Moving from gamma passing rates to patient DVH-based QA metrics in pretreatment dose QA.

Heming Zhen1, Benjamin E Nelms, Wolfgang A Tome.   

Abstract

PURPOSE: The purpose of this work is to explore the usefulness of the gamma passing rate metric for per-patient, pretreatment dose QA and to validate a novel patient-dose∕DVH-based method and its accuracy and correlation. Specifically, correlations between: (1) gamma passing rates for three 3D dosimeter detector geometries vs clinically relevant patient DVH-based metrics; (2) Gamma passing rates of whole patient dose grids vs DVH-based metrics, (3) gamma passing rates filtered by region of interest (ROI) vs DVH-based metrics, and (4) the capability of a novel software algorithm that estimates corrected patient Dose-DVH based on conventional phantom QA data are analyzed.
METHODS: Ninety six unique "imperfect" step-and-shoot IMRT plans were generated by applying four different types of errors on 24 clinical Head∕Neck patients. The 3D patient doses as well as the dose to a cylindrical QA phantom were then recalculated using an error-free beam model to serve as a simulated measurement for comparison. Resulting deviations to the planned vs simulated measured DVH-based metrics were generated, as were gamma passing rates for a variety of difference∕distance criteria covering: dose-in-phantom comparisons and dose-in-patient comparisons, with the in-patient results calculated both over the whole grid and per-ROI volume. Finally, patient dose and DVH were predicted using the conventional per-beam planar data as input into a commercial "planned dose perturbation" (PDP) algorithm, and the results of these predicted DVH-based metrics were compared to the known values.
RESULTS: A range of weak to moderate correlations were found between clinically relevant patient DVH metrics (CTV-D95, parotid D(mean), spinal cord D1cc, and larynx D(mean)) and both 3D detector and 3D patient gamma passing rate (3%∕3 mm, 2%∕2 mm) for dose-in-phantom along with dose-in-patient for both whole patient volume and filtered per-ROI. There was considerable scatter in the gamma passing rate vs DVH-based metric curves. However, for the same input data, the PDP estimates were in agreement with actual patient DVH results.
CONCLUSIONS: Gamma passing rate, even if calculated based on patient dose grids, has generally weak correlation to critical patient DVH errors. However, the PDP algorithm was shown to accurately predict the DVH impact using conventional planar QA results. Using patient-DVH-based metrics IMRT QA allows per-patient dose QA to be based on metrics that are both sensitive and specific. Further studies are now required to analyze new processes and action levels associated with DVH-based metrics to ensure effectiveness and practicality in the clinical setting.

Entities:  

Mesh:

Year:  2011        PMID: 21992366     DOI: 10.1118/1.3633904

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  81 in total

1.  Correlation between gamma index passing rate and clinical dosimetric difference for pre-treatment 2D and 3D volumetric modulated arc therapy dosimetric verification.

Authors:  X Jin; H Yan; C Han; Y Zhou; J Yi; C Xie
Journal:  Br J Radiol       Date:  2014-12-10       Impact factor: 3.039

2.  Correcting lateral response artifacts from flatbed scanners for radiochromic film dosimetry.

Authors:  David Lewis; Maria F Chan
Journal:  Med Phys       Date:  2015-01       Impact factor: 4.071

3.  Independent calculation of monitor units for VMAT and SPORT.

Authors:  Xin Chen; Karl Bush; Aiping Ding; Lei Xing
Journal:  Med Phys       Date:  2015-02       Impact factor: 4.071

Review 4.  Complexity metrics for IMRT and VMAT plans: a review of current literature and applications.

Authors:  Sophie Chiavassa; Igor Bessieres; Magali Edouard; Michel Mathot; Alexandra Moignier
Journal:  Br J Radiol       Date:  2019-07-24       Impact factor: 3.039

5.  Comparison of global and local gamma evaluation results using isodose levels.

Authors:  Liting Yu; Tanya Kairn; Jamie V Trapp; Scott B Crowe
Journal:  Phys Eng Sci Med       Date:  2021-02-08

6.  Three-dimensional gamma analysis of dose distributions in individual structures for IMRT dose verification.

Authors:  Yuuki Tomiyama; Fujio Araki; Takeshi Oono; Kazunari Hioki
Journal:  Radiol Phys Technol       Date:  2014-05-06

7.  Assessing the shift of radiobiological metrics in lung radiotherapy plans using 2D gamma index.

Authors:  Abdulhamid Chaikh; Jacques Balosso
Journal:  Transl Lung Cancer Res       Date:  2016-06

8.  A method for quantitative evaluations of scanning-proton dose distributions.

Authors:  Bryce C Allred; Jie Shan; Daniel G Robertson; Todd A DeWees; Jiajian Shen; Wei Liu; Joshua B Stoker
Journal:  J Appl Clin Med Phys       Date:  2021-03-29       Impact factor: 2.102

9.  Study of efficiency in five-field and field-by-field intensity modulated radiation therapy (IMRT) plan using DOSXYZnrc Monte Carlo code.

Authors:  Sitti Yani; Indra Budiansah; Mohamad Fahdillah Rhani; Freddy Haryanto
Journal:  Rep Pract Oncol Radiother       Date:  2020-04-27

10.  Verification of the machine delivery parameters of a treatment plan via deep learning.

Authors:  Jiawei Fan; Lei Xing; Ming Ma; Weigang Hu; Yong Yang
Journal:  Phys Med Biol       Date:  2020-09-30       Impact factor: 3.609

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.