Imelda Omana-Zapata1, Caren Mutschmann2, John Schmitz3, Sarah Gibson4, Kevin Judge1, Monika Aruda Indig5, Beverly Lu1, Doreen Taufman2, Alan M Sanfilippo3, Wendy Shallenberger4, Sharon Graminske5, Rachel McLean6, Rubal I Hsen6, Nicole d'Empaire7, Kimberly Dean1, Maurice O'Gorman6.
Abstract
Clinical flow cytometry is a reliable methodology for whole blood cell phenotyping for different applications. The BD FACSLyric™ system comprises a flow cytometer available in different optical configurations, BD FACSuite™ Clinical software, and optional BD FACS™ Universal Loader. BD FACSuite Clinical software used with BD™ FC Beads and BD CS&T Beads enable universal setup for performance QC, instrument control, data acquisition/storage, online/offline data analysis, and instrument standardization. BD Biosciences sponsored the clinical evaluation of the BD FACSLyric 10-color configuration at seven clinical sites using delinked and de-identified blood specimens from HIV-infected and uninfected subjects to enumerate T-, B-, and NK-lymphocytes with the BD Multitest™ reagents (BD Multitest IMK kit and BD Multitest 6-color TBNK). Samples were analyzed on the BD FACSLyric system with BD FACSuite Clinical software, and on the BD FACSCanto™ II system with BD FACSCanto clinical software and BD FACS 7-Color Setup beads. For equivalency between methods, data (n = 362) were analyzed with Deming regression for absolute count and percentage of lymphocytes. Results gave R2 ≥0.98, with slope values ≥0.96, and slope ranges between 0.90-1.05. The percent (%) bias values were <10% for T- and NK cells and <15% for B- cells. The between-site (n = 4) total precision was tested for 5 days (2 runs/day), and gave %coefficient of variation below 10% for absolute cell counts. The stability claims were confirmed (n = 186) for the two BD Multitest reagents. The reference intervals were re-established in male and female adults (n = 134). The analysis by gender showed statistically significant differences for CD3+ and CD4+ T-cell counts and %CD4. In summary, the BD FACSLyric and the BD FACSCanto II systems generated comparable measurements of T-, B-, and NK-cells using BD Multitest assays.Entities:
Mesh:
Substances:
Year: 2019 PMID: 30689658 PMCID: PMC6349332 DOI: 10.1371/journal.pone.0211207
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
T-, B-, and NK cell Deming regression results determined using BD Multitest reagents.
| Lymphocytes | IMK Kit | 6-color TBNK Reagent | |||||
|---|---|---|---|---|---|---|---|
| R2 | Slope [95% CI | Intercept | R2 | Slope [95% CI | Intercept | ||
| 0.99 | 1.04 [1.03, 1.05] | 1.04 | |||||
| 0.99 | 1.03 [1.02, 1.05] | 3.18 | 0.99 | 1.04 [1.03, 1.04] | -0.97 | ||
| 0.99 | 1.04 [1.03, 1.05] | -0.62 | |||||
| 0.99 | 1.02 [1.01, 1.04] | -0.05 | 1.00 | 1.04 [1.02, 1.05] | -0.79 | ||
| 0.99 | 1.02 [1.00, 1.04] | -1.35 | 0.99 | 1.03 [1.02, 1.05] | 0.23 | ||
| 0.99 | 0.96 [0.94, 0.98] | -3.79 | 1.00 | 0.94 [0.90, 0.99] | -0.23 | ||
| 1.00 | 1.02 [1.01, 1.04] | -0.05 | 1.00 | 1.01 [1.00, 1.03] | -0.21 | ||
| 0.99 | 1.00 [0.99, 1.01] | 0.60 | |||||
| 0.99 | 1.00 [0.99, 1.01] | 0.68 | 0.99 | 1.00 [0.98, 1.01] | 1.43 | ||
| 0.99 | 1.00 [0.99, 1.01] | 0.21 | |||||
| 1.00 | 1.01 [1.00, 1.02] | -.0.22 | 1.00 | 0.99 [0.99, 1.02] | 0.34 | ||
| 0.99 | 1.00 [0.98, 1.01 ] | -0.08 | 1.00 | 1.01 [1.00, 1.02] | 0.09 | ||
| 0.99 | 0.99 [0.98, 1.01] | -0.81 | 0.99 | 1.00 [0.97, 1.03] | -0.73 | ||
| 1.00 | 1.02 [1.01, 1.03] | 1.02 | 1.00 | 1.02 [1.01, 1.04] | -0.29 | ||
aWeighted Deming regression
bregular Deming regression
c95%confidence interval.
