Evaluation of aerosol generator devices at 3 locations in humidified and non-humidified circuits during adult mechanical ventilation.

Arzu Ari1, Hasan Areabi, James B Fink.   

Abstract

BACKGROUND: The position of the jet or ultrasonic nebulizer in the ventilator circuit impacts drug delivery during mechanical ventilation, but has not been extensively explored, and no study has examined all of the commonly used nebulizers.
METHODS: Drug delivery from jet, vibrating-mesh, and ultrasonic nebulizers and pressurized metered-dose inhaler (pMDI) with spacer was compared in a model of adult mechanical ventilation, with heated/humidified and non-humidified ventilator circuits. Albuterol sulfate was aerosolized at 3 circuit positions: (1) between the endotracheal tube and the Y-piece; (2) 15 cm from Y-piece; and (3) 15 cm from the ventilator, with each device (n = 3) using adult settings (tidal volume 500 mL, ramp flow pattern, 15 breaths/min, peak inspiratory flow 60 L/min, and PEEP 5 cm H(2)O). The drug deposited on an absolute filter distal to an 8.0-mm inner-diameter endotracheal tube was eluted and analyzed via spectrophotometry (276 nm), and is reported as mean +/- SD percent of total nominal or emitted dose.
RESULTS: The vibrating-mesh nebulizer, ultrasonic nebulizer, and pMDI with spacer were most efficient in position 2 with both non-humidified (30.2 +/- 1.0%, 24.7 +/- 4.4%, and 27.8 +/- 3.3%, respectively) and heated/humidified circuits (16.8 +/- 2.6%, 16.5 +/- 4.3%, and 17 +/- 1.0%, respectively). In contrast, the jet nebulizer was most efficient in position 3 under both non-humidified (14.7 +/- 1.5%) and heated/humidified (6.0 +/- 0.1%) conditions. In positions 2 and 3, all devices delivered approximately 2-fold more drug under non-humidified than under heated/humidified conditions (P < .01). At position 1 only the pMDI delivered substantially more drug than with the non-humidified circuit.
CONCLUSION: During mechanical ventilation the optimal drug delivery efficiency depends on the aerosol generator, the ventilator circuit, and the aerosol generator position.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20587094

Source DB:  PubMed          Journal:  Respir Care        ISSN: 0020-1324            Impact factor:   2.258


  28 in total

1.  Iloprost drug delivery during infant conventional and high-frequency oscillatory ventilation.

Authors:  Robert M DiBlasi; Dave N Crotwell; Shuijie Shen; Jiang Zheng; James B Fink; Delphine Yung
Journal:  Pulm Circ       Date:  2016-03       Impact factor: 3.017

Review 2.  Aerosol delivery via invasive ventilation: a narrative review.

Authors:  Hui-Ling Lin; James B Fink; Huiqing Ge
Journal:  Ann Transl Med       Date:  2021-04

Review 3.  Administration of dry powders during respiratory supports.

Authors:  Wei-Ren Ke; Rachel Yoon Kyung Chang; Philip Chi Lip Kwok; Patricia Tang; Lan Chen; Donghao Chen; Hak-Kim Chan
Journal:  Ann Transl Med       Date:  2021-04

4.  Inhaled Treprostinil Drug Delivery During Mechanical Ventilation and Spontaneous Breathing Using Two Different Nebulizers.

Authors:  Donna K Parker; Shuijie Shen; Jiang Zheng; D Dunbar Ivy; Dave N Crotwell; Justin C Hotz; Robert M DiBlasi
Journal:  Pediatr Crit Care Med       Date:  2017-06       Impact factor: 3.624

5.  The importance of protecting the mechanical ventilator during colistin methanesulfonate nebulization.

Authors:  Francesco Mojoli; Giorgio Antonio Iotti; Roberto Imberti; Antonio Braschi
Journal:  Intensive Care Med       Date:  2012-12-20       Impact factor: 17.440

6.  Usefulness of a handheld nebulizer in cough test to screen for silent aspiration.

Authors:  Yoko Wakasugi; Haruka Tohara; Ayako Nakane; Shino Murata; Shinya Mikushi; Chiaki Susa; Maho Takashima; Yoshiko Umeda; Ruriko Suzuki; Hiroshi Uematsu
Journal:  Odontology       Date:  2012-10-06       Impact factor: 2.634

7.  Pharmacokinetics of inhaled colistimethate sodium (CMS) in mechanically ventilated critically ill patients.

Authors:  Zoe E Athanassa; Sophia L Markantonis; Marina-Zoe F Fousteri; Pavlos M Myrianthefs; Eleni G Boutzouka; Athanassios Tsakris; George J Baltopoulos
Journal:  Intensive Care Med       Date:  2012-07-19       Impact factor: 17.440

Review 8.  Inhaled Antimicrobials for Ventilator-Associated Pneumonia: Practical Aspects.

Authors:  Garyphallia Poulakou; Dimitrios K Matthaiou; David P Nicolau; Georgios Siakallis; George Dimopoulos
Journal:  Drugs       Date:  2017-09       Impact factor: 9.546

9.  E-cigarette aerosol collection using converging and straight tubing Sections: Physical mechanisms.

Authors:  Markus Hilpert; Vesna Ilievski; Shao-Yiu Hsu; Ana M Rule; Pablo Olmedo; German Drazer
Journal:  J Colloid Interface Sci       Date:  2020-10-12       Impact factor: 8.128

10.  Guidance for nebulization during the COVID-19 pandemic.

Authors:  Rajesh Swarnakar; Neeraj M Gupta; Indranil Halder; Gopi C Khilnani
Journal:  Lung India       Date:  2021-03
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.