Literature DB >> 26405428

Real-Time Estimation of 3-D Needle Shape and Deflection for MRI-Guided Interventions.

Yong-Lae Park1, Santhi Elayaperumal2, Bruce Daniel3, Seok Chang Ryu2, Mihye Shin2, Joan Savall4, Richard J Black5, Behzad Moslehi5, Mark R Cutkosky2.   

Abstract

We describe a MRI-compatible biopsy needle instrumented with optical fiber Bragg gratings for measuring bending deflections of the needle as it is inserted into tissues. During procedures, such as diagnostic biopsies and localized treatments, it is useful to track any tool deviation from the planned trajectory to minimize positioning errors and procedural complications. The goal is to display tool deflections in real time, with greater bandwidth and accuracy than when viewing the tool in MR images. A standard 18 ga × 15 cm inner needle is prepared using a fixture, and 350-μm-deep grooves are created along its length. Optical fibers are embedded in the grooves. Two sets of sensors, located at different points along the needle, provide an estimate of the bent profile, as well as temperature compensation. Tests of the needle in a water bath showed that it produced no adverse imaging artifacts when used with the MR scanner.

Entities:  

Keywords:  Biomedical transducers; Bragg gratings; biopsy; magnetic resonance imaging (MRI); optical fiber sensors; strain measurement

Year:  2010        PMID: 26405428      PMCID: PMC4577522          DOI: 10.1109/TMECH.2010.2080360

Source DB:  PubMed          Journal:  IEEE ASME Trans Mechatron        ISSN: 1083-4435            Impact factor:   5.303


  21 in total

1.  Towards MRI guided surgical manipulator.

Authors:  K Chinzei; K Miller
Journal:  Med Sci Monit       Date:  2001 Jan-Feb

2.  In vitro study of needle deflection: a linear insertion technique versus a bidirectional rotation insertion technique.

Authors:  M N Hochman; M J Friedman
Journal:  Quintessence Int       Date:  2000-01       Impact factor: 1.677

3.  MRI-Compatible Pneumatic Robot for Transperineal Prostate Needle Placement.

Authors:  Gregory S Fischer; Iulian Iordachita; Csaba Csoma; Junichi Tokuda; Simon P Dimaio; Clare M Tempany; Nobuhiko Hata; Gabor Fichtinger
Journal:  IEEE ASME Trans Mechatron       Date:  2008-06-01       Impact factor: 5.303

4.  Brachytherapy needle deflection evaluation and correction.

Authors:  Gang Wan; Zhouping Wei; Lori Gardi; Donal B Downey; Aaron Fenster
Journal:  Med Phys       Date:  2005-04       Impact factor: 4.071

5.  MR-guided breast biopsy using an active marker: a phantom study.

Authors:  Richard Werner; Sascha Krueger; Axel Winkel; Christoph Albrecht; Tobias Schaeffter; Martin Heller; Christian Frahm
Journal:  J Magn Reson Imaging       Date:  2006-07       Impact factor: 4.813

Review 6.  Needle insertion into soft tissue: a survey.

Authors:  Niki Abolhassani; Rajni Patel; Mehrdad Moallem
Journal:  Med Eng Phys       Date:  2006-08-28       Impact factor: 2.242

7.  Dynamic MRI scan plane control for passive tracking of instruments and devices.

Authors:  S P DiMaio; E Samset; G Fischer; I Iordachita; G Fichtinger; F Jolesz; C M Tempany
Journal:  Med Image Comput Comput Assist Interv       Date:  2007

8.  Transperineal prostate biopsy under magnetic resonance image guidance: a needle placement accuracy study.

Authors:  Philip Blumenfeld; Nobuhiko Hata; Simon DiMaio; Kelly Zou; Steven Haker; Gabor Fichtinger; Clare M C Tempany
Journal:  J Magn Reson Imaging       Date:  2007-09       Impact factor: 4.813

Review 9.  The role of magnetic susceptibility in magnetic resonance imaging: MRI magnetic compatibility of the first and second kinds.

