Sang-Hyun Hwang1,2, John Jeongseok Yang1, Yoon-Hee Oh1, Dae-Hyun Ko1, Heungsup Sung1, Young-Uk Cho1, Seongsoo Jang1, Chan-Jeoung Park1, Heung-Bum Oh3.
Abstract
Affordable point-of-care (POC) CD4 + T lymphocyte counting techniques have been developed as alternatives to flow cytometry-based instruments caring for patients with human immunodeficiency virus (HIV)-1. However, POC CD4 enumeration technologies can be inaccurate. Here, we developed a microparticle-based visual detector of CD4 + T lymphocytes (ImmunoSpin) using microparticles conjugated with anti-CD4 antibodies, independent of microfluidic or fluorescence detection systems. Visual enumeration of CD4 + T cells under conventional light microscope was accurate compared to flow cytometry. Microparticle-tagged CD4 + T cells were well-recognized under a light microscope. ImmunoSpin showed very good precision (coefficients of variation of ImmunoSpin were ≤ 10%) and high correlation with clinical-grade flow cytometry for the enumeration of CD4 + T cells (y = 0.4232 + 0.9485 × for the %CD4 + T cell count, R2 = 0.99). At thresholds of 200 and 350 cells/µL, there was no misclassification of the ImmunoSpin system compared to the reference flow cytometry. ImmunoSpin showed clear differential classification of CD4 + T lymphocytes from granulocytes and monocytes. Because non-fluorescence microparticle-tags and cytospin slides are used in ImmunoSpin, they can be applied to an automatic digital image analyzer. Slide preparation allows long-term storage, no analysis time limitations, and image transfer in remote areas.Entities:
Keywords: CD4 + T cell; Human immunodeficiency virus; Image-based cell counting; ImmunoSpin; Microparticle
Mesh:
Year: 2021 PMID: 34822013 PMCID: PMC8616869 DOI: 10.1007/s00604-021-05070-y
Source DB: PubMed Journal: Mikrochim Acta ISSN: 0026-3672 Impact factor: 5.833
Fig. 1Process illustration for microparticle-tagged image-based cell counting (ImmunoSpin). Representation of specific labeling of CD4 + T cells from whole blood (WB) with the anti-CD4 antibody–microparticle complex and their subsequent RBC lysis and washing step (optional). Then, cytospin preparations were established for the enrichment of leukocytes, including anti-CD4 antibody–microparticle-labeled lymphocytes. Cytospin slides were counterstained with Wright (or methylene blue) stain. At least two cytospin slides were prepared from 100 µL WB and were analyzed
Fig. 2Microscopy images indicating the CD4 + T cells by anti-CD4 antibody-conjugated microparticles and a granulocyte (dashed box) in the background of other leukocytes (above). Several morphologies of microparticle-tagged CD4 + T cells are illustrated (below)
Precision of the ImmunoSpin system
| Mean %CD4 T cells of multicheck controls by ImmunoSpin (20 replicates) | Repeatability (%CV) | Between-run (%CV) | Within laboratory (between days, %CV) | |
|---|---|---|---|---|
| Multicheck Low CD4 control | 12.2% (135 cells/μL) | 4.1% | 7.1% | 8.2% |
| Multicheck Normal CD4 control | 48.0% (714 cells/μL) | 1.0% | 2.1% | 2.4% |
The bias between ImmunoSpin and flow cytometry
| %CD4 T cell | Mean%CD4 T cells by ImmunoSpin | %CD4 T cells by FACSCanto II | Mean bias | Relative bias |
|---|---|---|---|---|
| Multicheck Low CD4 control material (Lot. BM11202L) (3 replicates) | − 0.6% | − 4.6% | ||
| Multicheck Normal CD4 control material (Lot. BM1120N) (4 replicates) | − 0.5% | − 1.0% |
Fig. 3Comparison of ImmunoSpin with flow cytometry. Total WB samples (n = 45) were used for method comparison. CD4 + T cell proportions (above) and absolute CD4 T cell counts (below) measured on the ImmunoSpin were compared to the corresponding CD4 + T cell counting from the flow cytometer BD FACSCanto II (BD Biosciences). The regression equation was y = 0.4232 + 0.9485 × for %CD4 + T cell count (R2 = 0.99) and y = 4.1001 + 0.9590 × for absolute CD4 + T cell count (R2 = 0.99). The slopes were 0.95 (95%CI, 0.93–0.97) for %CD4 + T cell count and 0.96 (95%CI, 0.94–0.99) for absolute CD4 + T cell count. Absolute differences calculated with Bland–Altman plot analysis are plotted
Clinical misclassification at 200 and 350 cells/μL thresholds of clinical importance
| Threshold at 200 cells/μL (flow cytometry) | Threshold at 350 cells/μL (flow cytometry) | |||
|---|---|---|---|---|
| ImmunoSpin | < 200 cells/μL | > 200 cells/μL | < 350 cells/μL | > 350 cells/μL |
| < threshold | 11 | 0 | 27 | 0 |
| ≥ threshold | 0 | 34 | 0 | 18 |