Literature DB >> 34206675

Performance of a Handheld Ultrasound Device to Assess Articular and Periarticular Pathologies in Patients with Inflammatory Arthritis.

Giulia Corte1,2, Sara Bayat1,2, Koray Tascilar1,2, Larissa Valor-Mendez1,2, Louis Schuster1,2, Johannes Knitza1,2, Filippo Fagni1,2, Georg Schett1,2, Arnd Kleyer1,2, David Simon1,2.   

Abstract

The purpose of this study was to assess the accuracy and performance of a new handheld ultrasound (HHUS) machine in comparison to a conventional cart-based sonographic machine in patients with inflammatory arthritis (IA). IA patients with at least one tender and swollen joint count were enrolled. US was performed on the clinically affected joints using a cart-based sonographic device (Samsung HS40) and a HHUS device (Butterfly iQ). One blinded reader scored all images for the presence of erosions, bony enlargement, synovial hypertrophy, joint effusion, bursitis, tenosynovitis, and enthesitis. Synovitis was graded (B mode and power Doppler (PD)) by the 4-level EULAR-OMERACT scale. To avoid bias by the blinded reader, we included 67 joints of two healthy volunteers in the evaluation. We calculated the overall concordance and the concordance by type of joint and pathological finding. We also measured the time required for the US examination per joint with both devices. Thirty-two patients (20 with RA, 10 with PsA, and one each with gout and SLE-associated arthritis) were included, and 186 joints were examined. The overall raw concordance in B mode was 97% (κappa 0.90, 95% CI (0.89, 0.94)). In B mode, no significant differences were found in relation to type of joint or pathological finding examined. The PD mode of the HHUS device did not detect any PD signal, whereas the cart-based device detected a PD signal in 61 joints (33%). The portable device did not offer any time savings compared to the cart-based device (47.0 versus 46.3 s). The HHUS device was accurate in the assessment of structural damage and inflammation in patients with IA, but only in the B mode. Significant improvements are still needed for HHUS to reliably demonstrate blood flow detection in PD mode.

Entities:  

Keywords:  handheld ultrasound device; inflammatory arthritis; medical ultrasound; musculoskeletal ultrasound

Year:  2021        PMID: 34206675     DOI: 10.3390/diagnostics11071139

Source DB:  PubMed          Journal:  Diagnostics (Basel)        ISSN: 2075-4418


  27 in total

1.  Guidelines for musculoskeletal ultrasound in rheumatology.

Authors:  M Backhaus; G R Burmester; T Gerber; W Grassi; K P Machold; W A Swen; R J Wakefield; B Manger
Journal:  Ann Rheum Dis       Date:  2001-07       Impact factor: 19.103

2.  Prospective two year follow up study comparing novel and conventional imaging procedures in patients with arthritic finger joints.

Authors:  M Backhaus; G R Burmester; D Sandrock; D Loreck; D Hess; A Scholz; S Blind; B Hamm; M Bollow
Journal:  Ann Rheum Dis       Date:  2002-10       Impact factor: 19.103

3.  Portable ultrasonography in mass casualty incidents: The CAVEAT examination.

Authors:  Stanislaw Peter Stawicki; James M Howard; John P Pryor; David P Bahner; Melissa L Whitmill; Anthony J Dean
Journal:  World J Orthop       Date:  2010-11-18

4.  Current state of musculoskeletal ultrasound training and implementation in Europe: results of a survey of experts and scientific societies.

Authors:  Esperanza Naredo; Maria A D'Agostino; Philip G Conaghan; Marina Backhaus; Peter Balint; George A W Bruyn; Emilio Filippucci; Walter Grassi; Hilde B Hammer; Annamaria Iagnocco; David Kane; Juhani M Koski; Marcin Szkudlarek; Lene Terslev; Richard J Wakefield; Hans-Rudolf Ziswiler; Wolfgang A Schmidt
Journal:  Rheumatology (Oxford)       Date:  2010-09-13       Impact factor: 7.580

5.  Enhancement of Musculoskeletal Radiology Resident Education with the Use of an Individual Smart Portable Ultrasound Device (iSPUD).

Authors:  Troy H Maetani; Cody Schwartz; Robert J Ward; Daniel B Nissman
Journal:  Acad Radiol       Date:  2018-07-29       Impact factor: 3.173

6.  Pocket-size point-of-care ultrasound in rural Uganda - A unique opportunity "to see", where no imaging facilities are available.

