Literature DB >> 27282685

Early failure of symphysis pubis plating.

Jonathan G Eastman1, James C Krieg2, Milton L Chip Routt3.   

Abstract

INTRODUCTION: Operative fixation of a disrupted symphysis pubis helps return alignment and stability to a traumatized pelvic ring. Implant loosening or failure has been demonstrated to commonly occur at some subacute point during the postoperative period. The purpose of this study is to report on a series of patients with traumatic pelvic ring disruptions to determine the incidence and common factors associated with early postoperative symphyseal plate failure before 7 weeks.
MATERIALS AND METHODS: 126 patients retrospectively identified with unstable pelvic injuries treated with open reduction and plate fixation of the symphysis pubis and iliosacral screw fixation. Preoperative and postoperative radiographs, computed tomography (CT) images, and medical chart were reviewed to determine symphyseal displacement preoperatively and postoperatively, time until plate failure, patient symptoms and symphyseal displacement at failure, subsequent symphyseal displacement, incidence of additional surgery, and patient weight bearing compliance.
RESULTS: 14 patients (11.1%) sustained premature postoperative fixation failure. 13 patients had anteroposterior compression (APC)-II injuries and 1 patient had an APC-III injury. Preoperative symphyseal displacement was 35.6 millimeters (mm) (20.8-52.9). Postoperative symphyseal space measurement was 6.3mm (4.7-9.3). Time until plate failure was 29days (5-47). Nine patients (64.2%) noted a pop surrounding the time of failure. Symphyseal space measurement at failure was 12.4mm (5.6-20.5). All patients demonstrated additional symphyseal displacement averaging 2.6mm (0.2-9.4). Two patients (14.2%) underwent revision. Four patients (28.5%) were non-compliant.
CONCLUSION: Premature failure of symphysis pubis plating is not uncommon. In this series, further symphyseal displacement after plate failure was not substantial. The presence of acute symphyseal plate failure alone may not be an absolute indication for revision surgery. Making patient education a priority could lead to decreased postoperative non-compliance and potentially a decreased incidence of implant failure. Posterior pelvic ring fixation aides overall pelvic ring stability and may help minimize further displacement after early postoperative symphyseal plate failure. Further functional outcome and biomechanical studies surrounding early symphyseal plate failure are needed.
Copyright © 2016 Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Early failure; Pelvic implant; Pelvic ring; Symphysis plate

Mesh:

Year:  2016        PMID: 27282685     DOI: 10.1016/j.injury.2016.05.019

Source DB:  PubMed          Journal:  Injury        ISSN: 0020-1383            Impact factor:   2.586


  11 in total

1.  Tape suture constructs for instabilities of the pubic symphysis: is the idea of motion preservation a suitable treatment option? A cadaver study.

Authors:  Adrian Cavalcanti Kußmaul; Fanny Schwaabe; Manuel Kistler; Maximilian Jörgens; Korbinian F Schreyer; Axel Greiner; Wolfgang Böcker; Christopher A Becker
Journal:  Arch Orthop Trauma Surg       Date:  2022-07-13       Impact factor: 2.928

2.  The radiographic outcome after plating for pubic symphysis diastasis: does it matter clinically?

Authors:  Kuo-Yuan Tseng; Kai-Cheng Lin; Shan-Wei Yang
Journal:  Arch Orthop Trauma Surg       Date:  2022-03-12       Impact factor: 3.067

3.  Symphyseal fixation in open book injuries cannot fully compensate anterior SI joint injury-A biomechanical study in a two-leg alternating load model.

Authors:  Fabian M Stuby; Mark Lenz; Stefan Doebele; Yash Agarwal; Hristo Skulev; Björn G Ochs; Jörn Zwingmann; Boyko Gueorguiev
Journal:  PLoS One       Date:  2017-11-27       Impact factor: 3.240

4.  Fixation failure in patients with traumatic diastasis of pubic symphysis: impact of loss of reduction on early functional outcomes.

Authors:  Claudio Rojas; Ernesto Ewertz; Jose Miguel Hormazábal
Journal:  J Orthop Surg Res       Date:  2021-11-06       Impact factor: 2.359

5.  Finite element analysis of the Union Plate in treating elderly acetabular fracture patients.

Authors:  Guixiong Huang; Kaifang Chen; Yulong Wang; Xiaodong Guo
Journal:  J Orthop Surg Res       Date:  2022-01-29       Impact factor: 2.359

6.  How relevant is lumbar bone mineral density for the stability of symphyseal implants? A biomechanical cadaver study.

Authors:  Fanny Schwaabe; Johannes Gleich; Christoph Linhart; Alexander Martin Keppler; Matthias Woiczinski; Christian Kammerlander; Axel Greiner; Wolfgang Böcker; Adrian Cavalcanti Kußmaul
Journal:  Eur J Trauma Emerg Surg       Date:  2021-12-08       Impact factor: 2.374

7.  Trans-obturator cable fixation of open book pelvic injuries.

Authors:  Martin C Jordan; Veronika Jäckle; Sebastian Scheidt; Fabian Gilbert; Stefanie Hölscher-Doht; Süleyman Ergün; Rainer H Meffert; Timo M Heintel
Journal:  Sci Rep       Date:  2021-06-29       Impact factor: 4.379

8.  The role of anterior supra-acetabular external fixator as definitive treatment for anterior ring fixation in unstable pelvic fractures.

Authors:  Cristián Barrientos-Mendoza; Julián Brañes; Rodrigo Wulf; Alex Kremer; Maximiliano Barahona; Sebastián León
Journal:  Eur J Trauma Emerg Surg       Date:  2021-06-07       Impact factor: 2.374

9.  Novel minimally invasive tape suture osteosynthesis for instabilities of the pubic symphysis: a biomechanical study.

Authors:  Adrian Cavalcanti Kußmaul; Fanny Schwaabe; Manuel Kistler; Clara Gennen; Sebastian Andreß; Christopher A Becker; Wolfgang Böcker; Axel Greiner
Journal:  Arch Orthop Trauma Surg       Date:  2021-05-29       Impact factor: 2.928

10.  Plate fixation of the anterior pelvic ring in patients with fragility fractures of the pelvis.

Authors:  Michiel Herteleer; Mehdi Boudissa; Alexander Hofmann; Daniel Wagner; Pol Maria Rommens
Journal:  Eur J Trauma Emerg Surg       Date:  2021-03-11       Impact factor: 2.374

View more