| Literature DB >> 29296040 |
Prabhakar Ramachandran1,2, Abdulrahman Tajaldeen1,2, David Taylor1, Derrick Wanigaratne1, Karl Roozen1, Moshi Geso2.
Abstract
AIM: The aim of this study is to assess the use of ArcCHECK (AC) as an alternative method to replace film dosimetry for pre-treatment quality assurance (QA) of three-dimensional conformal radiation therapy, intensity-modulated radiation therapy (IMRT), and volumetric-modulated arc therapy (VMAT) stereotactic ablative radiotherapy (SABR) treatment plans.Entities:
Keywords: ArcCHECK; EBT3 films; stereotactic ablative radiotherapy
Year: 2017 PMID: 29296040 PMCID: PMC5744454 DOI: 10.4103/jmp.JMP_132_16
Source DB: PubMed Journal: J Med Phys ISSN: 0971-6203
Figure 1Treatment plans of six different SABR cases. (a) Lumber spine/intensity-modulated radiation therapy/20 Gy. (b) Ribs/volumetric modulated arc therapy/20 Gy. (c) Head of femur/three-dimensional conformal radiation therapy/20 Gy. (d) Left lung/three-dimensional conformal radiation therapy/26 Gy. (e) Right lung/three-dimensional conformal radiation therapy/28 Gy. (f) Sternum/three-dimensional conformal radiation therapy/20 Gy
Figure 2(a) ArcCHECK phantom aligned to the isocenter for measurement (b) In house designed Rod phantom with a CC13 chamber inserted to measure absolute dose at reference point
Figure 3(a) Films resized in multiple 2 cm × 2 cm for generating a calibration curve (b) EBT3 film placed in transverse plan (size: 12.5 cm × 20.5 cm) and measured for a lung stereotactic ablative radiotherapy plan (c) Measurement using ArcCHECK versus treatment planning system for a lung stereotactic ablative radiotherapy
Comparison of ArcCHECK and EBT3 results
Figure 4Comparison of passing rates of lung tumors for different gamma criteria
Figure 6Comparison of passing rates of sternum, ribs, scapula, and femur cases for different gamma criteria
Figure 7The relationship between equivalent square field size and passing rates of intensity-modulated radiation therapy plans