Ana I Barbosa1,2, Alexander D Edwards2,3, Nuno M Reis2,4. 1. Department of Chemical Engineering, Loughborough University, Loughborough LE11 3TU, United Kingdom. 2. Capillary Film Technology Ltd, Daux Road, Billingshurst RH14 9SJ, West Sussex, United Kingdom. 3. Reading School of Pharmacy, University of Reading, Whiteknights, Reading RG6 6AD, United Kingdom. 4. Department of Chemical Engineering and Centre for Biosensors, Bioelectronics and Biodevices (C3Bio), University of Bath, Claverton Down, Bath BA2 7AY, United Kingdom.
Abstract
The performance of biosensors is often optimized in buffers, which brings inconsistencies during applications with biological samples. Current strategies for minimizing sample (matrix) interference are complex to automate and miniaturize, involving, e.g., sample dilution or recovery of serum/plasma. This study shows the first systematic analysis using hundreds of actual microfluidic immunoassay fluoropolymer strips to understand matrix interference in microflow systems. As many interfering factors are assay-specific, we have explored matrix interference for a range of enzymatic immunoassays, including a direct mIgG/anti-mIgG, a sandwich cancer biomarker PSA, and a sandwich inflammatory cytokine IL-1β. Serum matrix interference was significantly affected by capillary antibody surface coverage, suggesting for the first time that the main cause of the serum matrix effect is low-affinity serum components (e.g., autoantibodies) competing with high-affinity antigens for the immobilized antibody. Additional experiments carried out with different capillary diameters confirmed the importance of antibody surface coverage in managing matrix interference. Building on these findings, we propose a novel analytical approach where antibody surface coverage and sample incubation times are key for eliminating and/or minimizing serum matrix interference, consisting in bioassay optimization carried out in serum instead of buffer, without compromising the performance of the bioassay or adding extra cost or steps. This will help establishing a new route toward faster development of modern point-of-care tests and effective biosensor development.
The performance of biosensors is often optimized in buffers, which brings inconsistencies during applications with biological samples. Current strategies for minimizing sample (matrix) interference are complex to automate and miniaturize, involving, e.g., sample dilution or recovery of serum/plasma. This study shows the first systematic analysis using hundreds of actual microfluidic immunoassay fluoropolymer strips to understand matrix interference in microflow systems. As many interfering factors are assay-specific, we have explored matrix interference for a range of enzymatic immunoassays, including a direct mIgG/anti-mIgG, a sandwich cancer biomarker PSA, and a sandwich inflammatory cytokine IL-1β. Serum matrix interference was significantly affected by capillary antibody surface coverage, suggesting for the first time that the main cause of the serum matrix effect is low-affinity serum components (e.g., autoantibodies) competing with high-affinity antigens for the immobilized antibody. Additional experiments carried out with different capillary diameters confirmed the importance of antibody surface coverage in managing matrix interference. Building on these findings, we propose a novel analytical approach where antibody surface coverage and sample incubation times are key for eliminating and/or minimizing serum matrix interference, consisting in bioassay optimization carried out in serum instead of buffer, without compromising the performance of the bioassay or adding extra cost or steps. This will help establishing a new route toward faster development of modern point-of-care tests and effective biosensor development.
