Literature DB >> 34742331

Fixation failure in patients with traumatic diastasis of pubic symphysis: impact of loss of reduction on early functional outcomes.

Claudio Rojas1,2, Ernesto Ewertz3, Jose Miguel Hormazábal1.   

Abstract

BACKGROUND: Failure of fixation (FF) in pubic symphysis diastasis (SD) ranges between 12 and 75%, though whether it influences functional outcomes is still debated. The objective of this study is to evaluate the impact of anterior pelvic plate failure and loss of reduction on Majeed's functional scores.
METHODS: Single center retrospective review of consecutive patients with acute SD treated by means of anterior pubic plating. Thirty-seven patients with a mean age 45.7 ± 14.4 years were included. Demographics, AO classification, pelvic fixation and secondary procedures were recorded. Majeed's functional scores at minimum 6 months follow-up were compared according to the presence of FF and loss of reduction.
RESULTS: Fifteen patients presented FF. Eight presented an additional loss of symphyseal reduction. Mean Majeed´s score (MMS) in patients with and without FF was 64.4 ± 13.04 and 81.8 ± 15.65, respectively (p = 0.0012). Differences in MMS in patients without FF and those with FF and maintained or loss of anterior reduction were 11.3 [70.5 vs 81.8] (p = 0.092) and 22.7 [59.1 vs 81.8] (p = 0.001), respectively. Significant association of FF with AO classification was noted. (OR 12.6; p = 0.002).
CONCLUSIONS: Differences in MMS in the analyzed groups suggest that loss of reduction might be more relevant than failure of the anterior osteosynthesis in functional outcomes.
© 2021. The Author(s).

Entities:  

Keywords:  APC pelvic fracture; Fixation failure; Functional outcomes; Loss of reduction; Majeed’s; Pubic diastasis; Symphyseal disruption

Mesh:

Year:  2021        PMID: 34742331      PMCID: PMC8572449          DOI: 10.1186/s13018-021-02802-x

Source DB:  PubMed          Journal:  J Orthop Surg Res        ISSN: 1749-799X            Impact factor:   2.359


Introduction

The pubic symphysis is a fibrocartilaginous joint composed of four strong ligaments that allows discrete multidirectional movement [1]. High energy mechanisms are required to injure these ligaments, and damage to surrounding structures can result in relatively high rates of hemodynamic instability and mortality in the acute setting [2-5]. Diverse fixation methods have been used to restore and maintain symphyseal reduction, though the most frequently utilized are external fixators and pubic plates [6-10]. As ligamentous structures heal and normal joint motion resumes, rigid fixation methods can result in metalwork loosening, plate breakage or loss of anatomic reduction. Failure rates of pubic plates ranging from 12 to 75% and significant loss of symphyseal reduction in 7 to 88% have been previously reported [11-15]. Furthermore, early failures within the first 30 days after surgery can present in as many as 11.6% of APC pelvic fractures [15]. Despite this, only 1.6 to 9% of patients will eventually require secondary surgical procedures [12, 14–19]. Therefore, whether plate loosening, breakage or loss of symphyseal reduction have an impact on functional outcomes is still a matter of debate [11, 20–22]. Given the paucity of literature regarding the relevance of anterior metalwork failure and loss of reduction, we sought to determine their influence in clinical outcomes. The objective of this study is to evaluate the impact of anterior pelvic plate failure and loss of reduction on early Majeed’s functional scores.

Methods

We conducted a retrospective cohort review at a single level I trauma center, of consecutive patients admitted with the diagnosis of pelvic fracture, between January 2016 and December 2018. Patients with acute symphyseal diastasis that were surgically treated using 4 to 6 hole 4.5 anterior symphyseal plates (Matta Pelvic System®; Stryker, Mahwah, NJ, USA) were included. Patients with incomplete radiological studies or unregistered Majeed functional scores were excluded. Population data regarding demographics, mechanism of injury, associated injuries, 2018 AO pelvic fracture classification [23] and osteosynthesis characteristics were recorded using a predefined data gathering spreadsheet. Patients were followed for a minimum of 6 months. Majeed’s functional scores [24] at 6 months were recorded. Digital radiological images (AGFA Xero viewer 8.1.2 system) after surgery were evaluated by two trained team members. Fixation failure (FF) was defined as presence of plate or screw breakage and screw loosening. Additional loss of symphyseal reduction was defined as an increase ≥ 5 mm in the distance between pubic bones in radiological follow up. Patients were categorized in three groups according to the presence of FF and loss of symphyseal reduction. The groups are as follows: Group A, patients without radiological failure, group B, patients with FF and maintained reduction and Group C, patients with FF and additional loss of symphyseal reduction. Majeed’s functional scores between groups were compared. The rates of FF and severity of pelvic injury were also analyzed. Statistical analysis was performed using Stata Statistical Software (Release 13. College Station, TX: StataCorp LP). Data are presented in means, ranges, and standard deviation. Normal distribution of the population was determined using the Shapiro–Wilk test. Differences were analyzed using t-student and Fisher’s exact test. A 95% confidence interval was used, and statistical significance was set at p value < 0.05 with IC 95%.