Authors:  Danny Epstein; Neta Petersiel; Erez Klein; Erez Marcusohn; Eyal Aviran; Reut Harel; Zaher S Azzam; Ami Neuberger; Lior Fuchs
Journal:  Travel Med Infect Dis       Date:  2018-01-06       Impact factor: 6.211

7.  Pocket ultrasound device as a complement to physical examination for ascites evaluation and guided paracentesis.

Authors:  Daniel Keil-Ríos; Hiram Terrazas-Solís; Alejandro González-Garay; Juan Francisco Sánchez-Ávila; Ignacio García-Juárez
Journal:  Intern Emerg Med       Date:  2016-02-19       Impact factor: 3.397

8.  Longitudinal power Doppler ultrasonographic assessment of joint inflammatory activity in early rheumatoid arthritis: predictive value in disease activity and radiologic progression.

Authors:  Esperanza Naredo; Paz Collado; Ana Cruz; Mercedes J Palop; Félix Cabero; Patricia Richi; Loreto Carmona; Manuel Crespo
Journal:  Arthritis Rheum       Date:  2007-02-15

9.  A novel ultrasonographic synovitis scoring system suitable for analyzing finger joint inflammation in rheumatoid arthritis.

Authors:  Alexander K Scheel; Kay-Geert A Hermann; Elke Kahler; Daniel Pasewaldt; Jacqueline Fritz; Bernd Hamm; Edgar Brunner; Gerhard A Müller; Gerd R Burmester; Marina Backhaus
Journal:  Arthritis Rheum       Date:  2005-03

Review 10.  2019 European League Against Rheumatism/American College of Rheumatology classification criteria for systemic lupus erythematosus.

Authors:  Martin Aringer; Karen Costenbader; David Daikh; Ralph Brinks; Marta Mosca; Rosalind Ramsey-Goldman; Josef S Smolen; David Wofsy; Dimitrios T Boumpas; Diane L Kamen; David Jayne; Ricard Cervera; Nathalie Costedoat-Chalumeau; Betty Diamond; Dafna D Gladman; Bevra Hahn; Falk Hiepe; Søren Jacobsen; Dinesh Khanna; Kirsten Lerstrøm; Elena Massarotti; Joseph McCune; Guillermo Ruiz-Irastorza; Jorge Sanchez-Guerrero; Matthias Schneider; Murray Urowitz; George Bertsias; Bimba F Hoyer; Nicolai Leuchten; Chiara Tani; Sara K Tedeschi; Zahi Touma; Gabriela Schmajuk; Branimir Anic; Florence Assan; Tak Mao Chan; Ann Elaine Clarke; Mary K Crow; László Czirják; Andrea Doria; Winfried Graninger; Bernadett Halda-Kiss; Sarfaraz Hasni; Peter M Izmirly; Michelle Jung; Gábor Kumánovics; Xavier Mariette; Ivan Padjen; José M Pego-Reigosa; Juanita Romero-Diaz; Íñigo Rúa-Figueroa Fernández; Raphaèle Seror; Georg H Stummvoll; Yoshiya Tanaka; Maria G Tektonidou; Carlos Vasconcelos; Edward M Vital; Daniel J Wallace; Sule Yavuz; Pier Luigi Meroni; Marvin J Fritzler; Ray Naden; Thomas Dörner; Sindhu R Johnson
Journal:  Ann Rheum Dis       Date:  2019-08-05       Impact factor: 27.973

View more
  2 in total

1.  Comparison of four handheld point-of-care ultrasound devices by expert users.

Authors:  Minh-Phuong T Le; Lara Voigt; Robert Nathanson; Anna M Maw; Gordon Johnson; Ria Dancel; Benji Mathews; Alvaro Moreira; Harald Sauthoff; Christopher Gelabert; Linda M Kurian; Jenna Dumovich; Kevin C Proud; Jessica Solis-McCarthy; Carolina Candotti; Christopher Dayton; Alexander Arena; Brandon Boesch; Saul Flores; Mark T Foster; Nicholas Villalobos; Tanping Wong; Gabriel Ortiz-Jaimes; Michael Mader; Craig Sisson; Nilam J Soni
Journal:  Ultrasound J       Date:  2022-07-07

2.  Abdominal aorta measurements by a handheld ultrasound device compared with a conventional cart-based ultrasound machine.

Authors:  Abdulrahman M Alfuraih; Abdulaziz I Alrashed; Saleh O Almazyad; Mohammed J Alsaadi
Journal:  Ann Saudi Med       Date:  2021-12-02       Impact factor: 1.526

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.