Richard N McNeely1, Salissou Moutari2, Stephen Stewart1, Jonathan E Moore3,4,5,6.
Abstract
PURPOSE: To assess the 1-month and 12-month postoperative visual performance and subjective outcomes following combined implantation of an extended depth of focus (EDOF) intraocular lens (IOL) and a trifocal IOL.Entities:
Keywords: Cataract surgery; Extended depth of focus IOL; Quality of vision; Refractive lens exchange; Trifocal IOL; Visual outcomes
Mesh:
Year: 2021 PMID: 34546493 PMCID: PMC8572828 DOI: 10.1007/s10792-021-01970-3
Source DB: PubMed Journal: Int Ophthalmol ISSN: 0165-5701 Impact factor: 2.031
Demographics and clinical data
| Parameter | Preoperative | |
|---|---|---|
| AT LARA | AT LISA | |
| No. of patients (eyes) | 58 (58) | 58 (58) |
| Age (y), mean ± SD (range) | 59 ± 7.0 (44−81) | |
| Gender, male / female (%) | 19/39 (32.8/67.2) | |
| Myopia / hyperopia (%) | 9/49 (15.5/84.5) | |
| Axial length (mm), mean ± SD (range) | 23.54 ± 1.11 (21.47−27.27) | 23.59 ± 1.17 (20.98−27.35) |
| Power of implanted IOL (D), mean ± SD (range) | 20.76 ± 3.43 (9.5 to 27.5) | 21.51 ± 3.97 (7.0 to 29.0) |
| Clinical, mean ± SD (range) | ||
| Sphere (D) | 0.88 ± 2.77 (− 10−4.50) | 1.00 ± 2.99 (− 12−5.25) |
| Cylinder (D) | − 0.48 ± 0.45 (− 2−0) | − 0.54 ± 0.52 (− 2.25−0) |
| MSE (D) | 0.63 ± 2.85 (− 10.63−3.88) | 0.72 ± 3.06 (− 12.63−4.50) |
| CDVA | − 0.01 ± 0.28 (− 0.2−1.85) | − 0.02 ± 0.20 (− 0.2−1.20) |
SD = standard deviation; IOL = intraocular lens; D = diopters; MSE = manifest spherical equivalent; CDVA = corrected distance visual acuity
Fig. 1Cumulative binocular uncorrected a distance, b intermediate, and c near visual acuity 1 month and 12 months postoperatively. (UDVA = uncorrected distance visual acuity; CDVA = uncorrected distance visual acuity)
Comparison of 1-month and 12-month postoperative clinical data
| Parameter, mean ± SD (range) | Postoperative 1 | Postoperative 2 | ||||
|---|---|---|---|---|---|---|
| AT LARA | AT LISA | AT LARA | AT LISA | AT LARA | AT LISA | |
| 58 (58) | 58 (58) | 58 (58) | 58 (58) | 58 (58) | 58 (58) | |
| Sphere (D) | 0.48 ± 0.43 (− 0.50−1.75) | 0.31 ± 0.40 (− 0.50 −1.50) | 0.50 ± 0.40 (0 −1.75) | 0.34 ± 0.41 (− 0.50 −1.50) | 0.766 | 0.592 |
| Cylinder (D) | − 0.25 ± 0.34 (− 1.50 −0) | − 0.32 ± 0.37 (− 1.50 −0) | − 0.25 ± 0.31 (− 1.50 −0) | − 0.30 ± 0.36 (− 1.25 −0) | 0.805 | 0.709 |
| MSE (D) | 0.36 ± 0.37 (− 0.50 −1.50) | 0.16 ± 0.34 (− 0.88 −0.88) | 0.37 ± 0.35 (− 0.25 −1.13) | 0.19 ± 0.37 (− 0.50 −1.00) | 0.796 | 0.416 |
| UDVA (logMAR) | − 0.01 ± 0.11 (− 0.20 −0.24) | 0 ± 0.09 (− 0.20 −0.24) | − 0.02 ± 0.09 (− 0.20 −0.20) | − 0.02 ± 0.09 (− 0.20 −0.20) | 0.439 | 0.315 |
| Binocular UDVA (logMAR) | − 0.08 ± 0.07 (− 0.20 −0.10) | − 0.09 ± 0.06 (− 0.20 −0) | 0.323 | |||
| UIVA (logMAR) | 0.18 ± 0.16 (− 0.10 −0.70) | 0.23 ± 0.14 (− 0.10 −0.70) | 0.17 ± 0.12 (0 −0.60) | 0.19 ± 0.12 (0 −0.60) | 0.513 | 0.128 |
| Binocular UIVA (logMAR) | 0.15 ± 0.14 (− 0.10 −0.70) | 0.11 ± 0.10 (− 0.10 −0.40) | 0.030 | |||
| UNVA (logMAR) | 0.37 ± 0.16 (0.10 −0.80) | 0.19 ± 0.12 (0 −0.50) | 0.38 ± 0.15 (0.10 −0.70) | 0.18 ± 0.14 (0 −0.50) | 0.802 | 0.927 |
| Binocular UNVA (logMAR) | 0.17 ± 0.11 (0 −0.50) | 0.13 ± 0.10 (0 −0.40) | 0.008 | |||
| CDVA | − 0.08 ± 0.06 (− 0.20 −0.10) | − 0.06 ± 0.08 (− 0.20 −0.10) | − 0.07 ± 0.07 (− 0.20 −0.10) | − 0.07 ± 0.08 (− 0.20 −0.12) | 0.619 | 0.699 |
SD = standard deviation; D = diopters; MSE = manifest spherical equivalent; UDVA = uncorrected distance visual acuity; UIVA = uncorrected intermediate visual acuity; UNVA = uncorrected near visual acuity; CDVA = corrected distance visual acuity
Fig. 2Histogram showing the efficacy of lines of difference between postoperative UDVA and CDVA for a EDOF IOL and b trifocal IOL, at 1 month and 12 months postoperatively. (UDVA = uncorrected distance visual acuity, CDVA = corrected distance visual acuity)
