustration with values of 1.0 and 1.0 s, respectively. Note the long ramp with sustained T
2 = 100 ms signal. The variation of the response is shown with respect to T
2 (B) and T
1 (C), where each line shows a different T
2 (in the range of 20–200 ms) or T
1 (with range of 0.6 to 1.6 s) species. Echo spacing is 6 ms; therefore, 60 ms denotes 10 echoesTo investigate effects of adjusting the plateau, three variable flip angle sequences with plateaus occurring at 50 (P50), 100 (P100), and 200 ms (P200) were designed and simulated, and the decay curves are shown in Figure 3. As mentioned, P100 was used as the starting point, as it mimics the common T
2‐weighted variable flip angle train used for clinical routine, while preliminary simulations showed an improved result in T
2 mapping when the plateau was optimized at a higher T
2 (P200). The P50 case was used to illustrate effects of using a lower T
2 plateau. The response of T
2 species in the in vivo range of 30–100 ms is illustrated for each variable flip angle train (P200, 3A; P100, 3B; P50, 3C) and compared with the constant flip angle, 32‐echo case (Figure 3D). The required flip angle arrays are shown in each as the dotted line on the secondary y‐axis. In the case of P50, larger flip angles are needed earlier in the evolution to maintain the plateau due to the more rapid transverse decay of the T
2 = 50 ms isochromat in comparison to species with slower relaxation rates. There is substantial T
2 variation in all sequences, indicating that T
2 mapping utility is possible with variable trains, with the most spreading occurring for the P50 case.
FIGURE 3
Spread of decay curves (multiple lines) for different T
2 values ranging from 20 to 200 ms for variable flip angle sequences. Sequences have been optimized to produce sustained signal at plateaus of T
2 = 200 ms (P200) (A), T
2 = 100 ms (P100) (B), and T
2 = 50 ms (P50) (C). The decay of the constant case is shown in (D). Vertical bars indicate the location of the chosen, optimized echoes for the T
2 mapping experiment. The flip angle train is shown as a dashed line at the top, corresponding to the secondary y‐axis. All experiments used an echo spacing of 6 ms, with a total of 96 echoes (A‐C) and 32 echoes (D)
Spread of decay curves (multiple lines) for different T
2 values ranging from 20 to 200 ms for variable flip angle sequences. Sequences have been optimized to produce sustained signal at plateaus of T
2 = 200 ms (P200) (A), T
2 = 100 ms (P100) (B), and T
2 = 50 ms (P50) (C). The decay of the constant case is shown in (D). Vertical bars indicate the location of the chosen, optimized echoes for the T
2 mapping experiment. The flip angle train is shown as a dashed line at the top, corresponding to the secondary y‐axis. All experiments used an echo spacing of 6 ms, with a total of 96 echoes (A‐C) and 32 echoes (D)Sequence evaluation and calculation of optimal TEs through S and oT2 was performed (Figure 4), and S, S
nd, and S for the optimal TEs are shown in Figure 4A–C. Although there is some T
1 variation at higher T
2 values (curvature of the plane in the T
1 direction in S
nd), there remains a very small variance of signal due to T
1 contributions in the in vivo range (<3% for P200). The largest signal difference occurs for P200 (4A), with a gradual decline in difference toward higher T
2 values. Although P50 shows some ability to distinguish T
2 values < 40 ms, S is close to zero around the T
2 values of interest (T
2 = 60 ms), and therefore should not perform as well as P100 or P200. P100 varies similarly to P200, with a reduced T
2 mapping capacity due to lower available contrast.
