Clinical significance of multi-leaf collimator positional errors for volumetric modulated arc therapy.

Mike Oliver1, Isabelle Gagne, Karl Bush, Sergei Zavgorodni, Will Ansbacher, Wayne Beckham.   

Abstract

BACKGROUND AND
PURPOSE: Multi-leaf collimator (MLC) positional errors occur during intensity modulated radiation therapy (IMRT) and volumetric modulated arc therapy (VMAT) deliveries. The impact of such errors has been evaluated for IMRT but not VMAT. The purpose of this work is to understand how random and systematic VMAT MLC positional errors affect the patient dose distribution.
MATERIALS AND METHODS: Eight head and neck single arc (360°) VMAT treatment plans were created. Random and two types of systematic MLC errors were simulated for error magnitudes of 0.25, 0.5, 1, 2 and 5mm. The two types of systematic MLC errors were: (1) MLC banks are shifted in the same direction (left or right) and (2) MLC banks are shifted in opposing directions resulting in smaller or larger field shapes. The MLC errors were simulated, for all control points, on both banks of active MLC leaves only.
RESULTS: There is a linear correlation of MLC errors with gEUD for all error types. The gEUD dose sensitivities with MLC error for the PTV70 were -0.2, -0.9, -2.8 and 1.9 Gy/mm for random, systematic shift, systematic close and systematic open MLC errors, respectively. The sensitivity of VMAT plans to MLC positional errors was similar to those of IMRT plans with less than 50 segments but much less than those created for a step and shoot with more than 50 segments or sliding-window delivery technique. To maintain the PTV70 to within 2% would require that MLC open/close errors be within 0.6mm.
CONCLUSIONS: Radiation therapy centers should have adequate quality assurance programs in place to assess open/close MLC errors (i.e. leaf gap errors) as they tend to be more impactful than random or systematic MLC shift errors.
Copyright © 2010 Elsevier Ireland Ltd. All rights reserved.

Entities:  

Mesh:

Year:  2010        PMID: 20817291     DOI: 10.1016/j.radonc.2010.06.013

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  32 in total

1.  FusionArc optimization: a hybrid volumetric modulated arc therapy (VMAT) and intensity modulated radiation therapy (IMRT) planning strategy.

Authors:  Martha M Matuszak; Jennifer M Steers; Troy Long; Daniel L McShan; Benedick A Fraass; H Edwin Romeijn; Randall K Ten Haken
Journal:  Med Phys       Date:  2013-07       Impact factor: 4.071

2.  Penalization of aperture complexity in inversely planned volumetric modulated arc therapy.

Authors:  Kelly C Younge; Martha M Matuszak; Jean M Moran; Daniel L McShan; Benedick A Fraass; Donald A Roberts
Journal:  Med Phys       Date:  2012-11       Impact factor: 4.071

3.  Commissioning and Acceptance Testing of the existing linear accelerator upgraded to volumetric modulated arc therapy.

Authors:  Ekambaram Varadharajan; Velayudham Ramasubramanian
Journal:  Rep Pract Oncol Radiother       Date:  2013-08-06

4.  Evaluation of beam matching accuracy among six linacs from the same vendor.

Authors:  Chockkalingam Krishnappan; Chandrasekaran Anu Radha; Karunakaran Balaji; Prasanna Kumar Mani; Vendhan Subramani; Velmurugan Thanigaimalai; Madhan Kumar Gunasekaran; Velayudham Ramasubramanian
Journal:  Radiol Phys Technol       Date:  2018-09-29

5.  Optical imaging provides rapid verification of static small beams, radiosurgery, and VMAT plans with millimeter resolution.

Authors:  Muhammad Ramish Ashraf; Petr Bruza; Brian W Pogue; Nathan Nelson; Benjamin B Williams; Lesley A Jarvis; David J Gladstone
Journal:  Med Phys       Date:  2019-10-04       Impact factor: 4.071

6.  Correlation Between Average Segment Width and Gamma Passing Rate as a Function of MLC Position Error in Volumetric Modulated Arc Therapy.

Authors:  Young Min Moon; Sang Il Bae; Moo Jae Han; Wan Jeon; Tosol Yu; Chul Won Choi; Jin Young Kim
Journal:  Technol Cancer Res Treat       Date:  2021 Jan-Dec

7.  Optimizing the Region for Evaluation of Global Gamma Analysis for Nasopharyngeal Cancer (NPC) Pretreatment IMRT QA by COMPASS: A Retrospective Study.

Authors:  Wenli Lu; Ying Li; Wei Huang; Haixia Cui; Hanyin Zhang; Xin Yi
Journal:  Front Oncol       Date:  2022-06-14       Impact factor: 5.738

8.  Is RapidArc more susceptible to delivery uncertainties than dynamic IMRT?

Authors:  Gregory T Betzel; Byong Yong Yi; Ying Niu; Cedric X Yu
Journal:  Med Phys       Date:  2012-10       Impact factor: 4.506

9.  The sensitivity of gamma-index method to the positioning errors of high-definition MLC in patient-specific VMAT QA for SBRT.

Authors:  Jung-In Kim; So-Yeon Park; Hak Jae Kim; Jin Ho Kim; Sung-Joon Ye; Jong Min Park
Journal:  Radiat Oncol       Date:  2014-07-28       Impact factor: 3.481

10.  Evaluation of dosimetric effect caused by slowing with multi-leaf collimator (MLC) leaves for volumetric modulated arc therapy (VMAT).

Authors:  Zhengzheng Xu; Iris Z Wang; Lalith K Kumaraswamy; Matthew B Podgorsak
Journal:  Radiol Oncol       Date:  2016-02-16       Impact factor: 2.991

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.