Friederike Nowotny1, Julian Schmidberger1, Patrycja Schlingeloff1, Andreas Binzberger1, Wolfgang Kratzer2.
Abstract
BACKGROUND: Few systematic comparative studies of the different methods of physical elastography of the spleen are currently available. AIM: To compare point shear wave and two-dimensional elastography of the spleen considering the anatomical location (upper, hilar, and lower pole).Entities:
Keywords: Acoustic radiation force impulse; Elastography; Healthy subjects; Point shear wave elastography; Spleen; Two-dimensional shear-wave elastography; Ultrasonography
Year: 2021 PMID: 34141093 PMCID: PMC8188838 DOI: 10.4329/wjr.v13.i5.137
Source DB: PubMed Journal: World J Radiol ISSN: 1949-8470
Figure 1Flow chart of study arm inclusion and exclusion.
Characteristics of subjects included in study arms A and B and their overlap, n (%)
|
|
|
| ||||
|
|
|
|
|
|
| |
| Gender | ||||||
| Male | 78 (39.00) | 38 (30.6) | 10 (22.73) | |||
| Female | 122 (61.00) | 75 (66.4) | 34 (77.27) | |||
| Age | 27.93 ± 8.13 | 25.95 ± 6.26 | 27.11 ± 5.64 | |||
| < 30 yr | 154 (77.00) | 98 (86.7) | 34 (77.27) | |||
| ≥ 30 yr | 46 (23.00) | 15 (13.3) | 10 (22.73) | |||
| BMI | 22.56 ± 2.57 | 21.64 ± 2.24 | 21.60 ± 2.67 | |||
| BMI < 25 | 166 (83.0) | 103 (91.2) | 38 (86.36) | |||
| BMI ≥25 | 34 (17.00) | 10 (8.9) | 6 (13.64) | |||
| Alcohol consumption | 8.17 ± 9.70 | 6.73 ± 8.35 | 6.22 ± 7.95 | |||
| None | 18 (9.00) | 9 (8.0) | 9 (20.45) | |||
| Less 1/mo | 28 (14.00) | 17 (15.0) | 3 (6.82) | |||
| Several times per month | 106 (53.00) | 71 (62.8) | 28 (63.64) | |||
| Several times per week | 48 (24.00) | 16 (14.2) | 4 (9.09) | |||
| Fasting time | 3.74 ± 1.84 | 4.47 ± 2.80 | 4.11 ± 1.65 | |||
| Spleen length (mm) | 105.44 ± 14.80 | 105.53 ± 14.16 | 101.66 ± 13.78 | |||
| Median (min-max) | 105.50 (61-144) | 106.00 (65-144) | 100.50 (75-137) | |||
| Spleen depth (mm) | 35.38 ± 6.19 | 35.28 ± 6.38 | 34.25 ± 5.59 | |||
| Median (min-max) | 35.50 (21-56) | 35.00 (21-56) | 34.00 (24-48) | |||
BMI: Body mass index.
Figure 2Illustration of the examination of the spleen. A: B-Mode Ultrasound image of the spleen; B: Upper spleen pole; C: Middle spleen pole; and D: Lower spleen pole.
Location and dispersion measures of shear wave velocity at the spleen measured with the Toshiba Aplio 500 version 5.0 and version 6.0, Siemens S3000, GE Logiq E9, and Philips Epiq 7 devices, n (%)
|
|
| |||||
|
|
|
|
|
|
|
|
| Toshiba Aplio 500, mean ± SD | 3.35 ± 0.92 | 2.90 ± 0.45 | 3.38 ± 0.65 | 2.34 ± 0.29 | 2.48 ± 0.26 | 2.48 ± 0.28 |
| Median (Min-Max) | 3.12 (2.20-6.94) | 2.84 (1.94-4.86) | 3.36 (1.50-5.92) | 2.36 (1.42-3.36) | 2.48 (1.82-3.02) | 2.45 (1.84-3.25) |
| Siemens S3000, mean ± SD | 2.05 ± 0.54 | 2.53 ± 0.44 | 2.53 ± 0.58 | 2.39 ± 0.33 | 2.63 ± 0.28 | 2.49 ± 0.34 |
| Median (Min-Max) | 2.04 (0.73-3.77) | 2.49 (1.27-3.73) | 2.47 (1.39-4.59) | 2.39 (1.52-3.86) | 2.62 (1.97-3.36) | 2.50 (1.53-3.58) |
| GE Logiq E9, mean ± SD | 2.20 ± 0.51 | 1.86 ± 0.44 | 1.53 ± 0.44 | - | - | - |
| Median (Min-Max) | 2.23 (1.10-6.26) | 1.88 (0.71-3.02) | 1.48 (0.77-2.84) | |||
| Philips Epiq 7, mean ± SD | 1.88 ± 0.40 | 1.89 ± 0.38 | 2.30 ± 0.87 | - | - | - |
| Median (Min-Max) | 1.90 (0.88-3.10) | 1.91 (0.93-3.12) | 2.17 (0.99-10.62) | |||
Position and stress measurements of shear wave velocities measured with the Toshiba Aplio 500 version 5.0 and version 6.0
|
|
|
|
|
| Toshiba Aplio 500 V.5.0, mean ± SD | 3.46 ± 0.52 | 2.94 ± 0.52 | 3.38 ± 1.02 |
| Median (Min-Max) | 3.45 (2.54-5.13) | 2.80 (2.27-4.86) | 3.14 (2.25-6.53) |
| Toshiba Aplio 500 V.6.0, mean ± SD | 2.32 ± 0.34 | 2.53 ± 0.25 | 2.52 ± 0.29 |
| Median (Min-Max) | 2.36 (1.42-3.36) | 2.56 (1.85-2.94) | 2.46 (1.93-3.25) |
Correlation of the heavy wave velocities of the Toshiba Aplio 500 version 5.0 and version 6.0, GE Logiq E9, and Philips Epiq 7 devices with the Siemens S3000
|
|
|
| |||
|
|
|
|
| ||
|
|
|
|
|
|
|
| Toshiba Aplio 500 | Lower pole | -0.08792 | 0.2193 | 0.19863 | 0.0375 |
| Middle pole | 0.13632 | 0.0561 | 0.13438 | 0.1616 | |
| Upper pole | 0.03951 | 0.5815 | 0.24951 | 0.0086 | |
| GE Logiq E9 | Lower pole | -0.03307 | 0.6446 | - | - |
| Middle pole | -0.04941 | 0.4905 | - | - | |
| Upper pole | 0.04744 | 0.5079 | - | - | |
| Philips Epiq 7 | Lower pole | 0.04894 | 0.4947 | - | - |
| Middle pole | 0.12321 | 0.0845 | - | - | |
| Upper pole | 0.33088 | < 0.0001 | - | - | |
Figure 3Boxplot diagram illustrating the measurements on the Toshiba Aplio 400 version 6.0 and Siemens devices for the different splenic sections. ARFI: Acoustic radiation force impulse.