FIGURE 4
Signal difference (S) (A), normalized difference (S
nd) (B), and T
2 mapping performance (S) (C) for the three investigated variable trains. The greatest signal difference occurs in the P200 case (highest plane in [A]) showing the greatest available SNR after signal difference. The T
2 variation shown in S
nd is greatest for P50 (largest curvature in T
2 direction in [B]). C, S shows the best potential for accurate mapping occurs for the P200 sequence (a combination of T
2 variation and SNR). D‐F, Optimized TEs using the ETL × ETL matrix approach for P50, P100, and P200. The coordinates of the peak of the optimization (oT2) are shown in brackets (proton density [PD] echo, T
2 echo). All sequences used an echo spacing of 6 ms. Note that the oT2 parameter is in arbitrary units
Signal difference (S) (A), normalized difference (S
nd) (B), and T
2 mapping performance (S) (C) for the three investigated variable trains. The greatest signal difference occurs in the P200 case (highest plane in [A]) showing the greatest available SNR after signal difference. The T
2 variation shown in S
nd is greatest for P50 (largest curvature in T
2 direction in [B]). C, S shows the best potential for accurate mapping occurs for the P200 sequence (a combination of T
2 variation and SNR). D‐F, Optimized TEs using the ETL × ETL matrix approach for P50, P100, and P200. The coordinates of the peak of the optimization (oT2) are shown in brackets (proton density [PD] echo, T
2 echo). All sequences used an echo spacing of 6 ms. Note that the oT2 parameter is in arbitrary unitsAnalysis of oT2 (echo selection) resulted in optimized echoes in each case occurring near the beginning and before the end of the plateau (Figure 4D–F). Consequently, an earlier echo is prescribed for S
2 (33, TEeff = 198 ms) in P50 versus S
2 for the P200 case (echo 49, TEeff = 294). Optimal TEeffs are shown as coordinates in the format (TEeffPD, TEeffT2). There is a considerable amount of variation in the P50 plot (Figure 4D) and a relative singular peak for optimal TEs, while more flexibility in TE selection is afforded with P100 and P200 (i.e., similar results can be achieved at several combinations due to the elongated maximum). Note that the optimal TEs in Figure 4D–F were used to produce the signal parameter maps in Figure 4A–C.Evaluation of signal parameters in terms of T
2 and space using an expected range of 0.7 to 1.25 (where nominal = 1.0) and a constant T
1 of 1 s (Supporting Information Figure S1) resulted in significant variation in both T
2 and directions. This is contrasted to measured signal parameters in T
2 and T
1 space (Figure 4), where variation was constrained primarily to T
2, indicating a need to take both T
2 and into account when performing fitting using an acquired map. P200 showed a signal variation of 0.1 (T
2 = 100 ms, B
1 = 0.7) to 0.95 (T
2 = 30 ms, = 1.25), with a similar shape to S
nd (Figure 4B) in the T
2 direction. P100 showed less variation (0 to 0.8) in T
2, while the P50 case produced similarly shaped S
nd () maps to the P50 case in Figure 4B, with the least amount of total variation of 0 to 0.6 in T
2. The S analysis indicated best performance should be expected for the P200 case, with the greatest available contrast and most variation in T
2 and directions.
Phantom experiments
Results of experiments using the six‐cylinder phantom are shown in Figure 5, with T
2 values recorded in Table 1. The SNR measurements quoted were measured using the first vial image (MESE T
2 = 82 ms), and contrast‐to‐noise ratio (CNR) used an average for all vials (corrected to voxel size). The constant train with 32 echoes (C32, 5C) and P200 (5F) compared the most favorably with the MESE T
2 maps, having percent differences of 2.2 and 1.3, respectively. As the TR for C32 was equivalent to the variable trains (1.4 s), the time for acquisition was three times longer due to a shorter ETL, making the C32 case less feasible for clinical studies. All phantom T
2 maps show more variation and noise compared with MESE. C16 (constant flip angle train with 16 echoes and TR = 0.2 s, not shown) and P50 (5D)—as predicted from the simulations—were the worst performers, with large deviations from MESE, and therefore were not included in the in vivo study. The SNR decreases of the variable echo trains from MESE and 2D PD‐T2 are due to increased resolution in the slice direction and longer ETL. Note that C32 maintains SNR near the 2D cases using an ETL one third the length of the P200/P100/P50 cases. Individual vial CNR measurements showed a small decrease with increasing T
2, as expected (Supporting Information Figure S2 and Supporting Information Table S1), although a strong relationship between CNR and accuracy of estimated T
2 values was not present between the different T
2 vials. However, the overall average CNR did correlate with increased accuracy of T
2 value estimation, indicating that a threshold of about 200 is required for accurate T
2 prediction. All sequences with CNRs below 200 performed poorly (P100, 79; P50, 88; C16, 140).