Fig 2Deming regression and Bland-Altman plots of the absolute counts of T-, B-, and NK lymphocytes identified with the BD Multitest 6-Color TBNK reagent.
This is the Fig 2 legend. The T- cells (CD3+, CD3+CD4+ and CD3+CD8+) shown as A, B, and C; the NK lymphocytes in D, and the B-lymphocytes are depicted in E stained with BD Multitest 6-Color TBNK reagent. The Deming regression are the upper plots with identity line (grey) and the Weighted Deming regression line (red) of each subset. The x-axis depicts BD FACSCanto II cell counts and the y-axis shows the BD FACSLyric counts. The Bland-Altman graphs are the lower plots of each subset, with the bias line (blue solid line) and the upper and lower limits of agreement (blue dotted lines). For these plots, the x-axis corresponds to the average of the lymphocyte marker, and the y-axis represents the estimated percent difference of the absolute counts between systems.
Agreement at the CD4 clinical cutoff of 200 cells/μL.
| Reagent | Systems | FACSCanto II (Cells/μL) | Agreement | Percent (%) | 95% CL (LCL | ||
|---|---|---|---|---|---|---|---|
| BD FACSLyric (Cells/μL) | Positive | Negative | Total | ||||
| Positive (≤200) | 85 | 2 | 87 | 98.9 | 97.2, 99.6 | ||
| Negative (>200) | 2 | 273 | 275 | 97.7 | 92.0, 99.4 | ||
| Total | 87 | 275 | 362 | 99.3 | 97.4, 99.8 | ||
| Positive (≤200) | 71 | 0 | 71 | 99.1 | 97.3, 99.7 | ||
| Negative (>200) | 3 | 243 | 246 | 96.0 | 88.8, 98.6 | ||
| Total | 74 | 243 | 317 | 100 | 98.4, 100.0 | ||
lower confidence limit
eupper confidence limit.
Predicted bias interval at the CD4 clinical cutoff.
| BD Multitest assay | CD4 cutoff | %Bias (%) | SE | 95% CI |
|---|---|---|---|---|
| 200 | 2.55 | 0.49 | 1.59, 3.51 | |
| 3.39 | 0.59 | 2.23, 4.54 |
cconfidence interval.
Inter-laboratory total precision lymphocyte subset results with BD Multitest 6-color TBNK reagent.
| Lymphocyte | CD4 Low | CD4 Normal | |||
|---|---|---|---|---|---|
| Mean | %CV | Mean | %CV | ||
| 875.81 | 4.86 | 1742.39 | 5.48 | ||
| 185.79 | 7.28 | 1175.19 | 5.95 | ||
| 616.88 | 5.14 | 557.86 | 8.32 | ||
| 294.49 | 6.82 | 231.03 | 9.48 | ||
| 335.03 | 6.37 | 276.52 | 7.72 | ||
| 57.46 | 1.14 | 76.98 | 1.02 | ||
| 12.19 | 0.73 | 51.91 | 1.09 | ||
| 40.47 | 1.05 | 24.64 | 1.59 | ||
| 19.31 | 0.84 | 10.20 | 0.75 | ||
| 21.97 | 0.84 | 12.21 | 0.65 | ||
fPercent coefficient of variation
gstandard deviation.
Fig 3Inter-laboratory total precision per site.
The variability per site is illustrated with the BD Multitest 6-color TBNK reagent total precision. The total precision results using CD Chex Plus control materials for the CD4 low are shown on the upper plot, and the total precision for CD4 normal on the bottom plot. The x-axis shows the participating laboratoriers per number and color. The y-axis displays the percentage of coefficient of variance (%CV) and the upper CV limit.
