Multi-institution validation of a new high spatial resolution diode array for SRS and SBRT plan pretreatment quality assurance.

Mark S Rose1, Lena Tirpak1, Kristin Van Casteren1, Jeff Zack1, Tom Simon1, Andreas Schoenfeld1, William Simon1.   

Abstract

PURPOSE: The SRS MapCHECK® , a recently developed patient-specific quality assurance (PSQA) tool for end-to-end testing of stereotactic radiosurgery (SRS) and stereotactic body radiation therapy (SBRT), was evaluated in a multi-institution study and compared with radiochromic film.
METHODS: The SRS MapCHECK was used to collect data on 84 SBRT or SRS PSQA plans/fields at nine institutions on treatment delivery devices (TDD) manufactured by Varian and Elekta. PSQA plans from five different treatment planning software (TPS) were selected and executed on TDDs operating at beam energies of 6 and 10 MV with and without a flattening filter. The patient plans were all VMAT except for ten conformal arc therapy fields. The plans were selected to encompass a range of size and tumor sites including brain, lung, spine, abdomen, ear, pancreas, and liver. Corresponding radiochromic film data was acquired in 50 plans/fields. Results were evaluated using gamma analysis with absolute dose criterion of 3% global dose-difference (DD) and 1 mm distance-to-agreement (DTA).
RESULTS: The mean 3% DD/1 mm DTA Gamma pass rate of SRS MapCHECK in comparison to film was 95.9%, whereas comparison of SRS MapCHECK to the treatment planning software was 94.7%. 80% of SRS MapCHECK comparisons against film exceed 95% pass rate, and about 30% of SRS MapCHECK comparisons against film exceed 99% pass rate. To maintain good agreement between SRS MapCHECK and film or TPS, authors recommend avoiding plans with a modified modulation complexity score (MMCS) <0.1 arbitrary units (a.u.). In the examples presented, this coincides with avoiding plans with a mu/dose limit of >3 µ/cGy.
CONCLUSIONS: Stereotactic radiosurgery MapCHECK has been validated for PSQA for a variety of clinical SRS/SBRT plans in a wide range of treatment delivery conditions. The SRS MapCHECK comparison with film demonstrates near-equivalence for analysis of patient-specific QA deliveries comprised of small field measurements.
© 2020 American Association of Physicists in Medicine.

Entities:  

Keywords:  End-to-End; QA; SBRT; SRS; SRS MapCHECK; StereoPHAN; patient QA; quality assurance

Mesh:

Year:  2020        PMID: 32215929     DOI: 10.1002/mp.14153

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  10 in total

1.  A patient-specific QA comparison between 2D and 3D diode arrays for single-lesion SRS and SBRT treatments.

Authors:  Yongsook C Lee; Yongbok Kim
Journal:  J Radiosurg SBRT       Date:  2021

2.  The current status and shortcomings of stereotactic radiosurgery.

Authors:  Hunter Mehrens; Trang Nguyen; Sharbacha Edward; Shannon Hartzell; Mallory Glenn; Daniela Branco; Nadia Hernandez; Paola Alvarez; Andrea Molineu; Paige Taylor; Stephen Kry
Journal:  Neurooncol Adv       Date:  2022-04-20

3.  Patient-specific dose quality assurance of single-isocenter multiple brain metastasis stereotactic radiosurgery using PTW Octavius 4D.

Authors:  James McCulloch; Jamie Pawlowski; Neil Kirby; Karl Rasmussen; Zheng Shi; Pamela Myers; Sotirios Stathakis; Niko Papanikolaou; Daniel L Saenz
Journal:  J Appl Clin Med Phys       Date:  2020-07-18       Impact factor: 2.102

4.  Implementing corrections of isocentric shifts for the stereotactic irradiation of cerebral targets: Clinical validation.

Authors:  Sonja Wegener; Robert Schindhelm; Otto A Sauer
Journal:  J Appl Clin Med Phys       Date:  2022-03-02       Impact factor: 2.243

5.  Impact of systematic MLC positional uncertainties on the quality of single-isocenter multi-target VMAT-SRS treatment plans.

Authors:  Georgia Prentou; Eleftherios P Pappas; Eleni Prentou; Nikolaos Yakoumakis; Chryssa Paraskevopoulou; Efi Koutsouveli; Evaggelos Pantelis; Panagiotis Papagiannis; Pantelis Karaiskos
Journal:  J Appl Clin Med Phys       Date:  2022-06-22       Impact factor: 2.243

6.  Development of a high-resolution two-dimensional detector-based dose verification system for tumor-tracking irradiation in the CyberKnife system.

Authors:  Fumitaka Kawabata; Takeshi Kamomae; Kuniyasu Okudaira; Masataka Komori; Hiroshi Oguchi; Motoharu Sasaki; Masaki Mori; Mariko Kawamura; Shinji Abe; Shunichi Ishihara; Shinji Naganawa
Journal:  J Appl Clin Med Phys       Date:  2022-07-05       Impact factor: 2.243

7.  Detection of rotational errors in single-isocenter multiple-target radiosurgery: Is a routine off-axis Winston-Lutz test necessary?

Authors:  Lauren M M Pudsey; Giordano Biasi; Anna Ralston; Anatoly Rosenfeld; Joel Poder
Journal:  J Appl Clin Med Phys       Date:  2022-06-17       Impact factor: 2.243

8.  Comparative study of SRS end-to-end QA processes of a diode array device and an anthropomorphic phantom loaded with GafChromic XD film.

Authors:  Seng Boh Lim; LiCheng Kuo; Tianfang Li; Xiang Li; Ase M Ballangrud; Michael Lovelock; Maria F Chan
Journal:  J Appl Clin Med Phys       Date:  2022-08-10       Impact factor: 2.243

9.  Evaluation of a two-dimensional diode array for patient-specific quality assurance of HyperArc.

Authors:  Richard A Popple; Rodney J Sullivan; Yuan Yuan; Xingen Wu; Elizabeth L Covington
Journal:  J Appl Clin Med Phys       Date:  2021-11-01       Impact factor: 2.102

10.  Implementing and evaluating a high-resolution diode array for patient-specific quality assurance of robotic brain stereotactic radiosurgery/radiotherapy.

Authors:  Qianyi Xu; Kiet Huynh; Wei Nie; Mark S Rose; Ashish K Chawla; Kevin S Choe; Samir Kanani; Gregory J Kubicek; Jiajin Fan
Journal:  J Appl Clin Med Phys       Date:  2022-03-12       Impact factor: 2.243

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.