FIGURE 5
Phantom T
2 maps for 2D multi‐echo spin‐echo (MESE) (A) and 2D PD‐T2 (B) experiments compared with the constant 32‐echo train (C32) (C) and variable trains P50 (D), P100 (E), and P200 (F). The expected circular axial cuts are slightly elongated vertically in the MESE and PD‐T2 cases due to thicker slices—illustrating the slight off‐axis alignment of the phantom. Nominal phantom T
2 values that mimic in vivo brain tissues from the top left, clockwise, as measured from MESE, were 82.2, 73.9, 57.9, 40.0, 64.1, and 46.0 ms. The color bars are in seconds
TABLE 1
Phantom T
2 measurements for constant and variable flip angle trains
Train
Time (s)
SNR (PD‐weighted)
SNR (T2‐weighted)
1b
2
3
4
5
6
Average CNRc
Diff from MESE (%)d
MESE
285
343
57a
82.2 ± 0.6
73.9 ± 0.5
57.9 ± 0.3
40.0 ± 0.0
64.1 ± 0.7
46.0 ± 0.1
418
–
PD‐T2
182
328
178
82.8 ± 3.2
73.7 ± 2.9
57.5 ± 1.7
39.5 ± 1.1
63.8 ± 2.2
45.5 ± 1.2
336
0.7
C32
528
281
196
84.7 ± 1.4
75.1 ± 1.1
58.9 ± 0.8
41.0 ± 0.7
65.2 ± 0.1
47.1 ± 0.6
818
2.6
C16
150
71
58
86.3 ± 6.1
77.4 ± 5.7
63.4 ± 3.5
46.4 ± 2.0
68.6 ± 4.6
53.5 ± 2.2
140
10.4
P200
176
97
71
79.9 ± 2.3
72.4 ± 2.3
57.1 ± 1.9
40.2 ± 1.6
64.4 ± 2.4
45.7 ± 1.7
203
1.5
P100
175
58
80
75.4 ± 3.5
67.7 ± 3.8
52.4 ± 2.5
35.7 ± 1.5
60.4 ± 3.4
41.3 ± 1.7
79
9.5
P50
176
70
96
43.4 ± 1.5
41.4 ± 1.4
36.2 ± 1.0
29.5 ± 0.8
39.6 ± 1.3
31.5 ± 0.8
88
37.7
Abbreviation: CNR, contrast‐to‐noise ratio.
SNR for MESE T
2 was measured on the 16th echo.
Mean (in ms) ± SD.
Corrected for voxel size.
Difference from MESE was measured as the mean of percent differences for each cylinder.
Phantom T
2 maps for 2D multi‐echo spin‐echo (MESE) (A) and 2D PD‐T2 (B) experiments compared with the constant 32‐echo train (C32) (C) and variable trains P50 (D), P100 (E), and P200 (F). The expected circular axial cuts are slightly elongated vertically in the MESE and PD‐T2 cases due to thicker slices—illustrating the slight off‐axis alignment of the phantom. Nominal phantom T
2 values that mimic in vivo brain tissues from the top left, clockwise, as measured from MESE, were 82.2, 73.9, 57.9, 40.0, 64.1, and 46.0 ms. The color bars are in secondsPhantom T
2 measurements for constant and variable flip angle trainsAbbreviation: CNR, contrast‐to‐noise ratio.SNR for MESE T
2 was measured on the 16th echo.Mean (in ms) ± SD.Corrected for voxel size.Difference from MESE was measured as the mean of percent differences for each cylinder.Spherical phantom experiments measuring T
2 variation due to (Supporting Information Figure S1D–F), illustrated a relatively constant T
2 across the phantom (Supporting Information Figure S1F), with an average value of 135 ± 3.5 ms and a minimum value of 117 ms occurring at the periphery. varied between 0.76 and 1.26 (mean = 1.12, SD = 0.13). Reducing the measurement circle (excluding peripheral values) to a diameter of 18 cm resulted in a minimum T
2 of 126 ms. This indicates that extreme edge values of low may experience reduced T
2 values.
In vivo experiments
Figure 6 shows demonstrative weighted images for P200 (PD image [Figure 6A] and T
2 image [Figure 6B]) and 2D PD‐T2 (Figure 6C,D) used for T
2 mapping. The map is also included (Figure 6E) to display the measured variation across the brain. The 3D‐PD image (Figure 6A) shows flatter PD tissue contrast than the 2D first‐echo image (Figure 6C), and less signal from CSF. This is likely due to the more pronounced T
1 effects in 3D (TR = 1.4 s for 3D, TR = 7 s for PD‐T2) as well as the removal of 2D incidental magnetization transfer effects.