Specimen and stained sample storage biases at different time points using BD Multitest reagents.
| Time point | Cell Type | Mean %Bias | LCL | UCL |
|---|---|---|---|---|
| -0.70 | -1.45 | 0.06 | ||
| -0.26 | -1.19 | 0.67 | ||
| -1.01 | -1.87 | -0.15 | ||
| -0.31 | -1.54 | 0.92 | ||
| 0.42 | -0.60 | 1.44 | ||
| -6.54 | -8.61 | -4.47 | ||
| 2.70 | 0.29 | 5.11 | ||
| -3.46 | -4.27 | -2.66 | ||
| -3.51 | -4.44 | -2.57 | ||
| -3.31 | -4.28 | -2.35 | ||
| -2.94 | -4.19 | -1.69 | ||
| -2.75 | -3.73 | -1.76 | ||
| -11.15 | -13.42 | -8.87 | ||
| -2.29 | -4.66 | 0.08 | ||
| -4.10 | -4.96 | -3.24 | ||
| -4.11 | -5.09 | -3.12 | ||
| -4.00 | -5.03 | -2.97 | ||
| -2.62 | -4.54 | -0.70 | ||
| -3.28 | -4.47 | -2.10 | ||
| -10.02 | -12.94 | -7.11 | ||
| -0.48 | -7.52 | 6.57 | ||
| 0.02 | -0.95 | 0.99 | ||
| -0.55 | -1.92 | 0.81 | ||
| 0.59 | -0.49 | 1.66 | ||
| -5.39 | -7.51 | -3.28 | ||
| -0.31 | -2.85 | 2.23 | ||
| 0.36 | -0.71 | 1.43 | ||
| -0.33 | -1.66 | 0.99 | ||
| 1.05 | -0.13 | 2.23 | ||
| -5.91 | -8.05 | -3.77 | ||
| 0.62 | -1.81 | 3.06 |
hlower confidence limit
iupper confidence limit.
Lymphocyte subset reference intervals for the BD Multitest reagents.
| Reagent and T-, B-, and NK Lymphocytes | Reference Interval | 90% CI | 90% CI | ||||
|---|---|---|---|---|---|---|---|
| Percentile 2.5 | Percentile 97.5 | Lower CI | Upper CI | Lower CI | Upper CI | ||
| 827 | 2547 | 613 | 875 | 2513 | 3357 | ||
| 840 | 2641 | 620 | 860 | 2413 | 3353 | ||
| 812 | 2655 | 606 | 858 | 2412 | 3361 | ||
| 488 | 1711 | 407 | 549 | 1588 | 1835 | ||
| 154 | 1097 | 115 | 185 | 981 | 2141 | ||
| 102 | 617 | 81 | 106 | 603 | 1316 | ||
| 60 | 551 | 21 | 108 | 509 | 600 | ||
| 56.86 | 82.5 | 52.43 | 58.02 | 82.13 | 89.44 | ||
| 56.65 | 83.36 | 53.28 | 57.96 | 81.86 | 89.96 | ||
| 56.74 | 82.54 | 51.58 | 58.07 | 81.64 | 88.93 | ||
| 32.42 | 63.19 | 26.95 | 32.8 | 61.16 | 65.66 | ||
| 8.99 | 38.99 | 6.82 | 11.94 | 37.11 | 52.31 | ||
| 5.42 | 29.65 | 3.92 | 5.63 | 24.95 | 35.96 | ||
| 5.14 | 22.96 | 1.24 | 6.79 | 21.6 | 24.57 | ||
| 856 | 2669 | 607 | 894 | 2342 | 3090 | ||
| 491 | 1734 | 363 | 543 | 1636 | 1834 | ||
| 162 | 1074 | 111 | 178 | 951 | 1946 | ||
| 108 | 680 | 83 | 117 | 583 | 1191 | ||
| 73 | 562 | 32 | 102 | 536 | 757 | ||
| 57.52 | 83.11 | 51.31 | 58.9 | 82.42 | 87.15 | ||
| 31.45 | 62.38 | 26.4 | 33.23 | 61.4 | 66.86 | ||
| 9.55 | 38.32 | 6.65 | 11.27 | 36.33 | 51.7 | ||
| 5.17 | 30.36 | 4.51 | 5.82 | 27.16 | 35.08 | ||
| 5.89 | 24.21 | 2.24 | 6.92 | 22.73 | 29.45 | ||
90% confidence interval
confidence interval.