Differences between values for plasma and serum in tests performed in the Ektachem 700 XR Analyzer, and evaluation of "plasma separator tubes (PST)".

B T Doumas1, L L Hause, D M Simuncak, D Breitenfeld.   

Abstract

We measured 25 analytes in plasma and serum from the same blood specimen, using the Kodak Ektachem 700 XR Analyzer. For 22 of the analytes, values for plasma were practically the same as those for serum; for the other three, differences between concentrations in plasma and serum were significant, both statistically and medically. Values for the same analytes in plasma from regular heparinized tubes were essentially indistinguishable from those for plasma obtained by use of Plasma Separator Tubes (PST).

Mesh:

Substances:

Year:  1989        PMID: 2910557

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  8 in total

1.  Impact of the ovarian cycle and pregnancy on plasma chemistry values in ewes.

Authors:  Micaela E Zywicki; Sharon E Blohowiak; Ronald R Magness; Jeffrey L Segar; Pamela J Kling
Journal:  J Vet Diagn Invest       Date:  2018-01-02       Impact factor: 1.279

2.  Stability of Routine Biochemical Analytes in Whole Blood and Plasma From Lithium Heparin Gel Tubes During 6-hr Storage.

Authors:  Denis Monneret; Alexandre Godmer; Ronan Le Guen; Clotilde Bravetti; Cecile Emeraud; Anthony Marteau; Rana Alkouri; Fouzi Mestari; Sylvie Dever; Françoise Imbert-Bismut; Dominique Bonnefont-Rousselot
Journal:  J Clin Lab Anal       Date:  2016-02-18       Impact factor: 2.352

3.  Chemistry Testing on Plasma Versus Serum Samples in Dialysis Patients: Clinical and Quality Improvement Implications.

Authors:  Roger Neill Carey; Chinu Jani; Curtis Johnson; Jim Pearce; Patricia Hui-Ng; Eduardo Lacson
Journal:  Clin J Am Soc Nephrol       Date:  2016-05-16       Impact factor: 8.237

4.  Potential inaccuracies in chloride measurements in patients with severe metabolic acidosis.

Authors:  Tetsuya Makiishi; Naomasa Nishimura; Keiko Yoshioka; Shinya Yamamoto; Ryuichi Mitsuhashi; Sayako Maeda; Takashi Konishi; Kunihiko Hirose
Journal:  Int J Nephrol       Date:  2012-06-12

5.  Improving home haemodialysis: Stability evaluation of routine clinical chemistry analytes in blood samples of haemodialysis patients.

Authors:  Lourens J P Nonkes; Maaike K van Gelder; Hans Kemperman; Alferso C Abrahams; Frans T J Boereboom; Maarten J Ten Berg; Karin G F Gerritsen
Journal:  Biochem Med (Zagreb)       Date:  2019-02-15       Impact factor: 2.313

6.  Comparative study of chemical pathology sample collection tubes at the largest hospital in South Africa.

Authors:  Siyabonga P Khoza; Sarah Ford; Ernest P Buthelezi; Donald M Tanyanyiwa
Journal:  J Med Biochem       Date:  2021-09-03       Impact factor: 3.402

7.  The local clinical validation of a new lithium heparin tube with a barrier: BD Vacutainer® Barricor LH Plasma tube.

Authors:  Fatma Demet Arslan; Inanc Karakoyun; Banu Isbilen Basok; Merve Zeytinli Aksit; Anil Baysoy; Yasemin Kilic Ozturk; Yusuf Adnan Guclu; Can Duman
Journal:  Biochem Med (Zagreb)       Date:  2017-08-28       Impact factor: 2.313

8.  Age- and Sex-Specific Pediatric Reference Intervals of Serum Electrolytes in Jilin Province of China Using the A Priori Approach.

Authors:  Xuetong Zhu; Kaijin Wang; Qi Zhou; Wenjia Guo; Yanan Jia; Jiancheng Xu
Journal:  Am J Clin Pathol       Date:  2020-10-13       Impact factor: 2.493

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.