Results

After inclusion and exclusion criteria were applied, 37 patients were eligible for final analysis. Thirty male and seven female patients, with a mean age of 45.7 years (SD: ± 14.4, range: 20–77 years), were included. The mechanism of injury was motor vehicle accidents in 23, fall from height in 9 and crushing injuries in 6 patients. Anterior pubic plating fixation was performed as follows. Four hole plate vs six hole 4.5 plates were used in 20 vs 2 patients in group A, 6 vs 1 in group B and 6 vs 2 in group C. Additional sacroiliac screw fixation was performed in thirty-five (94.6%) patients. No signs of FF were observed in 22 patients (Group A). Fifteen patients presented FF; 7 patients maintained postoperative symphysial reduction (Group B) and 8 patients presented an additional loss of the anterior reduction (Group C). No cases of plate breakage were recorded. Two patients presented early failures that required secondary interventions; a catastrophic failure 17 days after the initial surgery that was revised using a double plating fixation and one deep wound infection that required surgical debridement, metalwork removal and revised to an external fixator as definitive surgery. No late secondary surgeries were recorded. Immediate postoperative symphyseal distance in groups A, B and C was 6.2 mm (SD 1.87, range 4–10), 7.71 mm (DE 2.13, range 4–11) and 7 mm (DE 2.61, range 2–11), respectively. After failure, group C presented a mean distance of 15.5 mm (DE 2.82, range 11–20). Mean time from surgery to loss of reduction was 5 ± 2.7 weeks (range, 1–8). Detailed information is presented in Table 1.
Table 1

Details of patients with failure of pubic fixation (Groups B + C)

AgeSexAO ClassificationMajeed scoreAnterior fixationPosterior Fixation
Postoperative diastasis (mm)Loss of reduction (mm)Time to loss of reduction (weeks)Nº LevelsType (sacroiliac screws)
57MB2.2521111si
63MB2.368911 si
55MB3.362812 si
28FB3.384411 si
51MB3.371711 si*
49MC184812 tsti
51MC173722 si 1 tsti
61MB2.258818811 si
40MB2.3645118
53MB2.364814811 si
60MB3.343814411 si 1 tsti
67MC1441116112 tsti*
55MC151714423 si 1 tsti
41FC178720511 tsti
20FC271817213 si

* = secondary interventions

si = sacroiliac screw /

tsi Transiliac-Transsacral Screw

Details of patients with failure of pubic fixation (Groups B + C) * = secondary interventions si = sacroiliac screw / tsi Transiliac-Transsacral Screw The mean Majeed score (MMS) according to AO classification was 79.48 ± 15.8 and 62.1 ± 12.96 for AO-B and AO-C fractures, respectively (p = 0.019). The MMS in groups A, B and C was 81.8 ± 15.65, 70.5 ± 11.4 and 59.1 ± 12.54, respectively. Combined MMS for Groups B and C (failure group) was 64.4 ± 13.04. Significant differences between Group A and failure group (B + C groups) were observed (p = 0.0012). Failure group was further analyzed according to presence or absence of loss of reduction. No significant differences in MMS when comparing groups A and B were observed. (p = 0.092). However, significant differences were observed in MMS between Group A and C (p = 0.001). Moreover, patients from Group C presented a decreased MMS of 22.7. (Table 2).
Table 2

Mean majeeds functional scores

Group (n) = scoreNo fixation failureFixation failure
MMSA (22) = 81.8B(7) = 70.5p = 0.092
C(8) = 59.1p = 0.001

*MMS Mean Majeed`s Functional Score

p p value

Mean majeeds functional scores *MMS Mean Majeed`s Functional Score p p value Failure rates by severity of pelvic injury were analyzed. Five (20%) B type and 10 (77%) C type fractures presented FF. (OR 12.6; range, 2.4–64.2; p = 0.002) [2].