T
2 maps for the two cases are shown in Figure 7 for two locations: (1) corticospinal tract and deep gray matter (Figure 7A,B) and (2) cortical gray matter, cerebellum, and superior longitudinal fasciculus (Figure 7C,D). A marked increase in resolution is apparent in the P200 case, with easier delineation in gray‐matter folding and cerebellar structures. White‐matter tracts including the corticospinal tract have less blurring and more definition in P200.
FIGURE 6
Raw (no smoothing or gradient removal) weighted images acquired for use in T
2 mapping. A, P200 3D sequence, PD echo (echo 9, TEeff = 54 ms). B, T
2 echo of the 3D P200 sequence (echo 49, TEeff = 294 ms). C, PD image for the dual‐echo 2D sequence. D, T
2 image for the dual‐echo 2D sequence. E, Normalized map
FIGURE 7
T
2 maps for the P200 variable flip angle case (A,C) and 2D PD‐T2 (B,D), highlighting two locations. In #1 (A,B), the corticospinal tract (CST) is clearly visible on the 3D sagittal (sag) and coronal (cor) slices, while deep GM structures are illustrated in the axial view, particularly the internal capsule, thalamus, globus pallidus, and putamen. Location #2 features cortical GM folding (sag), a highly resolved cerebellum (Cer) in coronal 3D (C), and the tracts of the superior longitudinal fasciculus (SLF, axial). Of particular note is the three‐fold improved slice resolution in the 3D sagittal and coronal reformats (see sag and cor #2 and structural nuance present in the CST in #1)
Raw (no smoothing or gradient removal) weighted images acquired for use in T
2 mapping. A, P200 3D sequence, PD echo (echo 9, TEeff = 54 ms). B, T
2 echo of the 3D P200 sequence (echo 49, TEeff = 294 ms). C, PD image for the dual‐echo 2D sequence. D, T
2 image for the dual‐echo 2D sequence. E, Normalized mapT
2 maps for the P200 variable flip angle case (A,C) and 2D PD‐T2 (B,D), highlighting two locations. In #1 (A,B), the corticospinal tract (CST) is clearly visible on the 3D sagittal (sag) and coronal (cor) slices, while deep GM structures are illustrated in the axial view, particularly the internal capsule, thalamus, globus pallidus, and putamen. Location #2 features cortical GM folding (sag), a highly resolved cerebellum (Cer) in coronal 3D (C), and the tracts of the superior longitudinal fasciculus (SLF, axial). Of particular note is the three‐fold improved slice resolution in the 3D sagittal and coronal reformats (see sag and cor #2 and structural nuance present in the CST in #1)Several regions were manually segmented (locations shown in Supporting Information Figure S3), and measured T
2 values are provided in Table 2. T
2 values were similar in each region when comparing between subjects using the same method (other than red nucleus, likely due to relatively small number of voxels), with the main variance occurring between methods. When comparing MESE to 3D methods, the largest deviations occurred in the frontal white matter and internal capsule. P100 consistently projected lower T
2 values than other methods and produced a noisier T
2 map as predicted by S.