Entities:
Keywords:
biosensors; matrix effect; microcapillary film; microfluidics; protein biomarkers
Components
of biological samples
are known to interfere with the performance of diagnostics tests,
by affecting the response of the test to the analyte of interest.[1] This has a direct impact on sensitivity, specificity,
and variability of the test, leading to inaccurate analyte quantitation
in real biological samples.[2,3] According to the current
literature, this so-called matrix interference or effect can be caused
by different components; blood cells, sample viscosity, or plasma
components such as heterophilic antibodies (antibodies produced against
poorly defined antigens presenting low affinity and weak binding),[4] human antianimal antibodies (HAAA, high-affinity
antibodies generated when the immune system is in contact with animal
antibodies),[5] and other plasma proteins
such as albumin, lysozyme, fibrinogen, and paraprotein have been reported
to cause test interferences.[6] Boscato et
al. showed that analyte–antibody binding substances were detected
in 40% of studied samples (688 samples), being responsible for 15%
interference in assays.[7] Appropriate matrix
management is therefore essential to develop reliable bioassays and
biosensors; however, this is highly dependent on the molecular analysis
platform since the type of reagents (e.g., antibody purity) and the
antibody binding conditions (e.g., antibody affinity, diffusion distance,
surface interactions) are key contributors to the matrix effect. Although
current procedures for dealing with matrix interference can be effectively
implemented in a centralized pathology lab, involving conventional
sample preparation methods such as dilution, centrifugation, precipitation,
etc., these methods are not universal and fail to serve sensitive
and automated detection desired in portable point-of-care (POC) microfluidic
platforms.[8] Currently, little is known
in the literature about matrix interference in microfluidic systems,
which needs to be addressed to speed up the adoption of microfluidic
bioassays and biosensors.To find a universal way to deal with
the matrix effect at POC settings,
there are a plethora of microfluidic plasma separation devices aiming
to eliminate sample matrix interference in protein bioassays performed
by novel biosensor platforms.[9,10] However, plasma or
serum still contains interfering factors, which affect the accuracy
of the tests.[11] POC analytical approaches
would greatly benefit from overcoming biological matrix interference
without laboratory equipment, since any sample preparation required
for a POC test compromises the speed, complexity, and cost of the
test. Therefore, understanding the biological matrix interference
and finding strategies to overcome it are paramount for the POC diagnostic
industry and biosensor research,[12] which
aim to combine sensitive, accurate, and rapid protein quantitation
with cost-effective test development, demanded by the ever-increasing
biomarker use in patients’ stratification and personalized
medicine.[3,8,13,14] Many biosensors are incorporated into lab-on-a-chip
devices that test plasma or serum separated outside the microdevice
using centrifugation, reducing the benefits of miniaturization.[15] A growing number of microfluidic strategies
aim to incorporate in situ plasma separation from
whole blood[14] using microstructures,[16] gravity-driven separation,[17] microcentrifugation,[18] capillary-driven
contactless electrophoresis,[19] and the
plasma skimming effect sometimes referred to as the Zweifach–Fung
effect.[20,21] However, very few studies reported the measurement
of protein biomarkers after the blood plasma separation, which hinders
the validation of the developed devices and methods for protein biomarker
quantitation. Furthermore, the microfluidic studies that actually
report protein detection in plasma[22,23] do not report
recovery or sample variability studies, hampering the understanding
of how blood or plasma sample matrix affects protein biomarker detection
in microfluidic devices and consequently how to solve the sample variability
effect. In fact, data that reflects how sample components affect antibody–antigen
binding in a specific microfluidic device can be difficult to obtain,
not only due to the variety of interference factors but also due to
the prototype nature of microfluidic devices that are not manufactured
on a large scale, reducing the number of replicates needed for the
study. Several studies use real-time antibody–antigen detection
techniques such as optical waveguide lightmode spectroscopy (OWLS),
ellipsometry, or quartz crystal microbalance (QCM) that, although
very precise for antibody binding affinity determination, use polymer-coated
specific surfaces that not always replicate the surface chemistry
of the actual microfluidic devices. Also, these systems do not reflect
the geometry of the microfluidic devices, which can lead to errors
when translating assay conditions from real-time detection technique
to microfluidic systems.[24,25]In the present
work, we explored matrix interference in microfluidic
protein immunoassays using hundreds of fluorinated, Teflon FEP microfluidic
strips fabricated from a melt-extruded, mass-manufactured 10-bore
microcapillary film (MCF), connected to a multiple syringe aspirator
developed in-house (Figure A). Microfluidic protein bioassays presented significant variations
when performed in buffer or human serum (Figure B), confirming that matrix interference is
also present in microfluidic bioassays. Based on our previous experience
in carrying out high-performance immunoassays in this microfluidic
platform,[3,26] we hypothesized that the actuation mechanism
of the interfering factor(s) (Figure C) is closely related to the antibody surface coverage.
Consequently, in this study we explored the impact of antibody surface
coverage on sample matrix interference for three distinct protein
bioassays, as interference can be very assay-specific.[27,28] In addition, we studied other parameters that appear to contribute
to the matrix interference, with a particular focus on sample incubation
time and capillary diameter. We gathered the outcomes into a new bioanalytical
approach for minimizing matrix interference in immunoassay protein
detection.
