Influence of nebulizer type, position, and bias flow on aerosol drug delivery in simulated pediatric and adult lung models during mechanical ventilation.

Arzu Ari1, Orcin Telli Atalay, Robert Harwood, Meryl M Sheard, Essam A Aljamhan, James B Fink.   

Abstract

BACKGROUND: The effectiveness of aerosol drug delivery during mechanical ventilation is influenced by the patient, ventilator, and nebulizer variables. The impact of nebulizer type, position on the ventilator circuit, and bias flow on aerosol drug delivery has not been established for different age populations.
OBJECTIVE: To determine the influence of nebulizer position and bias flow with a jet nebulizer and a vibrating-mesh nebulizer on aerosol drug delivery in simulated and mechanically ventilated pediatric and adult patients.
METHOD: Albuterol sulfate (2.5 mg) was nebulized with a jet nebulizer and a vibrating-mesh nebulizer, using simulated pediatric and adult lung models. The 2 nebulizer positions were: (1) jet nebulizer placed 15 cm from the Y-piece adapter, and vibrating-mesh nebulizer attached directly to the Y-piece; and (2) jet nebulizer placed prior to the heated humidifier with 15 cm of large-bore tubing, and vibrating-mesh nebulizer positioned at an inlet to the humidifier. A ventilator with a heated humidifier and ventilator circuit was utilized in both lung models. The adult ventilator settings were V(T) 500 mL, PEEP 5 cm H2O, respiratory rate 20 breaths/min, peak inspiratory flow 60 L/min, and descending ramp flow waveform. The pediatric ventilator settings were V(T) 100 mL, PEEP 5 cm H2O, respiratory rate 20 breaths/min, inspiratory time 1 s. We tested bias flows of 2 and 5 L/min. The adult and pediatric lung models used 8-mm and 5-mm inner-diameter endotracheal tubes, respectively. Each experiment was run 3 times (n = 3). The albuterol sulfate was eluted from the filter and analyzed via spectrophotometry (276 nm).
RESULTS: Nebulizer placement prior to the humidifier increased drug delivery with both the jet nebulizer and the vibrating-mesh nebulizer, with a greater increase with the vibrating-mesh nebulizer. Higher bias flow reduced drug delivery. Drug delivery with the vibrating-mesh nebulizer was 2-4-fold greater than with the jet nebulizer at all positions (P < .05) in both lung models.
CONCLUSION: During simulated mechanical ventilation in pediatric and adult models, bias flow and nebulizer type and position impact aerosol drug delivery.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20587095

Source DB:  PubMed          Journal:  Respir Care        ISSN: 0020-1324            Impact factor:   2.258


  49 in total

Review 1.  Aerosol delivery to ventilated newborn infants: historical challenges and new directions.

Authors:  Jan Mazela; Richard A Polin
Journal:  Eur J Pediatr       Date:  2010-09-28       Impact factor: 3.183

Review 2.  A path to successful patient outcomes through aerosol drug delivery to children: a narrative review.

Authors:  Arzu Ari
Journal:  Ann Transl Med       Date:  2021-04

Review 3.  Narrative review of practical aspects of aerosol delivery via high-flow nasal cannula.

Authors:  Jie Li; James B Fink
Journal:  Ann Transl Med       Date:  2021-04

4.  Iloprost drug delivery during infant conventional and high-frequency oscillatory ventilation.

Authors:  Robert M DiBlasi; Dave N Crotwell; Shuijie Shen; Jiang Zheng; James B Fink; Delphine Yung
Journal:  Pulm Circ       Date:  2016-03       Impact factor: 3.017

Review 5.  Aerosol delivery via invasive ventilation: a narrative review.

Authors:  Hui-Ling Lin; James B Fink; Huiqing Ge
Journal:  Ann Transl Med       Date:  2021-04

Review 6.  Inhaled antibiotics during mechanical ventilation-why it will work.

Authors:  Maxime Desgrouas; Stephan Ehrmann
Journal:  Ann Transl Med       Date:  2021-04

7.  Inhaled Treprostinil Drug Delivery During Mechanical Ventilation and Spontaneous Breathing Using Two Different Nebulizers.

Authors:  Donna K Parker; Shuijie Shen; Jiang Zheng; D Dunbar Ivy; Dave N Crotwell; Justin C Hotz; Robert M DiBlasi
Journal:  Pediatr Crit Care Med       Date:  2017-06       Impact factor: 3.624

8.  Pharmacokinetics of inhaled colistimethate sodium (CMS) in mechanically ventilated critically ill patients.

Authors:  Zoe E Athanassa; Sophia L Markantonis; Marina-Zoe F Fousteri; Pavlos M Myrianthefs; Eleni G Boutzouka; Athanassios Tsakris; George J Baltopoulos
Journal:  Intensive Care Med       Date:  2012-07-19       Impact factor: 17.440

9.  Pulmonary Delivery of Vancomycin Dry Powder Aerosol to Intubated Rabbits.

Authors:  Bradley P Sullivan; Nashwa El-Gendy; Christopher Kuehl; Cory Berkland
Journal:  Mol Pharm       Date:  2015-05-06       Impact factor: 4.939

10.  Optimal delivery of aerosols to infants during mechanical ventilation.

Authors:  P Worth Longest; Mandana Azimi; Michael Hindle
Journal:  J Aerosol Med Pulm Drug Deliv       Date:  2013-12-03       Impact factor: 2.849

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.