Jos J M Westenberg1, Hans C van Assen1, Pieter J van den Boogaard1, Jelle J Goeman2, Hicham Saaid3, Jason Voorneveld4, Johan Bosch4, Sasa Kenjeres5, Tom Claessens6, Pankaj Garg7, Marc Kouwenhoven8, Hildo J Lamb1.
Abstract
PURPOSE: To assess errors associated with EPI-accelerated intracardiac 4D flow MRI (4DEPI) with EPI factor 5, compared with non-EPI gradient echo (4DGRE).Entities:
Keywords: 4D flow MRI; echo planar imaging; intracardiac blood flow; phase-contrast MRI
Mesh:
Year: 2021 PMID: 34866236 PMCID: PMC9300143 DOI: 10.1002/mrm.29112
Source DB: PubMed Journal: Magn Reson Med ISSN: 0740-3194 Impact factor: 3.737
FIGURE 1(A) Phantom setup for experiment 1. A tank with straight tube phantom is positioned inside the MRI scanner. A pump outside the scanner room maintains steady flow, which is calibrated by timed volumetric collection at the returning tube. (B) The orientations of the 4D flow MRI acquisitions with respect to the main flow direction in the tube. In (1), 4DEPI5_ro_coronal refers to the coronally oriented EPI acquisition with the readout gradient aligned parallel to the main flow direction; in (2), 4DEPI5_bp refers to the coronally oriented EPI acquisition with the blip phase‐encoding gradient aligned parallel to the main flow direction; and in (3), 4DEPI5_nbp_axial refers to the axially oriented EPI acquisition with the readout and blip phase‐encoding gradient both perpendicular to the main flow direction. The 4D gradient echo (4DGRE) has the same orientation as 4DEPI5_ro_coronal. The schematically drawn scan volumes are not up to scale with the phantom dimensions
Orientation of the MRI acquisitions for three experiments
| Acquisition | Volume orientation | Main flow direction |
|---|---|---|
| Experiment 1—straight tubes | ||
| 4DEPIx_ro_coronal | Coronal | Readout |
| 4DEPIx_bp_coronal | Coronal | Blip phase‐encoding |
| 4DEPIx_nbp_axial | Axial | Nonblip phase‐encoding |
| 4DGRE_coronal | Coronal | Readout |
| Experiment 2—LV phantom | ||
| 4DEPI5_ro_LA | Axial (long‐axis) | Readout |
| 4DEPI5_bp_LA | Axial (long‐axis) | Blip phase‐encoding |
| 4DEPI5_nbp_SA | Coronal (short‐axis) | Nonblip phase‐encoding |
| 4DGRE_LA | Axial (long‐axis) | Phase‐encoding |
| Experiment 3—healthy volunteers | ||
| 4DEPI5_ro_LA | Long‐axis | Readout |
| 4DEPI5_bp_LA | Long‐axis | Blip phase‐encoding |
| 4DEPI5_nbp_SA | Short‐axis | Nonblip phase‐encoding |
| 4DGRE_LA | Long‐axis | Readout |
For experiment 1, 4DEPIx_ro_coronal refers to EPI acquisition with the readout gradient aligned parallel to the main flow direction and reconstructed planes in coronal orientation; 4DEPIx_bp_coronal refers to EPI acquisition with the blip phase‐encoding gradient aligned parallel to the main flow direction and reconstructed planes in coronal orientation; and 4DEPIx_nbp_axial refers to EPI acquisition with the readout and blip phase‐encoding gradient both perpendicular to the main flow direction and reconstructed planes in axial orientation. The “x” after EPI is either 5 or 3, and refers to EPI factor 5 or 3. For experiments 2 and 3, 4DEPI5_ro_LA refers to EPI factor of 5 acquisition with the readout gradient aligned parallel to the main flow direction; 4DEPI5_bp_LA refers to EPI acquisition with the blip phase‐encoding gradient aligned parallel to the main flow direction; and 4DEPI5_nbp_SA refers to EPI acquisition with the readout and blip phase‐encoding gradient both perpendicular to the main flow direction.
Abbreviations: LA, long‐axis; SA, short‐axis.
