Nurhaslina Hasan1,2, Nur Fatihah Ronny Sham1, Muhammad Khalis Abdul Karim3, Syed Baharom Syed Ahmad Fuad1, Narimah Abdul Hamid Hasani1, Effat Omar1, Mohammad Johari Ibahim4.
Abstract
We presented a development of a custom lead shield and mouse strainer for targeted irradiation from the gamma-cell chamber. This study was divided into two parts i.e., to (i) fabricate the shield and strainer from a lead (Pb) and (ii) optimize the irradiation to the mice-bearing tumour model with 2 and 8 Gy absorbed doses. The lead shielding was fabricated into a cuboid shape with a canal on the top and a hole on the vertical side for the beam path. Respective deliveries doses of 28 and 75 Gy from gamma-cell were used to achieve 2 and 8 Gy absorbed doses at the tumour sites.Entities:
Year: 2021 PMID: 34267293 PMCID: PMC8282798 DOI: 10.1038/s41598-021-93964-5
Source DB: PubMed Journal: Sci Rep ISSN: 2045-2322 Impact factor: 4.379
Figure 1The shape of the lead shield in different views angle (a) Full dimension (b) Top view (c) Front view (d) Side view (Scale in cm).
Figure 2The shape of the mouse strainer in different views angle. There is a hole (1 × 1 cm) for the beam path. (a) Full dimension (b) Side view (c) Top view. (Scale in cm).
Figure 3Illustration of cross-sectional from the lateral side of the lead shield in the gamma cell chamber from Co60 source (Scale in cm).
Figure 4The calibration curve of gamma cell dose using polynomial power of 3. (Data n = 2, mean ± standard deviation). The calibration curve equation obtained is y = 122.78x3 + (− 42.542) x2 + 4.6201 + (− 0.0112).
The absorbed dose in the beam path from different delivery dose.
| Dose delivery from Gamma-cell (Gy) | Dose absorb (Measure by EBT Gafchromic Film) (Gy) | |
|---|---|---|
| Mean | Standard deviation | |
| 30 | 2.13 | 0.12 |
| 32 | 2.56 | 0.05 |
| 34 | 3.06 | 0.24 |
| 70 | 6.84 | 0.94 |
| 80 | 11.45 | 1.45 |
| 90 | 13.26 | 0.87 |
n = 2.
The absorbed dose from 28 and 75 Gy delivery dose.
| Mice No | Absorbed dose (Gy) from 28 Gy dose delivery | Absorbed dose (Gy) from 75 Gy dose delivery |
|---|---|---|
| 1 | 2.03 | 8.24 |
| 2 | 2.18 | 8.77 |
| 3 | 1.84 | 8.84 |
| 4 | 1.94 | 9.28 |
| 5 | 2.12 | 8.82 |
| 6 | 2.23 | 8.97 |
| Mean | 2.06 | 8.82 |
| Standard deviation | 0.15 | 0.34 |