Monitoring for and Diagnosis of Lead Dysfunction.

Sandeep G Nair1, Charles D Swerdlow2.   

Abstract

The predominant structural mechanisms of transvenous lead dysfunction (LD) are conductor fracture and insulation breach. LD typically presents as an abnormality of electrical performance; the earliest sign usually is either oversensing or out-of-range pacing or shock impedance. Accurate diagnosis of LD requires discriminating patterns of oversensing and impedance trends that are characteristic of LD from similar patterns that occur in other conditions. Implantable cardioverter-defibrillators have advanced features to detect and mitigate the consequences of LD; these features operate both independently and in conjunction with remote monitoring networks.
Copyright © 2018 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Implantable cardioverter-defibrillator; Lead dysfunction; Lead failure; Pacemaker

Mesh:

Year:  2018        PMID: 30396573     DOI: 10.1016/j.ccep.2018.07.004

Source DB:  PubMed          Journal:  Card Electrophysiol Clin        ISSN: 1877-9182


  3 in total

1.  Case Report of Novel, Automatic Shocking Vector Adjustment Algorithm: A Life-Saving Feature of a Modern Defibrillator.

Authors:  Mark R Heckle; Sunil K Jha
Journal:  Perm J       Date:  2022-04-05

2.  A case report of lead dysfunction presenting as high ventricular premature complex burden.

Authors:  Saurabh Ajit Deshpande; Ameya Udyavar
Journal:  Indian Pacing Electrophysiol J       Date:  2021-10-16

3.  Defibrillation failure with an electrical short circuit caused by internal insulation breach.

Authors:  Takuo Tsurugi; Junjiroh Koyama; Kazuhisa Kodama; Hiroshi Nakajima; Tomohiro Sakamoto; Ken Okumura
Journal:  HeartRhythm Case Rep       Date:  2021-04-29
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.