FIGURE 2Left‐ventricle (LV) phantom setup in experiment 2. (A) Positioning of the pressurized acrylic tank in the scanner (1 indicates the LV phantom; 2 and 3 indicate the bioprosthetic mitral [MV] and aortic valve [AoV], respectively). (B–D) Illustrations of the gradient orientation for the three 4D‐EPI (4DEPI) and 4DGRE acquisitions (4DEPI5_ro_LA refers to the long‐axis‐oriented EPI factor of 5 acquisition with the readout gradient aligned parallel to the main flow direction; 4DEPI5_bp_LA refers to the long‐axis‐oriented EPI acquisition with the blip phase‐encoding gradient aligned parallel to the main flow direction; and 4DEPI5_nbp_SA refers to the short‐axis‐oriented EPI acquisition with the readout and blip phase‐encoding gradient both perpendicular to the main flow direction). Color‐coded streamlines are shown in the phantom during peak inflow. A blue cylinder‐shaped control volume is positioned below the MV inside the ventricle. The schematically drawn scan volumes are not up to scale with the phantom dimensions
FIGURE 3The MRI images from a healthy volunteer for experiment 3. (A) The four‐chamber view (acquired by cine SSFP) with respect to the scanner axis. (B–D) The gradient orientations for the three 4DEPI and the 4DGRE acquisitions (4DEPI5_ro_LA refers to the long‐axis‐oriented EPI factor of 5 acquisition with the readout gradient aligned parallel to the main flow direction; 4DEPI5_bp_LA refers to the long‐axis‐oriented EPI acquisition with the blip phase‐encoding gradient aligned parallel to the main flow direction; and 4DEPI5_nbp_SA refers to the short‐axis‐oriented EPI acquisition with the readout and blip phase‐encoding gradient both perpendicular to the main flow direction). Color‐coded streamlines are shown in the ventricle during peak inflow. A blue cylinder‐shaped control volume is positioned below the MV inside the ventricle. The schematically drawn scan volumes are not up to scale with the in vivo dimensions. (E,F) Flow‐rate time curves obtained from valve tracking at the MV and AoV, respectively. Abbreviations: LAt, left atrium; RA, right atrium; RV, right ventricle
Flow‐rate assessment for steady flow in a straight tube (experiment 1)
| Acquisition | Volumetric flow rate (mL/s) | MRI‐assessed flow rate (mL/s) | Relative difference (%) (ΔVolumetric) | MRI‐assessed max velocity (cm/s) | Relative difference (%) (Δ4DGRE) | MRI‐assessed flow rate (mL/s) | Relative difference (%) (ΔVolumetric) | MRI‐assessed max velocity (cm/s) | Relative difference (%) (Δ4DGRE) |
|---|---|---|---|---|---|---|---|---|---|
| Tube 1 (diameter 1.0 cm) | Tube 2 (diameter 0.7 cm) | ||||||||
| 4DEPI5_ro_coronal | 50 | 53.6 | 7.2 | 96 | 9.1 | 44.0 | −12.0 | 216 | −1.4 |
| 4DEPI5_bp_coronal | 50 | 53.0 | 6.1 | 97 | 10.2 | 45.0 | −10.0 | 216 | −1.4 |
| 4DEPI5_nbp_axial | 50 | 52.5 | 5.0 | 98 | 11.4 | 48.0 | −4.0 | 215 | −1.8 |
| 4DEPI3_ro_coronal | 50 | 54.0 | 8.0 | 88 | 0.0 | 49.1 | −1.8 | 233 | 6.4 |
| 4DEPI3_bp_coronal | 50 | 52.3 | 4.6 | 93 | 5.7 | 50.0 | 0 | 215 | −1.8 |
| 4DEPI3_nbp_axial | 50 | 51.6 | 3.1 | 92 | 4.5 | 46.7 | −6.6 | 229 | 4.6 |
| 4DGRE_coronal | 50 | 49.4 | −1.2 | 88 | – | 47.8 | −4.4 | 219 | – |
| 4DEPI5_ro_coronal | 23 | 24.3 | 5.7 | 44 | 2.3 | ||||
| 4DEPI5_bp_coronal | 23 | 24.4 | 6.1 | 46 | 7.0 | ||||
| 4DEPI5_nbp_axial | 23 | 23.9 | 3.9 | 42 | −2.3 | ||||
| 4DGRE_coronal | 23 | 24.4 | 3.0 | 43 | – | ||||
4DEPIx_ro refers to acquisition with EPI factor (5 or 3) with the readout gradient aligned parallel to the main flow direction; 4DEPIx_bp refers to EPI acquisition with the blip phase‐encoding gradient aligned parallel to the main flow direction; and 4DEPIx_nbp refers to EPI acquisition with the readout and blip phase‐encoding gradient both perpendicular to the main flow direction.