Figure 1
Human serum matrix effect in MCF diagnostic strips. (A) MCF and
the fluid handling setup for diagnostic procedures. (B) PSA sandwich
assay full response curves in human serum and buffer, showing the
matrix effect interference. (C) Schematic of the capillary immunoassays
in the MCF platform.
Human serum matrix effect in MCF diagnostic strips. (A) MCF and
the fluid handling setup for diagnostic procedures. (B) PSA sandwich
assay full response curves in human serum and buffer, showing the
matrix effect interference. (C) Schematic of the capillary immunoassays
in the MCF platform.
Experimental
Section
Materials and Reagents
Mouse IgG (mIgG, whole antibody)
was purchased from Life Technologies (Paisley, U.K.); rabbit anti-mIgG
(whole molecule) conjugated with peroxidase and SIGMAFAST OPD (o-phenylenediamine dihydrochloride) tablets was purchased
from Sigma-Aldrich (Dorset, U.K.). The IL-1β recombinant protein,
anti-Human IL-1β biotin, and anti-Human IL-1β (purified)
were obtained from eBiosciences (Hatfield, U.K.). High-sensitivity
streptavidin-HRP was supplied by Thermo Scientific (Lutterworth, U.K.).
Human kallikrein 3/prostate specific antigen (PSA) ELISA kit was purchased
from R&D Systems (Minneapolis). The kit contained a monoclonal
mouse Human Kallikrein 3/PSA antibody (CapAb), a Human Kallikrein
3/PSA polyclonal biotinylated antibody (DetAb), and recombinant Human
Kallikrein 3/PSA (standard). Phosphate-buffered solution (PBS) and
bovine serum albumin (BSA) were sourced from Sigma-Aldrich (Dorset,
U.K.). PBS, pH 7.4, 10 mM was used as the main experimental buffer.
The blocking solutions consisted of 3% w/v protease-free BSA diluted
in PBS buffer and a SuperBlock blocking solution purchased from Thermo
Scientific (Lutterworth, U.K.). For washings, PBS with 0.05% v/v Tween-20
(Sigma-Aldrich, Dorset, U.K.) was used. A female human serum sample
was supplied by BBI solutions (Cardiff, U.K.), aliquot, and stored
at −20 °C. Human blood, supplied by healthy volunteers
at Loughborough University, was collected into a 5 mL vial with citrate
phosphate dextrose (CPD) as the anticoagulant.
Microfluidic Fluoropolymer
MCF Strips
The microengineered
MCF material (materials and geometry detailed in the Supporting Information, SI)[29,30] is particularly
well suited to study systematically the role of the sample matrix
on heterogeneous immunoassays, enabling simple and rapid manufacturing
of hundreds or thousands of disposable strips under very identical
conditions at a negligible cost, which would be hard to match with
other microfluidic devices. Also, the whole inner section of the cylindrical/elliptical
capillaries is homogeneously coated with the capture antibody in contrast
to immobilization on a single surface as it happens for many other
microfluidic devices, which offers advantages in studying surface
coverage and specific/nonspecific surface binding.[31]
Effect of Antibody Surface Coverage
To understand how
antibody surface coverage influences human serum interference in MCF
protein tests, three different assays (mIgG/anti-mIgG, IL-1β,
and PSA assay), presenting different analytical antibodies, were studied
in a 10-bore, 212 μm mean internal diameter MCF. The antibody
surface coverage of these assays was varied by loading captured antibody
solutions in the range of 0–200 μg/mL. The antigen concentration
was kept constant in the three assays, being 0.6 μg/mL, 0.125
ng/mL, and 3.75 ng/mL for peroxidase-conjugated anti-mIgG, IL-1β,
and PSA, respectively, as well as the antigen/sample incubation time
was fixed at 5 min. IL-1β and PSA assays follow the same conditions
as previously reported[3,32] and briefly described in the SI. Digital images of MCF strips in the three
studied assays were taken after 5 min of OPD enzymatic substrate loading.
The described assays were performed in the exact same conditions preparing
antigen solutions in buffer and in nondiluted human serum (refer to
the SI document for more details).
Effect
of Sample Incubation Time
To better understand
the sample incubation effect in the matrix interference in MCF protein
assays, an IL-1β sandwich assay[32] was performed in nondiluted human serum, whole blood, and buffer.
