Sabine Visser1, Hendrike Neh2, Cássia Oraboni Ribeiro1, Erik W Korevaar1, Arturs Meijers1, Björn Poppe2, Nanna M Sijtsema1, Stefan Both1, Johannes A Langendijk1, Christina T Muijs1, Antje C Knopf1,2.
Abstract
PURPOSE: To ensure target coverage in the treatment of esophageal cancer, a density override to the region of diaphragm motion can be applied in the optimization process. Here, we evaluate the benefit of this approach during robust optimization for intensity modulated proton therapy (IMPT) planning.Entities:
Keywords: 3D robust optimization; density override; diaphragm; esophageal cancer; intensity modulated proton therapy
Mesh:
Year: 2021 PMID: 34289123 PMCID: PMC9291176 DOI: 10.1002/mp.15114
Source DB: PubMed Journal: Med Phys ISSN: 0094-2405 Impact factor: 4.506
Analyzed amplitudes and baseline shifts of the left and right diaphragm during all weeks of treatment. A positive shift reflects a more cranial position of the diaphragm
| Patient | Week 0 | Week 1 | Week 2 | Week 3 | Week 4 | Week 5 | Week 6 | ||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| A (cm) | A (cm) | S (cm) | A (cm) | S (cm) | A (cm) | S (cm) | A (cm) | S (cm) | A (cm) | S (cm) | A (cm) | S (cm) | |
|
Amplitude and baseline shift left diaphragm | |||||||||||||
| 1 | 1.4 | 1.4 | −0.9 | 1.8 | −1.7 | 1.8 | −0.9 | 2.0 | −1.1 | ||||
| 2 | 1.2 | 0.8 | +0.2 | 1.4 | +0.2 | 1.4 | +0.5 | 1.6 | +0.2 | 1.4 | +0.0 | ||
| 3 | 1.0 | 1.4 | +0.6 | 1.2 | −0.1 | 0.8 | −0.2 | 1.0 | +0.3 | 1.0 | +0.5 | 1.0 | +1.3 |
| 4 | 0.8 | 1.6 | −0.7 | 0.8 | +0.0 | 1.2 | −1.1 | 0.8 | −1.3 | ||||
| 5 | 1.2 | 1.4 | +0.2 | 1.6 | +0.4 | 1.2 | +0.8 | 1.2 | −0.3 | 1.2 | +0.2 | ||
| 6 | 1.2 | 0.6 | +0.6 | 1.2 | +0.3 | 0.4 | −0.6 | 1.2 | +0.5 | 0.8 | +0.2 | ||
| 7 | 1.0 | 0.6 | −0.2 | 0.8 | −0.1 | 0.8 | −0.5 | 1.0 | +0.1 | 1.2 | −0.3 | ||
| 8 | 1.2 | 0.8 | −0.5 | 1.0 | −0.3 | 1.0 | −0.1 | 1.0 | −0.7 | 1.2 | −0.5 | ||
| 9 | 0.8 | 1.0 | +3.2 | 0.8 | +2.2 | 1.6 | +2.8 | 1.0 | +2.4 | 1.8 | +2.6 | 1.6 | +2.9 |
| 10 | 1.0 | 1.0 | +0.7 | 1.4 | +0.3 | 1.2 | +0.5 | 1.0 | +0.5 | 1.0 | +0.2 | ||
| Amplitude and baseline shift right diaphragm | |||||||||||||
| 1 | 2.0 | 0.8 | −1.3 | 0.8 | −1.7 | 0.6 | −1.3 | 1.2 | −0.3 | ||||
| 2 | 1.2 | 1.2 | +0.6 | 1.6 | +0.2 | 1.4 | +0.5 | 1.6 | +0.0 | 2.2 | +0.0 | ||
| 3 | 2.2 | 1.6 | +0.8 | 1.2 | −1.5 | 1.8 | −0.6 | 1.6 | −0.1 | 1.2 | +0.5 | 1.4 | +0.9 |
| 4 | 1.2 | 1.8 | −0.9 | 1.4 | −0.6 | 0.8 | −2.1 | 0.8 | −2.1 | ||||
| 5 | 1.0 | 0.8 | +0.2 | 1.2 | 0.4 | 0.4 | +0.6 | 0.6 | 0.3 | 1.0 | +0.4 | ||
| 6 | 1.0 | 1.0 | 0.0 | 1.0 | +0.1 | 0.6 | −0.8 | 1.2 | −0.5 | 1.2 | −0.4 | ||
| 7 | 1.4 | 1.6 | 0.0 | 1.0 | −0.1 | 1.0 | −0.5 | 1.6 | −0.5 | 0.8 | −0.5 | ||
| 8 | 1.4 | 1.2 | −0.5 | 1.2 | −0.1 | 1.8 | +0.3 | 1.4 | −0.1 | 1.6 | −0.1 | ||
| 9 | 1.2 | 1.6 | +3.2 | 1.6 | +2.6 | 1.8 | +2.6 | 1.4 | +2.2 | 2.2 | +2.6 | 2.0 | +2.7 |
| 10 | 0.4 | 0.8 | +0.9 | 0.8 | +0.5 | 0.8 | +0.9 | 1.0 | +0.5 | 0.6 | +0.8 | ||
Abbreviations: A, amplitude; S, baseline shift.
