Differential lead component pulling as a possible mechanism of inside-out abrasion and conductor cable externalization.

Ernest W Lau1.   

Abstract

BACKGROUND: Conductor cable externalization with protrusion (CCE*) is highly prevalent among the Riata 8F and ST 7F defibrillation (DF) leads and infrequently present in the QuickSite and the QuickFlex coronary sinus (CS) leads (St. Jude Medical, Sylmar, CA, USA). A model for CCE* based on differential lead component pulling and conjugate extension with reciprocal compression-bending was developed. Extension of a proximal lead body segment by pectoral or cardiac movements causes reciprocal compression-bending of a distal lead body segment mediated by inextensible conductor cables running down a lead body fixed at various points by fibrous adhesions. The "sawing" action of these cables under tension causes inside-out abrasion of insulation leading to CCE*.
METHODS: DF leads from different manufacturers and the QuickFlex and QuickFlex μ CS leads were subjected to simulated differential pulling.
RESULTS: Restitution from differential pulling followed three patterns: complete, partial without escalation, and incomplete with escalation. Only the last pattern (only shown by the Riata 8F and ST 7F leads) was associated with an increased risk to CCE*. For CS leads, deformation concentrated on the more flexible segment when the lead body did not have a uniform construction.
CONCLUSIONS: The Durata, Riata ST Optim, QuickFlex μ, and Quartet leads should be relatively immune to CCE*. The Durata leads are extremely resistant to longitudinal deformation and probably cause mediastinal displacement rather than differential pulling in response to pectoral movements in vivo. Implantation techniques and lead designs can be used to minimize the risk of CCE*. A bench test for CCE* can be constructed. ©2013, The Author. Journal compilation ©2013 Wiley Periodicals, Inc.

Entities:  

Keywords:  Durata; Quartet; QuickFlex; QuickSite; Riata; conductor cable externalization; differential lead component pulling; inside-out abrasion

Mesh:

Year:  2013        PMID: 23746039     DOI: 10.1111/pace.12185

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  4 in total

1.  A case of Durata ICD lead coil externalization: Inside-out lead abrasion?

Authors:  Ian E Mann; Oliver R Segal
Journal:  HeartRhythm Case Rep       Date:  2016-04-25

2.  Inside-out abrasion and contained conductor cable externalization in a defibrillation lead with asymmetric conductor cable lumen distribution.

Authors:  Ernest W Lau; Michael J D Roberts
Journal:  HeartRhythm Case Rep       Date:  2018-01-11

3.  Defibrillation failure with an electrical short circuit caused by internal insulation breach.

Authors:  Takuo Tsurugi; Junjiroh Koyama; Kazuhisa Kodama; Hiroshi Nakajima; Tomohiro Sakamoto; Ken Okumura
Journal:  HeartRhythm Case Rep       Date:  2021-04-29

4.  Conductor cable externalization in an atrial hemodynamic sensor lead in a patient presenting with inappropriate shocks.

Authors:  Sing-Chien Yap; Mark G Hoogendijk; Suzanne D A Valk; Sweder W Van de Poll; Peter van der Kemp; Tamas Szili-Torok
Journal:  HeartRhythm Case Rep       Date:  2019-09-10