Instead of full response curves, only four IL-1β concentrations
were tested (0, 0.03, 0.125, and 0.5 ng/mL). The sandwich assays were
performed with 5 and 30 min of sample incubation. To plot the IL-1β
full response curve, a solution of 40 μg/mL of anti-IL-1β
CapAb was used as the coating solution in a 212 ± 16 μm
diameter MCF, and 1:2 serial dilutions of 0–1 ng/mL range of
IL-1β were spiked in buffer, 50% serum, and 100% serum as sample
diluents. The samples were incubated for 5 and 30 min. The 4-parameter
logistic (4PL) mathematical model was fitted to the experimental data
by the minimum square difference for each full IL-1β response
curve. The lower limit of detection was calculated by the mean absorbance
of the blank plus three times the standard deviation of the blank
samples.
Effect of Capillary Diameter
Several transversal sections
of three FEP MCFs with different capillary diameters were trimmed,
and a long focal distant point microscope (Nikon SMZ 1500 stereo microscope)
was used for imaging. ImageJ software (NIH, Maryland) was used to
measure the diameter of the 10 capillaries from the microphotographs.[33] A solution of 200 μg/mL of mIgG was filled
into MCF strips of three different diameters (109, 212, and 375 μm)
with 35 cm length each. A negative control strip was filled with PBS
buffer. The solutions were incubated for 30 min at room temperature
and washed with 1 mL of PBS-Tween. A solution of 0.6 μg/mL peroxidase-conjugated
mouse anti-mIgG, prepared in PBS buffer, was added to the MCF strip
and 4 cm long strips were trimmed and individually washed with PBS-Tween
after variable incubation times of anti-mIgG. OPD substrate (1 mg/mL)
was added to the strips, and digital images were taken with a flatbed
scanner after 5 min of enzymatic substrate incubation. The procedure
was repeated for 0.6 μg/mL peroxidase-conjugated anti-mIgG solutions
prepared in nondiluted human serum.
Kinetics of Antibody–Antigen
Binding
Equation was solved analytically
for a constant analyte concentration and used to estimate the rates
of association and dissociation of antibody binding in the MCF system.[34]where Abs
is the optical absorbance signal
corresponding to the antigen surface density at time t; C is the antigen bulk concentration; Kon is the association rate and Koff is the dissociation rate; and Absmax is the
maximum Abs signal corresponding to the maximum antigen surface coverage.
Image Analysis of the Microfluidic MCF Strips
RGB digital
images of the immunoassay strips were split into three separated channel
images by ImageJ software (NIH, Maryland). The blue channel images
were used to calculate Abs values, based on the gray-scale peak height
of each individual capillary of Teflon FEP MCF, as described previously.[3,29,35] Therefore, the absorbance signal
is calculated for each capillary, according to the Beer–Lambert
equation. The absorbance values presented averages of absorbance from
10 capillaries in a given MCF strip.
Results and Discussion
Matrix
Interference is Linked to Antibody Surface Coverage
It has
been previously shown that antibody surface coverage is
related to immobilized antibodies’ functionality since it interferes
with their orientation and steric hindrance.[31] Therefore, we explored the impact of an antibody monolayer on the
serum matrix effect, since it would favor the binding of high-affinity
components—antigens. As matrix interference is usually dependent
on diagnostic antibodies,[8] we have tested
three different immunoassays: a direct mouse IgG/anti-mouse IgG, a
sandwich human PSA, and a sandwich human IL-1β, covering a range
of high-performance immunoassays. We manipulated the antibody surface
coverage by varying the concentration of the capture antibody loaded
into the microcapillaries, with absorbance responses tested in both
buffer and undiluted serum. Surprisingly, we noticed full agreement
of optical signals between buffer and undiluted human serum for a
narrow range of concentrations of capture antibody (Figure ), with the window of agreement
being very immunoassay-specific.
Figure 2
Effect of antibody surface coverage on
the matrix effect of human
serum of three different MCF protein assays. (A) Effect of human serum
on anti-mIgG detection using a range of 0–200 μg/mL of
capture antibody loading. Antigen concentration was kept constant,
anti-mIgG = 0.6 μg/mL. (B) Effect of human serum on IL-1β
detection using a range of 0–200 μg/mL of capture antibody
loading. Antigen concentration was kept constant, IL-1β = 0.13
ng/mL. (C) Effect of human serum on PSA using a range of 0–200
μg/mL of capture antibody loading. Antigen concentration was
kept constant, PSA = 3.75 ng/mL. (i) Assay schematics; (ii) shows
the assay signal in buffer and serum, while (ii) shows the ratio of
the two signals. The red dashed line indicates the limit of 20% variation
above which the variability is not acceptable for immunoassay performance.
