Takuo Tsurugi1, Junjiroh Koyama1, Kazuhisa Kodama1, Hiroshi Nakajima2, Tomohiro Sakamoto1, Ken Okumura1.
Abstract
Entities:
Keywords: Defibrillation failure; Riata ST Optim lead; Short circuit; Silent lead malfunction; Ventricular fibrillation
Year: 2021 PMID: 34307037 PMCID: PMC8283545 DOI: 10.1016/j.hrcr.2021.04.009
Source DB: PubMed Journal: HeartRhythm Case Rep ISSN: 2214-0271
Figure 1Episode summary revealed that ventricular fibrillation (VF) detection occurred following the charge of capacitor to 35 J, but a 0.0 J shock was delivered owing to the excessive current protection system.
Figure 2Ventricular fibrillation (VF) could not be terminated by a 0.0 J shock and burst pacing (black arrows), which continued for 20 minutes.