Abstract
BACKGROUND: Preclinical radiation biology has become increasingly sophisticated due to the implementation of advanced small animal image guided radiation platforms into laboratory investigation. These small animal radiotherapy devices enable state-of-the-art image guided therapy (IGRT) research to be performed by combining high-resolution cone beam computed tomography (CBCT) imaging with an isocentric irradiation system. Such platforms are capable of replicating modern clinical systems similar to those that integrate a linear accelerator with on-board CBCT image guidance.Entities:
Keywords: Commissioning; Preclinical radiotherapy; Radiobiology; Small animal radiotherapy research platform (SARRP)
Mesh:
Year: 2017 PMID: 29282134 PMCID: PMC5745702 DOI: 10.1186/s13014-017-0936-3
Source DB: PubMed Journal: Radiat Oncol ISSN: 1748-717X Impact factor: 3.481
Fig. 1Depth dose deposition profile 0.5 mm apperture using a broad a) and c) and a fine focus b) and d) beam for the three SSDs:31 cm (black), 34 cm (red) and 38 cm (blue). Data was analysed and quantified using FilmQA software a) and b) and Matlab codes c) and d). Data represents the average ± standard error (n = 3)
Fig. 2Beam uniformity across the irradiated area using a 0.5 mm diameter aperture and a broad focus at: 31 cm SSD a) and b), 34 cm SSD c) and d), and 38 cm SSD e) and f). Depths of 0.15 mm: a),c),e), and 72.55 mm: b),d),f) are shown. Measured profiles along x axis are black solid lines, and profiles along y axis are shown in red solid lines. Monte Carlo calculated beam profiles are dashed lines (black – x axis and red- y axis)
Fig. 3Beam uniformity across the irradiated area using a 0.5 mm diameter aperture and a fine focus at: 31 cm SSD for a) and b), 34 cm SSD c) and d), and 38 cm SSD e) and f) . Depths of 0.15 mm a),c),e), and 72.55 mm b),d),f) are shown. Measured profiles along x axis are black solid lines, and profiles along y axis are shown in red solid lines. Monte Carlo calculated beam profiles are dashed lines (black – x axis and red- y axis)
Fig. 4Beam Penumbra presented after 3 independent repeats at different times. FilmQA Pro data is presented for bright focus a) and fine focus b). Manufacturer data is presented for bright focus c) and fine focus d). Data presented represents an average ± standard error (n = 3)
Fig. 5Monte Carlo simulations (solid lines) for bright focus (black) and fine focus (red) as compared with measured data (symbols) for a) 31 cm SSD, b)34 cm SSD and c) 38 cm SSD