| Literature DB >> 34616225 |
Siyabonga P Khoza1, Sarah Ford2, Ernest P Buthelezi3, Donald M Tanyanyiwa4,2.
Abstract
BACKGROUND: BarricorTM Lithium heparin plasma tubes are new blood tubes that have been introduced to overcome the effects of gel in serum separator tubes (SST) and the shortcomings of standard Lithium heparin plasma. We aimed to evaluate BarricorTM tubes as an alternative to serum separator tubes and compare the stability between the tubes.Entities:
Keywords: BarricorTM; blood collection devices; serum separator tubes
Year: 2021 PMID: 34616225 PMCID: PMC8451222 DOI: 10.5937/jomb0-27216
Source DB: PubMed Journal: J Med Biochem ISSN: 1452-8266 Impact factor: 3.402
Comparison between serum separator tubes and BarricorTM lithium heparin tubes for the five chemistry analytes
SST = serum separator tube; AST = Aspartate aminotransferase; K = Potassium; LDH = Lactate dehydrogenase; CI = Confidence intervals; aThe baseline value considered serum/plasma measured less 6h post collection; bPassing-Bablok regression analysis; cSpearman correlation coef- ficient; dWilcoxon matched paired test, statistical significant if P< 0.05; eDifferences are calculated as = (Concentration in BD Barricor – Concentration in SST)/Concentration in SST*100; fTEa = 1.65(0.5xCVI) + 0.25(CVI2+CVG2)½ using biological variation data from EFLM
| SST tube | Barricor | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Analyte | Unit | Sample Number | Median Baseline valua (IQR) | Median Baseline valua (IQR) | PBb 95% CI Intercept | PBb 95% CI Slope | Rsc 95% CI | Pd | Mean Difference (%)e | TEaf |
| AST | U/L | 44 | 22 (17–30) | 23 (18–30) | 1.20 (-0.195–2.516) | 0.962 (0.9032–1.0209) | 0.919 (0.855–1.021) | <0.001 | 1.20 | 13.8 |
| K | mmol/L | 44 | 4.3 (4.0–4.5) | 3.9 (3.6–4.2) | -0.669 (-1.588–0.700) | 1.067 (0.923–1.289) | 0.730 (0.553–0.844) | <0.001 | -8.68 | 4.85 |
| Phosphate | mmol/L | 44 | 1.06 (0.93–1.18) | 0.95 (0.84–1.05) | -0.100 (-0.1802–0.016) | 1.00 (0.889–1.080) | 0.917 (0.853– 0.954) | <0.001 | -8.79 | 8.37 |
| LDH | U/L | 44 | 198 (178–240) | 210 (177–237) | 3.00 (-20.417–20.431) | 0.979 (0.896–1.081) | 0.862 (0.760–0.923) | <0.001 | -0.70 | 7.67 |
| Creatinine | mmol/L | 44 | 77 (63–89) | 74 (58–87) | -3.00 (-5.429–0.816) | 1.00 (0.947–1.029) | 0.984 (0.971–0.991) | <0.001 | -4.2 | 7.37 |
Figure 1Differences between BarricorTM and Serum separator tube for aspartate aminotransferase values by Bland-Altman analysis (bias% 1.7%). The dashed lines are the limit of agreements (LOA), which correspond to the mean ± 1.96 SD of the difference between the tubes.
Figure 2Differences between BarricorTM and Serum separator tube for potassium values by Bland-Altman analysis (bias%=-9.35%). The dashed lines are the limit of agreements (LOA), which correspond to the mean ± 1.96 SD of the difference between the tubes.
Figure 3Differences between BarricorTM and Serum separator tube for lactate dehydrogenase (LDH) values by Bland-Altman analysis (bias%=-0.46%). The dashed lines are the limit of agreements (LOA), which correspond to the mean ± 1.96 SD of the difference between the tubes.
Figure 4Differences between BarricorTM and Serum separator tube for phosphate values by Bland-Altman analysis (bias%=- 9.59 %). The dashed lines are the limit of agreements (LOA), which correspond to the mean ± 1.96 SD of the difference between the tubes.
Figure 5Differences between BarricorTM and Serum separator tube for creatinine values by Bland-Altman analysis (bias%=-4.56 %). The dashed lines are the limit of agreements (LOA), which correspond to the mean ± 1.96 SD of the difference between the tubes.
Mean difference over time for the five analytes
SST = serum separator tube, AST = Aspartate aminotransferase, K = Potassium, LDH = Lactate dehydrogenasea Differences are calculated as = (Concentration at point X – Concentration at baseline)/Concentration at baseline *100bTEa = 1.65(0.5xCVI) + 0.25(CVI2 + CVG2)½ using biological variation data from EFLM c RCV = Reference change value.
| Mean differences % at difference time pointsa | |||||||||
|---|---|---|---|---|---|---|---|---|---|
| Analyte | Tube | 24 hrs | 48 hrs | 72 hrs | 96 hrs | 120 hrs | 144 hrs | TEa | RCVc (%) |
| AST | SST | 6.4 | 7.7 | 8.2 | 8.4 | 8.4 | 8.8 | 13.8 | ±34 |
| Barricor | 4.3 | 7.5 | 8.3 | 10.4 | 11.6 | 12.7 | |||
| K | SST | 0.90 | 3.15 | 5.08 | 7.11 | 10.66 | 14.96 | 4.85 | ±13.7 |
| Barricor | 1.89 | 4.77 | 5.74 | 6.54 | 7.63 | 8.25 | |||
| LDH | SST | 3 | 0 | -1 | -4 | -6 | -7 | 7.67 | ±24 |
| Barricor | 1.1 | 3.1 | 4.8 | 6.8 | 9.5 | 11.0 | |||
| Phosphate | SST | 0.1 | 0.1 | 0.9 | 2.0 | 4.3 | 5.1 | 8.37 | ±23.5 |
| Barricor | 0.1 | 1.4 | 3.9 | 5.7 | 6.6 | 9.4 | |||
| Creatinine | SST | 2.0 | 2.0 | 2.8 | 2.8 | 3.4 | 5.0 | 7.37 | ±17.4 |
| Barricor | 0.34 | 0.81 | 1.87 | 2.41 | 2.77 | 4.73 | |||