Prospective long-term evaluation of Optim-insulated (Riata ST Optim and Durata) implantable cardioverter-defibrillator leads.

John A Cairns1, Andrew E Epstein2, John Rickard3, Stuart J Connolly4, Christopher Buller5, Bruce L Wilkoff6, Janice Pogue4, Ellison Themeles4, Jeff S Healey4.   

Abstract

BACKGROUND: St Jude Medical Optim-insulated implantable cardioverter-defibrillator leads were designed to impart lubricity, strength, and abrasion resistance while maintaining flexibility and biostability. No long-term prospective follow-up data have been published.
OBJECTIVE: The objective of this study was to determine the rates of all-cause mechanical failure and its subtypes (conductor fracture, insulation abrasion, externalized conductors, and other mechanical failures) in a prospective cohort of Optim-insulated implantable cardioverter-defibrillator leads.
METHODS: St Jude Medical established 3 prospective registries and enrolled 11,016 leads implanted in 10,835 patients beginning in 2006. There was standardized baseline documentation, 6-monthly follow-up, adverse events reports (verified by expert staff using detailed algorithms), and documentation of lead revisions or inactivation, study withdrawal, and death. The Population Health Research Institute (McMaster University) was engaged to review database functions, adjudicate all potential mechanical lead failures, and conduct independent analyses of the data.
RESULTS: During a median follow-up of 3.2 years, there were 51 mechanical failures (0.46%), with 99.0% survival free of this outcome by 5 years of follow-up. Freedom from conductor fracture was identified in 99.4% and from all-cause abrasion in 99.8% of the leads, and there were no reports of externalized conductors. There were no significant differences in survival among Durata DF4, Durata DF1, and Riata ST Optim leads.
CONCLUSION: Over a mean follow-up of 3.2 years, Optim-insulated leads have low rates of all-cause mechanical failure and no observed externalized conductors. Independent analyses of these registries are designed to provide reliable long-term follow-up information and are ongoing.
Copyright © 2014 Heart Rhythm Society. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  ICD leads; Mechanical lead failure; Optim insulation; Prospective cohort study; Prospective registries

Mesh:

Year:  2014        PMID: 25131665     DOI: 10.1016/j.hrthm.2014.08.023

Source DB:  PubMed          Journal:  Heart Rhythm        ISSN: 1547-5271            Impact factor:   6.343


  3 in total

1.  What did we learn from Riata™?

Authors:  Avi Fischer; C Jenney
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2015-06-09

2.  Prospective long-term follow-up of silicone-polyurethane-insulated implantable cardioverter-defibrillator leads.

Authors:  John A Cairns; Jeff S Healey; Andrew E Epstein; Ellison Themeles; Kumar Balasubramanian; Stuart J Connolly
Journal:  Heart Rhythm O2       Date:  2021-12-01

3.  Defibrillation failure with an electrical short circuit caused by internal insulation breach.

Authors:  Takuo Tsurugi; Junjiroh Koyama; Kazuhisa Kodama; Hiroshi Nakajima; Tomohiro Sakamoto; Ken Okumura
Journal:  HeartRhythm Case Rep       Date:  2021-04-29