Voxel‐based comparison of speed sampled inside the control volume at peak inflow in the LV phantom and comparison of median speed and interquartile range (in centimeters/second) (experiment 2)
| Voxel‐based comparison at peak LV inflow | Mean difference ± SEM (cm/s) |
| Pearson correlation ( |
|---|---|---|---|
| 4DEPI5_ro_LA–4DGRE_LA | −1.28 ± 0.12 | << 0.01 | 0.94 ( |
| 4DEPI5_bp_LA–4DGRE_LA | −2.16 ± 0.09 | << 0.01 | 0.97 ( |
| 4DEPI5_nbp_SA–4DGRE_LA | −1.01 ± 0.11 | << 0.01 | 0.95 ( |
| 4DEPI5_ro_LA–4DEPI5_bp_LA | 0.87 ± 0.12 | << 0.01 | 0.94 ( |
| 4DEPI5_ro_LA–4DEPI5_nbp_SA | −0.27 ± 0.08 | << 0.01 | 0.97 ( |
| 4DEPI5_bp_LA–4DEPI5_nbp_SA | −1.14 ± 0.12 | << 0.01 | 0.94 ( |
|
|
|
| |
| 4DEPI5_ro_LA | 24.2 (15.6, 39.7) | 0.945 | |
| 4DEPI5_bp_LA | 24.2 (15.6, 40.6) | 0.954 | |
| 4DEPI5_nbp_SA | 24.8 (15.6, 42.9) | 0.981 | |
| 4DGRE_LA | 25.4 (15.7, 42.1) | ||
4DEPI5_ro_LA refers to EPI factor of 5 acquisition with the readout gradient aligned parallel to the main flow direction; 4DEPI5_bp_LA refers to EPI acquisition with the blip phase‐encoding gradient aligned parallel to the main flow direction; and 4DEPI5_nbp_SA refers to EPI acquisition with the readout and blip phase‐encoding gradient both perpendicular to the main flow direction. Abbreviation: IQR, interquartile range.
FIGURE 4Artifacts present in velocity images of one 4DEPI5_ro_LA data set in a healthy volunteer (arrows in [B] and [D]). (A,B) Systolic phase. (C,D) Diastolic phase. Artifacts appear with phase wrapping but in areas with moderate velocities, and therefore remain unresolved by anti‐aliasing algorithm. This particular data set was removed from analysis
Valvular flow assessment by automated retrospective valve tracking over the MV and AoV and median speed assessed from the control volume in the LV of 10 healthy volunteers (experiment 3)
| 4DEPI5_ro_LA | 4DEPI5_bp_LA | 4DEPI5_nbp_SA | 4DGRE_LA | |
|---|---|---|---|---|
| Mean heart rate ± SD (bpm) | 57 ± 7 | 59 ± 7 | 60 ± 6 | 60 ± 4 |
| Mean MV FF ± SD (mL/beat) | 92 ± 26 | 100 ± 23 | 93 ± 21 | 103 ± 17 |
| Mean MV BF ± SD (mL/beat) | 2 ± 2 | 4 ± 4 | 2 ± 1 | 4 ± 3 |
| Mean MV NFV ± SD (mL/beat) | 90 ± 24 | 95 ± 22 | 92 ± 20 | 99 ± 16 |
| Mean AoV FF ± SD (mL/beat) | 93 ± 22 | 99 ± 21 | 98 ± 18 | 99 ± 18 |
| Mean AoV BF ± SD (mL/beat) | 1 ± 1 | 1 ± 1 | 1 ± 1 | 2 ± 2 |
| Mean AoV NFV ± SD (mL/beat) | 93 ± 21 | 96 ± 20 | 97 ± 17 | 96 ± 17 |
| Mean relative difference (AoV NFV − MV NFV) ± SD (%) | 4.2 ± 6.8 | 1.6 ± 6.0 | 6.4 ± 8.5 | −2.9 ± 7.4 |
| t‐test (AoV NFV vs MV NFV) | 0.16 | 0.60 | 0.08 | 0.28 |
| Pearson r (AoV NFV vs MV NFV) | 0.97 ( | 0.97 ( | 0.89 ( | 0.91 ( |
| Median speed at peak diastole (IQR) (cm/s) | 51.4 (41.5, 63.8) | 46.7 (44.8, 61.7) | 56.6 (45.8, 59.4) | 51.6 (45.2, 68.6) |
| Mean EPI/GRE ratio ± SD (over all cardiac phases) | 0.91 ± 0.08 | 0.94 ± 0.10 | 0.93 ± 0.06 |
4DEPI5_ro_LA refers to EPI factor of 5 acquisition with the readout gradient aligned parallel to the main flow direction; 4DEPI5_bp_LA refers to EPI acquisition with the blip phase‐encoding gradient aligned parallel to the main flow direction; and 4DEPI5_nbp_SA refers to EPI acquisition with the readout and blip phase‐encoding gradient both perpendicular to the main flow direction.
Abbreviations: BF, backward flow; FF, forward flow; NFV, net forward volume.
FIGURE 5Net forward volume (NFV) measured over MV and AoV assessed in 10 healthy volunteers by four acquisitions. Bland‐Altman plot is presented in (B)