Minimal artifact actively shimmed metallic needles in MRI.

Saikat Sengupta1,2, Xinqiang Yan1,2, Tamarya L Hoyt2, Gary Drake1, Anthony Gunderman3, Yue Chen4.   

Abstract

PURPOSE: Signal voids caused by metallic needles pose visualization and monitoring challenges in many MRI applications. In this work, we explore a solution to this problem in the form of an active shim insert that fits inside a needle and corrects the field disturbance (ΔB0 ) caused by the needle outside of it.
METHODS: The ΔB0 induced by a 4 mm outside-diameter titanium needle at 3T is modeled and a two-coil orthogonal shim set is designed and fabricated to shim the ΔB0 . Signal recovery around the needle is assessed in multiple orientations in a water phantom with four different pulse sequences. Phase stability around the needle is assessed in an ex-vivo porcine tissue dynamic gradient echo experiment with and without shimming. Additionally, heating of the shim insert is assessed under 8 min of continuous operation with 1A current and concurrent imaging.
RESULTS: An average recovery of ~63% of lost signal around the needle across orientations is shown with active shimming with a maximum current of 1.172 A. Signal recovery and correction of the underlying ΔB0 is shown to be independent of imaging sequence. Needle-induced phase gradients outside the perceptible signal void are also minimized with active shimming. Temperature rise of up to 0.9° Celsius is noted over 8 min of continuous 1A active shimming operation.
CONCLUSION: A sequence independent method for minimization of metallic needle induced signal loss using an active shim insert is presented. The method has potential benefits in a range of qualitative and quantitative interventional MRI applications.
© 2021 International Society for Magnetic Resonance in Medicine.

Entities:  

Keywords:  active shimming; field inhomogeneity; interventional MRI; metallic needles; shim coils

Mesh:

Year:  2021        PMID: 34411348      PMCID: PMC8868020          DOI: 10.1002/mrm.28977

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  47 in total

1.  A perspective on needle artifacts in MRI: an electromagnetic model for experimentally separating susceptibility effects.

Authors:  A Glowinski; G Adam; A Bücker; J van Vaals; R W Günther
Journal:  IEEE Trans Med Imaging       Date:  2000-12       Impact factor: 10.048

2.  Remote control of catheter tip deflection: an opportunity for interventional MRI.

Authors:  T P L Roberts; W V Hassenzahl; S W Hetts; R L Arenson
Journal:  Magn Reson Med       Date:  2002-12       Impact factor: 4.668

3.  Numerical modeling of needle tip artifacts in MR gradient echo imaging.

Authors:  Bernd Müller-Bierl; Hansjörg Graf; Ulrike Lauer; Günter Steidle; Fritz Schick
Journal:  Med Phys       Date:  2004-03       Impact factor: 4.071

4.  Magnetic resonance-guided upper abdominal biopsies in a high-field wide-bore 3-T MRI system: feasibility, handling, and needle artefacts.

Authors:  Jens-Peter Kühn; Sönke Langner; Katrin Hegenscheid; Matthias Evert; Antje Kickhefel; Norbert Hosten; Ralf Puls
Journal:  Eur Radiol       Date:  2010-05-26       Impact factor: 5.315

5.  Real-Time Estimation of 3-D Needle Shape and Deflection for MRI-Guided Interventions.

Authors:  Yong-Lae Park; Santhi Elayaperumal; Bruce Daniel; Seok Chang Ryu; Mihye Shin; Joan Savall; Richard J Black; Behzad Moslehi; Mark R Cutkosky
Journal:  IEEE ASME Trans Mechatron       Date:  2010-12       Impact factor: 5.303

6.  Compensation of magnetic field distortions from paramagnetic instruments by added diamagnetic material: measurements and numerical simulations.

Authors:  Bernd Müller-Bierl; Hansjörg Graf; Günter Steidle; Fritz Schick
Journal:  Med Phys       Date:  2005-01       Impact factor: 4.071

7.  Characterization of the susceptibility artifact around a prostate brachytherapy seed in MRI.

Authors:  K Wachowicz; S D Thomas; B G Fallone
Journal:  Med Phys       Date:  2006-12       Impact factor: 4.071

8.  MR imaging--guided large-core (14-gauge) needle biopsy of small lesions visible at breast MR imaging alone.

Authors:  C K Kuhl; N Morakkabati; C C Leutner; A Schmiedel; E Wardelmann; H H Schild
Journal:  Radiology       Date:  2001-07       Impact factor: 11.105

9.  SEMAC: Slice Encoding for Metal Artifact Correction in MRI.

Authors:  Wenmiao Lu; Kim Butts Pauly; Garry E Gold; John M Pauly; Brian A Hargreaves
Journal:  Magn Reson Med       Date:  2009-07       Impact factor: 4.668

10.  A novel MRI marker for prostate brachytherapy.

Authors:  Steven J Frank; R Jason Stafford; James A Bankson; Chun Li; David A Swanson; Rajat J Kudchadker; Karen S Martirosyan
Journal:  Int J Radiat Oncol Biol Phys       Date:  2008-05-01       Impact factor: 7.038

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.