TABLE 2
T
2 values from 3 healthy subjects for 3D and 2D methods
Time (s)
SNR PD
SNR T2
FWM
a
,
b
ILF
CGM
IC
Putamen
GP
Thalamus
RN
Diff from MESE (%)
V1
MESE
396
180
65.0c
59.5 ± 1.2
73.8 ± 4.3
73.9 ± 3.0
85.0 ± 11
64.5 ± 4.1
50.4 ± 1.8
63.3 ± 2.7
51.6 ± 1.7
PD‐T2
238
155
73.8
57.4 ± 2.0
73.1 ± 3.5
74.0 ± 3.7
82.5 ± 3.7
61.5 ± 2.0
47.2 ± 2.6
60.1 ± 3.1
48.7 ± 3.0
3.7
P200
646
95.5
114
54.5 ± 1.3
73.6 ± 3.3
74.6 ± 4.2
79.4 ± 3.8
63.1 ± 2.6
48.6 ± 2.2
61.1 ± 2.0
49.8 ± 2.5
3.6
P100
646
56.5
43.9
49.4 ± 2.5
73.0 ± 5.3
72.6 ± 4.7
72.5 ± 3.0
62.7 ± 3.8
49.6 ± 2.4
62.7 ± 2.4
47.8 ± 3.1
5.9
V2
MESE
396
203
79.0c
58.5 ± 1.5
74.6 ± 3.8
75.6 ± 3.9
81.4 ± 9.7
62.6 ± 3.5
52.0 ± 2.2
63.6 ± 3.0
–d
PD‐T2
238
175
91.0
55.6 ± 2.4
70.0 ± 2.7
74.0 ± 3.6
80.3 ± 5.6
62.3 ± 3.2
48.8 ± 3.6
61.7 ± 3.8
50.3 ± 3.5
3.5
P200
646
43.0
52.3
55.1 ± 2.6
72.0 ± 2.2
73.0 ± 4.7
78.9 ± 2.8
62.2 ± 3.4
51.6 ± 2.1
65.6 ± 2.0
61.1 ± 1.7
2.9
P100
646
33.8
37.2
49.4 ± 2.2
71.9 ± 2.5
73.8 ± 0.9
74.3 ± 1.9
61.0 ± 3.3
50.2 ± 3.3
62.1 ± 1.8
52.4 ± 3.0
5.5
V3
MESE
396
212
78.9c
56.4 ± 2.7
71.0 ± 4.0
70.5 ± 2.8
81.9 ± 13
63.4 ± 3.6
46.9 ± 2.3
63.3 ± 1.8
53.4 ± 1.9
PD‐T2
238
132
91.6
55.6 ± 2.8
73.3 ± 3.4
72.6 ± 3.0
82.3 ± 6.1
59.8 ± 3.0
47.6 ± 5.7
62.6 ± 4.0
50.3 ± 5.2
2.8
P200
646
78.1
76.2
52.8 ± 2.7
70.5 ± 2.9
72.3 ± 3.2
76.3 ± 3.1
62.5 ± 2.7
49.8 ± 3.3
62.5 ± 2.7
53.1 ± 3.7
3.2
P100
646
62.9
73.4
45.1 ± 2.1
67.2 ± 1.9
82.1 ± 7.4
74.3 ± 4.7
57.2 ± 2.6
47.2 ± 3.4
59.0 ± 2.5
47.7 ± 3.6
9.9
Manual segmentation locations are as follows with region‐of‐interest pixel counts in brackets: FWM, frontal white matter (64); ILF, inferior longitudinal fasciculus (20); CGM, cortical gray matter (21); IC, corticospinal tract/internal capsule (16); putamen (56); GP, globus pallidus (48); thalamus (48); and RN, red nucleus (25).
Mean value (in ms) ± SD.
SNR for the MESE T
2 case used the 16th echo image.
Red nuclei were not included in the MESE slices for the second subject.
T
2 values from 3 healthy subjects for 3D and 2D methodsManual segmentation locations are as follows with region‐of‐interest pixel counts in brackets: FWM, frontal white matter (64); ILF, inferior longitudinal fasciculus (20); CGM, cortical gray matter (21); IC, corticospinal tract/internal capsule (16); putamen (56); GP, globus pallidus (48); thalamus (48); and RN, red nucleus (25).Mean value (in ms) ± SD.SNR for the MESE T
2 case used the 16th echo image.Red nuclei were not included in the MESE slices for the second subject.Using Bland‐Altman analysis, differences between MESE T
2 values for each region and the other methods were computed and tested for normality. The resultant plots illustrating the differences of the measurements are shown in Figure 8, with color coding based on the three different volunteers. PD‐T2 has the smallest CI followed by P200 and P100. In each case, a negative bias occurs ranging from 1.5 ms in the P200 case to 3.7 ms in P100, indicating underestimated T
2 values when compared with MESE. As expected, P100 is the poorest match for MESE, with CI ranging from 13 ms from the mean to +6.1 ms, and a total range of 19.1 ms, well beyond the defined acceptable value of 8 ms. P200 and PD‐T2 are both within acceptable agreement, with CI ranges of 7.8 and 7.3 ms, respectively.
