Correlation Between Average Segment Width and Gamma Passing Rate as a Function of MLC Position Error in Volumetric Modulated Arc Therapy.

Young Min Moon1, Sang Il Bae1, Moo Jae Han2, Wan Jeon1, Tosol Yu1, Chul Won Choi1, Jin Young Kim1.   

Abstract

OBJECTIVE: This study analyzed the correlation between the average segment width (ASW) and gamma passing rate according to the multi-leaf collimator (MLC) position error.
METHOD: To evaluate the changes in the gamma passing rate according to the MLC position error, 21 volumetric modulated arc therapy (VMAT) plans were generated using pelvic lymph node metastatic prostate cancer patient's data which is sensitive to MLC position errors as they involve several long, narrow, irregular fields. The ASW for each VMAT plan was calculated using our own code developed using Visual Basic for Applications (VBA). The gamma passing rate of the VMAT plan according to the MLC position error was evaluated using ArcCHECK (Sun Nuclear, Melbourne, FL, USA) while inducing symmetric MLC position errors in 0.25 mm intervals from -1 mm to +1 mm in the infinity medical linear accelerator (Elekta AB, Stockholm, Sweden). Finally, we examined the correlation between the change in the passing rate (γgradient) due to the MLC position error and the ASW in VMAT through linear regression analysis using the least squares method.
RESULTS: The ASW and γgradient were found to have a linear correlation according to the MLC position error, and the coefficient of determination was 0.88. For a 1 mm position error of MLC in VMAT, the gamma passing rate improved by approximately 11.9% as the ASW increased by 10 mm.
CONCLUSION: These results are expected to be employed as guidelines to minimize the dose uncertainty due to MLC position error in VMAT.

Entities:  

Keywords:  MLC position error; VMAT; average segment width; gamma passing rate; radiation therapy

Mesh:

Year:  2021        PMID: 34821195      PMCID: PMC8649092          DOI: 10.1177/15330338211059937

Source DB:  PubMed          Journal:  Technol Cancer Res Treat        ISSN: 1533-0338


  11 in total

Review 1.  The physical basis of IMRT and inverse planning.

Authors:  S Webb
Journal:  Br J Radiol       Date:  2003-10       Impact factor: 3.039

2.  A comparison of the gamma index analysis in various commercial IMRT/VMAT QA systems.

Authors:  Mohammad Hussein; Pejman Rowshanfarzad; Martin A Ebert; Andrew Nisbet; Catharine H Clark
Journal:  Radiother Oncol       Date:  2013-10-04       Impact factor: 6.280

3.  Tolerances on MLC leaf position accuracy for IMRT delivery with a dynamic MLC.

Authors:  Alejandra Rangel; Peter Dunscombe
Journal:  Med Phys       Date:  2009-07       Impact factor: 4.071

4.  Dosimetric impact of systematic MLC positional errors on step and shoot IMRT for prostate cancer: a planning study.

Authors:  N M Ung; C S Harper; L Wee
Journal:  Australas Phys Eng Sci Med       Date:  2011-03-16       Impact factor: 1.430

5.  On the sensitivity of common gamma-index evaluation methods to MLC misalignments in Rapidarc quality assurance.

Authors:  G Heilemann; B Poppe; W Laub
Journal:  Med Phys       Date:  2013-03       Impact factor: 4.071

Review 6.  Influence of segment width on plan quality for volumetric modulated arc based stereotactic body radiotherapy.

Authors:  Karthikeyan Nithiyanantham; Ganesh Kadirampatti Mani; Vikraman Subramani; Karrthick Karukkupalayam Palaniappan; Mohanraj Uthiran; Sennniandavar Vellengiri; Sambasivaselli Raju; Sanjay S Supe; Tejinder Kataria
Journal:  Rep Pract Oncol Radiother       Date:  2014-04-16

7.  Clinical significance of multi-leaf collimator positional errors for volumetric modulated arc therapy.

Authors:  Mike Oliver; Isabelle Gagne; Karl Bush; Sergei Zavgorodni; Will Ansbacher; Wayne Beckham
Journal:  Radiother Oncol       Date:  2010-12       Impact factor: 6.280

8.  Moving from gamma passing rates to patient DVH-based QA metrics in pretreatment dose QA.

Authors:  Heming Zhen; Benjamin E Nelms; Wolfgang A Tome
Journal:  Med Phys       Date:  2011-10       Impact factor: 4.071

9.  Analysis of direct clinical consequences of MLC positional errors in volumetric-modulated arc therapy using 3D dosimetry system.

Authors:  Karthikeyan Nithiyanantham; Ganesh K Mani; Vikraman Subramani; Lutz Mueller; Karrthick K Palaniappan; Tejinder Kataria
Journal:  J Appl Clin Med Phys       Date:  2015-09-08       Impact factor: 2.102

10.  A study of minimum segment width parameter on VMAT plan quality, delivery accuracy, and efficiency for cervical cancer using Monaco TPS.

Authors:  Yuanyuan Wang; Li Chen; Fengying Zhu; Wanjing Guo; Dandan Zhang; Wenzhao Sun
Journal:  J Appl Clin Med Phys       Date:  2018-07-30       Impact factor: 2.102

View more
  1 in total

1.  Sensitivity of Three Patient-Specific Quality Assurance Systems to MLC Aperture Errors With Volumetric Modulated Arc Therapy.

Authors:  Jinyou Hu; Shaoxian Gu; Ningyu Wang; Fengjie Cui; Shengyuan Zhang; Chuou Yin; Yunzhu Cai; Chengjun Gou; Lian Zou; Zhangwen Wu
Journal:  Technol Cancer Res Treat       Date:  2022 Jan-Dec