Fig. 3Binocular defocus curve at 1 month. D = diopters
Fig. 4Predictability for a EDOF IOL and b trifocal IOL, at 1 month and 12 months postoperatively. (SE = spherical equivalent; D = dioptres)
Fig. 5Postoperative refractive cylinder for a EDOF IOL and b trifocal IOL, at 1 month and 12 months postoperatively
Fig. 6Stability up to 12 months postoperatively plotted as the mean ± SD of the SE refraction, for both IOL designs
Change in refractive error between 1 and 12 months postoperatively for the EDOF IOL and trifocal IOLs
| No change SE (within ± 0.25) | % myopic 0.26 −0.50 | % myopic 0.51 −1.00 | % myopic > 1.00 | % hyperopic 0.26 −0.50 | % hyperopic 0.51 −1.00 | % hyperopic > 1.00 | |
|---|---|---|---|---|---|---|---|
| EDOF IOL ( | 75.9 ( | 6.9 ( | 6.9 ( | − | 3.4 ( | 5.2 ( | 1.7 (n = 1) |
| Trifocal IOL ( | 58.9 ( | 14.3 ( | 1.8 ( | − | 19.6 ( | 5.4 ( | − |
Comparison of 1-month and 12-month patient experience postoperative data
| Postoperative assessment | Question | ||||
|---|---|---|---|---|---|
| Compare your vision preoperative to postoperative | |||||
| Better | Unchanged | Worse | Can’t remember | ||
| Postop 1 | 94.8% | 3.5% | 1.7% | 0% | |
| Postop 2 | 100% | 0% | 0% | 0% | |
| Would you choose this procedure again? | |||||
| Yes | No | Maybe | |||
| Postop 1 | 94.9% | 1.7% | 3.4% | ||
| Postop 2 | 98.3% | 0% | 1.7% | ||
| Would you recommend this procedure? | |||||
| Yes | No | Maybe | |||
| Postop 1 | 94.9% | 1.7% | 3.4% | ||
| Postop 2 | 98.3% | 0% | 1.7% | ||
| How often do you require reading glasses? | |||||
| Never | Occasionally | Quite often | Always | ||
| Postop 1 | 84.5% | 13.8% | 1.7% | 0% | |
| Postop 2 | 86.3% | 10.3% | 3.4% | 0% | |
| How much difficulty do you have doing a regular task that requires you to see well in the distance? | |||||
| Distance vision is clear | Slight problem | Moderate problem | Severe problem | Intolerable problem | |
| Postop 1 | 89.7% | 6.9% | 0% | 3.4% | 0% |
| Postop 2 | 89.7% | 10.3% | 0% | 0% | 0% |
| How much difficulty do you have doing a regular task that requires you to see well at intermediate working distances? | |||||
| Intermediate vision is clear | Slight problem | Moderate problem | Severe problem | Intolerable problem | |
| Postop 1 | 86.2% | 8.6% | 5.2% | 0% | 0% |
| Postop 2 | 89.6% | 5.2% | 5.2% | 0% | 0% |
| How much difficulty do you have doing a regular task that requires you to see well at near working distances? | |||||
| Near vision is clear | Slight problem | Moderate problem | Severe problem | Intolerable problem | |
| Postop 1 | 75.9% | 19.0% | 3.4% | 1.7% | 0% |
| Postop 2 | 77.6% | 19.0% | 3.4% | 0% | 0% |
| How were your expectations fulfilled with the procedure? | |||||
| More than fulfilled | Fulfilled | Sufficiently fulfilled | Not fulfilled at all | ||
| Postop 1 | 43.1% | 50.0% | 5.2% | 1.7% | |
| Postop 2 | 44.8% | 48.3% | 6.9% | 0% | |
Comparison of 1-month and 12-month postoperative visual phenomena data
| Postop 1 | Postop 2 | P Value | |
|---|---|---|---|
| Glare | 0.53 ± 0.92 (0, 3) | 0.43 ± 0.75 (0, 3) | .498 |
| Haloes | 0.64 ± 0.87 (0, 3) | 0.47 ± 0.68 (0, 2) | .176 |
| Starburst | 0.31 ± 0.75 (0, 3) | 0.29 ± 0.62 (0, 2) | .903 |
| Hazy vision | 0.14 ± 0.44 (0, 2) | 0.03 ± 0.18 (0, 1) | .063 |
| Blurred vision | 0.16 ± 0.45 (0, 2) | 0.14 ± 0.44 (0, 2) | .709 |
| Distortion | 0.02 ± 0.13 (0, 1) | 0 | .317 |
| Double vision | 0.05 ± 0.29 (0, 2) | 0.03 ± 0.26 (0, 2) | .317 |
| Vision Fluctuate | 0.17 ± 0.42 (0, 2) | 0.21 ± 0.45 (0, 2) | .317 |
| Depth perception | 0.10 ± 0.36 (0, 2) | 0.10 ± 0.36 (0, 2) | 1.000 |
| QoV Day | 9.12 ± 0.94 (7, 10) | 9.24 ± 0.78 (7, 10) | .183 |
| QoV Night | 7.88 ± 1.74 (2, 10) | 8.26 ± 1.38 (5, 10) | .043 |
Visual phenomena calculated on a scale 0 (not at all) to 3 (very). Values represent mean ± SD (range); QoV is calculated on a scale 0 (worst) to 10 (best). Values represent mean ± SD (range)