Analysis of the EBT3 Gafchromic film irradiated with 6 MV photons and 6 MeV electrons using reflective mode scanners.

Nicolas Farah1, Ziad Francis2, Marie Abboud3.   

Abstract

We explore in our study the effects of electrons and X-rays irradiations on the newest version of the Gafchromic EBT3 film. Experiments are performed using the Varian "TrueBeam 1.6" medical accelerator delivering 6 MV X-ray photons and 6 MeV electron beams as desired. The main interest is to compare the responses of EBT3 films exposed to two separate beams of electrons and photons, for radiation doses ranging up to 500 cGy. The analysis is done on a flatbed EPSON 10000 XL scanner and cross checked on a HP Scanjet 4850 scanner. Both scanners are used in reflection mode taking into account landscape and portrait scanning positions. After thorough verifications, the reflective scanning method can be used on EBT3 as an economic alternative to the transmission method which was also one of the goals of this study. A comparison is also done between single scan configuration including all samples in a single A4 (HP) or A3 (EPSON) format area and multiple scan procedure where each sample is scanned separately on its own. The images analyses are done using the ImageJ software. Results show significant influence of the scanning configuration but no significant differences between electron and photon irradiations for both single and multiple scan configurations. In conclusion, the film provides a reliable relative dose measurement method for electrons and photons irradiations in the medical field applications.
Copyright © 2014 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Film dosimetry; Gafchromic EBT3; Radiotherapy

Mesh:

Year:  2014        PMID: 24880678     DOI: 10.1016/j.ejmp.2014.04.010

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  5 in total

1.  Measurement of percentage dose at the surface for a 6 MV photon beam.

Authors:  O O Galván De la Cruz; M A Rodríguez-Ávila; T Rivera-Montalvo; O A García Garduño
Journal:  Rep Pract Oncol Radiother       Date:  2019-10-18

2.  Gafchromic EBT3 film dosimetry in electron beams - energy dependence and improved film read-out.

Authors:  Petri Sipilä; Jarkko Ojala; Sampsa Kaijaluoto; Ilkka Jokelainen; Antti Kosunen
Journal:  J Appl Clin Med Phys       Date:  2016-01-08       Impact factor: 2.102

3.  Physical and Dosimetric Aspect of Euromechanics Add-on Multileaf Collimator on Varian Clinac 2100 C/D.

Authors:  S A Rohani; S R Mahdavi; A Mostaar; S Ueltzhöffer; R Mohammadi; Gh Geraily
Journal:  J Biomed Phys Eng       Date:  2019-02-01

4.  Skin Dosimetry with EBT3 Radiochromic Film in Radiotherapy of Parotid Cancer.

Authors:  Mohammad Taghi Bahreyni Toossi; Mahdi Ghorbani; Farideh Khorshidi; Mohammad Mohammadi; Nastaran Mohamadian; Fateme Akbari; Mahdie Dayani
Journal:  J Biomed Phys Eng       Date:  2021-10-01

5.  Development of custom lead shield and strainer for targeted irradiation for mice in the gamma cell chamber.

Authors:  Nurhaslina Hasan; Nur Fatihah Ronny Sham; Muhammad Khalis Abdul Karim; Syed Baharom Syed Ahmad Fuad; Narimah Abdul Hamid Hasani; Effat Omar; Mohammad Johari Ibahim
Journal:  Sci Rep       Date:  2021-07-15       Impact factor: 4.379

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.