Discussion

In our series, 15 (41%) patients presented failure of the anterior pelvic fixation, which is similar to other reports in the literature [11]. Additional loss of anterior reduction (7.75 mm ± 2.71) was observed in eight cases (19%) presenting as late as 8 weeks postoperatively. Despite this, only two patients required early secondary interventions and no patients required late plate removal due to symptoms associated with loosening. Even though no patient-reported outcome measures have been developed for pelvic trauma; generic outcome instruments (SF-36) and disease specific instruments have been used to assess patient functional performance [25]. The Majeed score has been widely used in research concerning the quality of life of patients with pelvic injuries [24]. We investigated how implant failure and loss of reduction influence the patient’s functional scores. The impact of FF has been previously studied. Giannoudis et al. found no correlation of its occurrence with worse functional outcome [14]. On the other hand, Frietman et al. [25] retrospectively evaluated 37 patients treated with anterior pelvic plating and reported higher Majeed scores and similar SF-36 scores in patients with implant failure. Conversely, we observed significant differences in mean Majeed scores of patients with and without FF (p = 0.012). Moreover, additional loss of anterior reduction resulted in significant differences in mean Majeed Score (p = 0.001); meanwhile, no differences were observed in cases with failure and maintained reduction (p = 0.092). These findings suggest that loss of anatomic reduction might be more relevant than loosening or failure of the anterior construct. In this series, patients with FF were 12.6 times more likely to belong to the C type fracture group. Posterior fixation has been proposed as a method of augmentation in fractures with some component of posterior ligamentous injury in order to reduce anterior implant failure [26]. Currently, B3 fractures and C type fractures are managed with posterior fixation, most frequently using sacroiliac screws, though failure rates remain high [3, 4, 15, 27, 28]. This study has many limitations arising from its retrospective nature, and functional outcomes scores were not recorded at the same point in time. Also, the influence of associated injuries on Majeed scores was not evaluated. The studied population is composed entirely of workers entitled to compensation benefits, making it difficult to compare with other patient groups. Even though the severity of pelvic fracture and associated injuries are the main determinants of functional outcomes in patients with traumatic pubic diastasis; loss of anatomic reduction does result in poorer outcomes. The rates of FF and loss of reduction are higher in pelvic fractures with significant injury to posterior structures (AO B3—C type fractures) and posterior arch stabilization should be performed to address them, though they do not prevent failure of the anterior fixation. Augmentation of the anterior construct by means of double plating, cables, tight rope supplementation or the use of larger plates should be considered in order to reduce failure rates, loss of reduction and improve the functional outcome in this group of patients.
  28 in total

Review 1.  Implant retention and removal after internal fixation of the symphysis pubis.

Authors:  Raghu Raman; Craig S Roberts; Hans-Christoph Pape; Peter V Giannoudis
Journal:  Injury       Date:  2005-02-25       Impact factor: 2.586

Review 2.  Biomechanics of pelvic fixation.

Authors:  P T Simonian; M L Routt
Journal:  Orthop Clin North Am       Date:  1997-07       Impact factor: 2.472

3.  Indications for anterior fixation of pelvic fractures.

Authors:  J M Matta
Journal:  Clin Orthop Relat Res       Date:  1996-08       Impact factor: 4.176

4.  Biomechanical simulation of the anteroposterior compression injury of the pelvis. An understanding of instability and fixation.

Authors:  P T Simonian; M L Routt; R M Harrington; K A Mayo; A F Tencer
Journal:  Clin Orthop Relat Res       Date:  1994-12       Impact factor: 4.176

5.  Mobility of the pubic symphysis. Measurements by an electromechanical method.

Authors:  G Walheim; S Olerud; T Ribbe
Journal:  Acta Orthop Scand       Date:  1984-04

6.  Pelvic ring disruptions with symphysis pubis diastasis. Indications, technique, and limitations of anterior internal fixation.

Authors:  R H Lange; S T Hansen
Journal:  Clin Orthop Relat Res       Date:  1985-12       Impact factor: 4.176

Review 7.  The adult human pubic symphysis: a systematic review.

Authors:  Ines Becker; Stephanie J Woodley; Mark D Stringer
Journal:  J Anat       Date:  2010-09-14       Impact factor: 2.610

8.  Does Posterior Fixation of Partially Unstable Open-Book Pelvic Ring Injuries Decrease Symphyseal Plate Failure? A Biomechanical Study.

Authors:  Rachelle M Metz; J Gary Bledsoe; Berton R Moed
Journal:  J Orthop Trauma       Date:  2018-02       Impact factor: 2.512

9.  Posterior Fixation of APC-2 Pelvic Ring Injuries Decreases Rates of Anterior Plate Failure and Malunion.

Authors:  Frank R Avilucea; Paul S Whiting; Hassan Mir
Journal:  J Bone Joint Surg Am       Date:  2016-06-01       Impact factor: 5.284

10.  Early failure of symphysis pubis plating.

Authors:  Jonathan G Eastman; James C Krieg; Milton L Chip Routt
Journal:  Injury       Date:  2016-05-17       Impact factor: 2.586

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.