Authors:  J F Schenck
Journal:  Med Phys       Date:  1996-06       Impact factor: 4.071

10.  Intravascular MR tracking catheter: preliminary experimental evaluation.

Authors:  D A Leung; J F Debatin; S Wildermuth; G C McKinnon; D Holtz; C L Dumoulin; R D Darrow; E Hofmann; G K von Schulthess
Journal:  AJR Am J Roentgenol       Date:  1995-05       Impact factor: 3.959

View more
  30 in total

1.  Modeling of active shimming of metallic needles for interventional MRI.

Authors:  Saikat Sengupta
Journal:  Magn Reson Med       Date:  2020-06-29       Impact factor: 4.668

2.  Large Deflection Shape Sensing of a Continuum Manipulator for Minimally-Invasive Surgery.

Authors:  Hao Liu; Amirhossein Farvardin; Sahba Aghajani Pedram; Iulian Iordachita; Russell H Taylor; Mehran Armand
Journal:  IEEE Int Conf Robot Autom       Date:  2015-05-26

3.  Enhancement of accuracy in shape sensing of surgical needles using optical frequency domain reflectometry in optical fibers.

Authors:  Francois Parent; Sebastien Loranger; Koushik Kanti Mandal; Victor Lambin Iezzi; Jerome Lapointe; Jean-Sébastien Boisvert; Mohamed Diaa Baiad; Samuel Kadoury; Raman Kashyap
Journal:  Biomed Opt Express       Date:  2017-03-16       Impact factor: 3.732

4.  Teleoperation System with Hybrid Pneumatic-Piezoelectric Actuation for MRI-Guided Needle Insertion with Haptic Feedback.

Authors:  Weijian Shang; Hao Su; Gang Li; Gregory S Fischer
Journal:  Rep U S       Date:  2013

5.  Vessel-based registration of an optical shape sensing catheter for MR navigation.

Authors:  Koushik Mandal; Francois Parent; Sylvain Martel; Raman Kashyap; Samuel Kadoury
Journal:  Int J Comput Assist Radiol Surg       Date:  2016-03-16       Impact factor: 2.924

6.  Development of a compact continuum tubular robotic system for nasopharyngeal biopsy.

Authors:  Liao Wu; Shuang Song; Keyu Wu; Chwee Ming Lim; Hongliang Ren
Journal:  Med Biol Eng Comput       Date:  2016-05-18       Impact factor: 2.602

7.  Preliminary study on magnetic tracking-based planar shape sensing and navigation for flexible surgical robots in transoral surgery: methods and phantom experiments.

Authors:  Shuang Song; Changchun Zhang; Li Liu; Max Q-H Meng
Journal:  Int J Comput Assist Radiol Surg       Date:  2017-10-05       Impact factor: 2.924

Review 8.  Needle-tissue interactive mechanism and steering control in image-guided robot-assisted minimally invasive surgery: a review.

Authors:  Pan Li; Zhiyong Yang; Shan Jiang
Journal:  Med Biol Eng Comput       Date:  2018-04-21       Impact factor: 2.602

9.  Real-time active MR-tracking of metallic stylets in MR-guided radiation therapy.

Authors:  Wei Wang; Charles L Dumoulin; Akila N Viswanathan; Zion T H Tse; Alireza Mehrtash; Wolfgang Loew; Isaiah Norton; Junichi Tokuda; Ravi T Seethamraju; Tina Kapur; Antonio L Damato; Robert A Cormack; Ehud J Schmidt
Journal:  Magn Reson Med       Date:  2014-06-05       Impact factor: 4.668

10.  Autonomous real-time interventional scan plane control with a 3-D shape-sensing needle.

Authors:  Santhi Elayaperumal; Juan Camilo Plata; Andrew B Holbrook; Yong-Lae Park; Kim Butts Pauly; Bruce L Daniel; Mark R Cutkosky
Journal:  IEEE Trans Med Imaging       Date:  2014-06-23       Impact factor: 10.048

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.