FIGURE 8
Correlation (left) and Bland‐Altman (right) plots for comparison of MESE with 2D PD‐T2 (A), P200 (B), and P100 (C) methods. Limits of agreement (confidence intervals) are shown in the Bland‐Altman plots as dotted horizontal lines, with the mean bias as solid. Individual volunteers have been color‐coded (volunteer 1, blue; volunteer 2, black; and volunteer 3, red)
Correlation (left) and Bland‐Altman (right) plots for comparison of MESE with 2D PD‐T2 (A), P200 (B), and P100 (C) methods. Limits of agreement (confidence intervals) are shown in the Bland‐Altman plots as dotted horizontal lines, with the mean bias as solid. Individual volunteers have been color‐coded (volunteer 1, blue; volunteer 2, black; and volunteer 3, red)
DISCUSSION
A development of a dual‐echo 3D T
2 mapping method with optimized flip angles is presented and compared with 2D PD‐T2 and MESE standard methods. Simulations were used to compute signal response for optimization of flip angles and other sequence parameters as well as dictionary production for T
2 mapping. Three variable flip angle trains and a 32‐echo constant flip angle case (reflecting 2D methods) were compared and verified in phantom measurements, and tissue T
2 values were compared with 2D methods using the variable trains. The standard 2D MESE measures for in vivo tissues were in agreement with past studies from our laboratory.The P200 variable flip angle train performed better than P100, as expected from simulations and phantom experiments, and was within the defined acceptable limits of agreement with MESE based on Bland‐Altman analysis. Consequently, this allows higher spatial resolution and simplicity of modeling RF pulses, offering potential improvements over 2D methods. The PD‐T2 method also showed a slight deviation from MESE values with an approximately 2‐ms negative bias, similar to P200. Both methods use substantially fewer TEs for mapping compared with MESE (2 vs. 32) and likely contain similar errors in decay curve matching.Although T
2 mapping provides a quantitative measure of tissue in pathology, clinical exams continue to rely on weighted images for diagnosis. High‐resolution images with typical contrasts, in this case PD‐weighted and T
2‐weighted, were part of the design criteria for the 3D mapping sequence. Recent methods focusing more on quantitative imaging aspects, such as the stack of stars technique,
show promise in speeding up acquisition—the main drawback of the dual‐echo 3D approach—while providing high‐resolution 3D volumes with T
2‐weighted images (~7 min vs. ~11 min for dual‐echo 3D), although further comparison with previous methods and literature T
2 values is necessary. Additional gradient modules can also be introduced for faster sampling,
or can replace part or all of the typical SE readouts. For example, in a recent study,
3D‐FLASH readouts were used to determine quantitative measures in brain, although T
2 values are elevated 10%–15% from those presented in the 3D dual‐echo case previously. Combination SE/GE sequences introduce complexities in modeling for dictionary values compared with the Bloch or EPG nonselective approaches; therefore, reference phantom values are required or machine learning techniques are used, as commonly employed in MR fingerprinting.
,
Furthermore, the 3D dual‐echo approaches developed here are relatively T
1‐invariant and substantial simplifications can be made when modeling signal evolution.The acquisition of multiple volumes for quantification is subject to motion and can lead to erosion of resolution in the resultant T
2 map. As classification of tissue values is typically based on voxel averaging of segmented regions, T
2 reporting should not be greatly affected. It should be noted that using extended regions of interest for quantification does not necessarily degrade the gains of 3D methods—a finer region of interest (sharper edges including less of the surrounding tissues) can be delineated, resulting in less region‐of‐interest contamination and more accurate T
2 measures. Motion‐correction techniques could alleviate the problem
; however, a combined dual‐echo sequence—a true analogue of 2D PD‐T2—would effectively remove interscan motion. The dual‐contrast sequence could be designed to share echoes
to reduce TA while maintaining production of PD‐weighted and T
2‐weighted base images, with extension to multi‐echo methods for further quantitative measures (such as T
1). Techniques to alleviate intrascan motion have recently been proposed for constant flip angle train FSE sequences
and will be investigated for variable flip angle trains in future work.Limitations of our study include the use of healthy‐tissue T