FIGURE 1In red, the diaphragm override structure is shown on the maximal inspiration phase (left), the maximal expiration phase (middle), and on the average CT (right)
FIGURE 2Mean heart dose (MHD) and mean lung dose (MLD) obtained with the WDO and the NDO plans
FIGURE 3Vwmax dose distributions on the average CT at planning stage with (WDO) and without (NDO) the diaphragm density override of patient 1 (a) and patient 2 (b). The diaphragm override region is delineated in red. The beam setup is shown in white
(Top) Hotspots in the weekly robustness evaluations of both WDO and NDO plans. The numbers highlighted in red (bold) resemble higher hotspots (>110%). (Bottom) Target coverage in the weekly robustness evaluations of both WDO and NDO plans. The numbers highlighted in red (bold) indicate where the target coverage criteria (D98% > 94%) was not met
| Patient | Week 0 | Week 1 | Week 2 | Week 3 | Week 4 | Week 5 | Week 6 | |||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| WDO | NDO | WDO | NDO | WDO | NDO | WDO | NDO | WDO | NDO | WDO | NDO | WDO | NDO | |
| Body D2cm3 | ||||||||||||||
| 1 | 108% | 108% | 109% |
| 109% |
| 109% |
| 109% |
| 108% | 109% | ||
| 2 | 108% | 109% | 108% | 110% | 108% | 108% | 109% | 111% | 108% | 108% | 110% | 108% | ||
| 3 | 108% | 108% | 108% | 108% | 108% | 108% | 108% | 108% | 108% | 108% | 108% | 108% | 108% | 108% |
| 4 | 108% | 108% | 109% | 109% | 109% | 109% | 109% | 109% | 108% | 108% | ||||
| 5 | 107% | 107% | 107% | 107% | 107% | 107% | 107% | 107% | 107% | 107% | 107% | 107% | ||
| 6 | 106% | 107% | 106% | 106% | 106% | 106% | 107% | 107% | 106% | 107% | 107% | 107% | ||
| 7 | 108% | 107% | 108% | 108% | 107% | 107% | 108% | 107% | 109% | 108% | 108% | 107% | ||
| 8 | 107% | 107% | 108% | 107% | 108% | 108% | 108% | 108% | 108% | 108% | 108% | 108% | ||
| 9 | 108% | 108% | 110% |
| 110% | 110% | 110% |
| 110% |
| 110% |
|
|
|
| 10 | 108% | 109% | 109% | 109% | 109% | 109% | 109% | 109% | 109% | 109% | 109% | 109% | ||
| ITV D98% | ||||||||||||||
| 1 | 97% | 97% | 96% | 96% | 96% | 97% |
|
| 95% | 96% | 96% | 96% | ||
| 2 | 96% | 97% | 95% | 96% | 96% | 96% | 96% | 96% | 96% | 96% | 96% | 96% | ||
| 3 | 97% | 97% | 97% | 97% | 97% | 97% | 97% | 97% | 97% | 97% | 96% | 96% |
|
|
| 4 | 97% | 97% | 97% | 97% | 96% | 97% |
|
| 96% | 96% | ||||
| 5 | 96% | 96% | 96% | 96% | 95% | 96% | 96% | 96% | 96% | 96% | 96% | 96% | ||
| 6 | 96% | 96% | 96% | 96% | 96% | 96% | 96% | 96% | 96% | 97% | 96% | 96% | ||
| 7 | 97% | 97% | 97% | 97% | 97% | 97% | 97% | 97% | 97% | 97% | 97% | 97% | ||
| 8 | 97% | 97% | 97% | 97% | 97% | 97% | 97% | 97% | 97% | 97% | 97% | 97% | ||
| 9 | 97% | 97% |
|
|
|
|
|
|
|
|
|
|
|
|
| 10 | 97% | 97% | 97% | 97% | 97% | 97% | 97% | 97% | 96% | 97% | 97% | 97% | ||
FIGURE 4Visualization of the target dose coverage in the weekly robustness evaluations of both WDO and NDO plans based on the evaluation on the average CT (ITV) and on all the phases (CTV). Phases‐based evaluation gave typically higher target dose coverage than average CT‐based evaluation
FIGURE 5(a) Axial and coronal view of the image registration between the average CT of week 0 (blue) and the average CT of week 1 (orange) of patient 9. The ITV is shown for week 0 and week 1 in blue and orange, respectively. The position of the diaphragm was found higher in the repeated CT compared to the planning CT. The diaphragm variation caused a displacement of the ITV and higher densities in the beam path of the right posterior oblique field (white arrow). (b) Presentation of the 95% dose coverage (green) and the 80% dose (blue) in the Vwmin dose in week 1 of patient 9. The ITV is shown in orange