Evaluation of BD Vacutainer® Barricor™ blood collection tubes for routine chemistry testing on a Roche Cobas® 8000 Platform.

Jeffrey E Fournier1, Victoria Northrup2, Claudette Clark3, Jacqueline Fraser4, Michael Howlett5, Paul Atkinson5, Jennifer L Shea6.   

Abstract

INTRODUCTION: Barricor™ Vacutainers® are a novel non-gel separator blood collection tube. These tubes enable faster pre-analytical processing which could reduce turnaround time and be beneficial in an acute care setting. We sought to evaluate the bias, stability, and integrity of plasma generated from these tubes compared to Plasma Separator Tubes™ (PST) for 50 routine chemistry analytes on a Roche Cobas® 8000 analyzer.
METHODS: Paired samples were collected from 150 patients originating in the emergency department and outpatient collections at the Saint John Regional Hospital. Barricor™ vacutainers were centrifuged for 3 min at 4000g and PST™ vacutainers for 10 min at 1300 g within two hours of collection. Plasma samples (n = 126) were then analyzed for 50 chemistry analytes and bias determined between tubes. Ten-day stability of AST, glucose, potassium, phosphate, and LDH was also assessed in a subset of paired samples (n = 4). Lastly, the quality of plasma (n = 20) was assessed through measurement of cell counts on a DxH Hematology Analyzer.
RESULTS: All 50 analytes demonstrated comparable results across a broad concentration range between Barricor™ and PST™ vacutainers (average percent bias -1.5% to 6.1%; Deming linear regression slopes 0.933-1.041; correlation coefficients ≥ 0.9144). AST, potassium, glucose and LDH were stable for 10 days in Barricor™ vacutainers (change from baseline < 10%) but <5 days in PST™ vacutainers while phosphate was stable for 4 days in Barricor™ vs 2 days in PST™ vacutainers. Platelet counts were statistically lower in Barricor™ compared to PST™ vacutainers.
CONCLUSION: Our data suggest that Barricor™ vacutainers are an acceptable alternative to PST™ vacutainers while offering the added benefit of decreased pre-analytical processing time, increased stability of certain analytes, and possibly less cellular contamination. Crown
Copyright © 2018. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Barricor™; PST™; Roche Cobas 8000; Routine chemistry testing

Mesh:

Year:  2018        PMID: 29885310     DOI: 10.1016/j.clinbiochem.2018.06.002

Source DB:  PubMed          Journal:  Clin Biochem        ISSN: 0009-9120            Impact factor:   3.281


  4 in total

1.  Comparison of Three Blood Collection Tubes for 35 Biochemical Analytes: The Becton Dickinson Barricor Tube, Serum Separating Tube, and Plasma Separating Tube.

Authors:  Sunghwan Shin; Jongwon Oh; Hyung-Doo Park
Journal:  Ann Lab Med       Date:  2020-08-25       Impact factor: 3.464

2.  A comparison of stability of chemical analytes in plasma from the BD Vacutainer® Barricor™ tube with mechanical separator versus tubes containing gel separator.

Authors:  Ghassaan Gawria; Linda Tillmar; Eva Landberg
Journal:  J Clin Lab Anal       Date:  2019-10-11       Impact factor: 2.352

3.  Evaluation of the Barricor Tube in 28 Routine Chemical Tests and Its Impact on Turnaround Time in an Outpatient Clinic.

Authors:  Soo Young Moon; Han Sol Lee; Min Soon Park; In-Suk Kim; Sun Min Lee
Journal:  Ann Lab Med       Date:  2021-05-01       Impact factor: 3.464

4.  Comparative study of chemical pathology sample collection tubes at the largest hospital in South Africa.

Authors:  Siyabonga P Khoza; Sarah Ford; Ernest P Buthelezi; Donald M Tanyanyiwa
Journal:  J Med Biochem       Date:  2021-09-03       Impact factor: 3.402

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.