1 and T
2 ranges in the simulations (800–1400 ms and 40–90 ms, respectively) and the adjustment of flip angle arrays within the sequence source code. The T
1 and T
2 ranges used are insufficient to quantify the long T
2 times in CSF, although the focus of this study was the viability of variable flip angle trains using a T
2 range corresponding to healthy tissues. In clinical study, the expected T
2 range for the pathology should be considered. In terms of sequence programming, the modification of the flip angle array is the only requirement to enable these sequences and is possible using any scanner development software. However, some sites may not have access to the source code for sequences, making the implementation difficult. In these cases, researchers can test site‐specific use of the T
2 variable flip angle train sequence (CUBE, VISTA, or SPACE) through manipulation of TEeff for PD‐weighted and T
2‐weighted image production. Additionally, future work should include a robust discussion of the point‐spread function effects on the selected echo trains. Although the results shown here have been compared with MESE experiments that do not exhibit this problem, they are limited to the sequence parameters explored, and may not necessarily be extrapolated to variations in the sequence due to effects including the point‐spread function.Finally, while acquisition of images used for T
2 mapping is on the order of minutes, calculation of T
2 maps based on dictionary fitting using minimization typically takes several hours. Previous studies have reported computation times of 22–50 min for single‐slice MESE experiments,
comparable to our study, whereas PD‐T2 methods are on the order of 5 min/slice. While processing times are generally viewed as a nuisance to researchers, the applicability of T
2 mapping methods to clinical practice may hinge on the future ability to provide an efficient and accurate T
2 map calculation in line with base image acquisition on the scanner.
CONCLUSIONS
The dual‐echo 3D approach using variable flip angle trains is a viable T
2 mapping method based on simulated dictionary values while providing high‐resolution PD‐weighted and T
2‐weighted base images. Optimization of the sequence parameters resulted in tissue T
2 values in brain in agreement with previous studies, indicating that the 3D approach can provide useful T
2 maps with superior anatomical detail in comparison to commonly used 2D methods.FIGURE
S1 Signal difference (S) (A), normalized difference (S
nd) (B), and T
2 mapping performance (S) (C) for the three investigated variable trains with a constant T
1 of 1 s, and ranges of T
2 = [20, 100] ms and = [0.7, 1.25]. A large amount of variation is present in both the T
2 and directions (in contrast to nearly constant performance in the T
1 direction in Figure 4A–C). P200 shows the greatest available contrast (highest plane in A) and best mapping performance (C), followed by P100 and P50. Variation in the direction reinforces the need for maps when fitting decay curves. D‐F, Spherical phantom tests of T
2 mapping performance with variation comparable to that in the human head. Proton density (PD) and T
2 images are illustrated in the split image in (D), with the map in (E) and resulting T
2 map in (F) for the P200 sequence. The mean T
2 was 135 ms, with SD = 3.4 ms, and a range of −13% to +5%. Extreme values occurred at the periphery.FIGURE S2 Phantom experimental data plotting contrast‐to‐noise ratio (CNR) versus T
2, as measured in the multi‐echo spin‐echo (MESE) experiment. The size of the markers corresponds to the percent difference in T
2 value as compared with MESEFIGURE S3 Voxel locations as denoted on volunteer T
2 maps for the eight regions. Abbreviations: FWM, frontal white matter; GM, gray matter; GP, globus pallidus; IC, internal capsule; ILF, inferior longitudinal fasciculus; Pu, putamen; RN, red nucleus; Th, thalamusTABLE S1 Phantom T
2 CNR measurements for constant and variable flip angle trainsClick here for additional data file.
Authors: Thomas Prasloski; Alexander Rauscher; Alex L MacKay; Madeleine Hodgson; Irene M Vavasour; Corree Laule; Burkhard Mädler Journal: Neuroimage Date: 2012-07-06 Impact factor: 6.556
Authors: Mahesh Bharath Keerthivasan; Manojkumar Saranathan; Kevin Johnson; Zhiyang Fu; Craig C Weinkauf; Diego R Martin; Ali Bilgin; Maria I Altbach Journal: Magn Reson Med Date: 2019-03-18 Impact factor: 4.668
Authors: Dan Ma; Vikas Gulani; Nicole Seiberlich; Kecheng Liu; Jeffrey L Sunshine; Jeffrey L Duerk; Mark A Griswold Journal: Nature Date: 2013-03-14 Impact factor: 49.962