Effect of antibody surface coverage on
the matrix effect of human
serum of three different MCF protein assays. (A) Effect of human serum
on anti-mIgG detection using a range of 0–200 μg/mL of
capture antibody loading. Antigen concentration was kept constant,
anti-mIgG = 0.6 μg/mL. (B) Effect of human serum on IL-1β
detection using a range of 0–200 μg/mL of capture antibody
loading. Antigen concentration was kept constant, IL-1β = 0.13
ng/mL. (C) Effect of human serum on PSA using a range of 0–200
μg/mL of capture antibody loading. Antigen concentration was
kept constant, PSA = 3.75 ng/mL. (i) Assay schematics; (ii) shows
the assay signal in buffer and serum, while (ii) shows the ratio of
the two signals. The red dashed line indicates the limit of 20% variation
above which the variability is not acceptable for immunoassay performance.For the mIgG/anti-mIgG immunoassay (Figure A), where both antibodies are
polyclonal
and do not present site-specific binding, larger antibody surface
coverages obtained from 50 to 200 μg/mL mIgG fully eliminated
the matrix interference in undiluted serum. Similar results were observed
for the IL-1β sandwich assay, where matrix interference was
fully eliminated with antibody surface coverages in the range of 40–200
μg/mL (Figure B). Based on a previous FEP adsorption study in the same MCF material,[31] it is known that these CapAb concentrations
promote the formation of half of the antibody monolayer with antibodies
oriented “end-on” with Fab regions in line, enhancing
antigen capture in microcapillary bioassays. This agrees with findings
in the literature for a thyroxin assay, where the replacement of an
antibody coverage with low affinity by high affinity eliminated the
matrix effect of serum samples,[36] explained
by the low-affinity binding of the interfering factor(s) to the immobilized
antibody. Consequently, higher antibody coverages with properly oriented
antibodies present higher antigen-binding capacity, minimizing sample
matrix interference. This is in line with conclusions in another study
that reported that matrix proteins bind nonspecifically to the immobilized
receptors in IL-6 and acute phase protein (PCT) immunoassays, however
not preventing the analyte binding.[37]The sandwich PSA (Figure C), where the immobilized antibody is monoclonal and the detection
antibody polyclonal, showed a contrasting behavior, with matrix interference
minimized for a narrow window of concentrations (10–20 μg/mL)
of the capture antibody, which is significantly lower than for the
other antibody systems shown in Figure A,B. The polyclonal anti-PSA detection antibody binds
directly to the monoclonal CapAb in the absence of the antigen; therefore,
an increment in CapAb promotes a higher increment of the signal in
buffer than in serum, suggesting a competition of the DetAb with the
interfering components from the serum.These results mean that
the narrow window for anti-PSA loading
will reduce the assay limit of detection, as sensitivity is linked
to the degree of antigen capture, which relates to a higher amount
of functionalized antibodies on the surface. Nevertheless, this assay
presented the necessary sensitivity for its application, since the
PSA clinical threshold is 4 ng/mL.[3] It
is also important to note that optimizing the capture antibody loading
in buffer could lead to significant errors in terms of assay performance.
In comparison, the IL-1β assay composed of two monoclonal antibodies,
showed improved limit of detection since the antibodies are less prone
to interference, which is coherent with the general knowledge that
assay performance is dependent on antibody nature.[32]
Matrix Interference is Time-Dependent
Longer sample
incubation times increase the probability of lower-affinity components
to be desorbed and higher-affinity compounds to be bound. In line
with our previous experience with the PSA sandwich immunoassay,[3] where we found a significant impact of sample
incubation time on the matrix interference using both whole blood
and serum, we have further studied the effect of sample incubation
time using a monoclonal pair sandwich assay system. Therefore, we
have separately fully tested the effect of sample incubation time
and different sample diluents for monoclonal quantitation of IL-1β
(Figure ). Human serum
matrix interference was fully eliminated by extending the sample incubation
from 5 to 30 min. For whole blood, matrix interference was mostly
eliminated for the range of antigen concentrations tested, only with
the negative control showing an increase in the background signal
(Figure A,B). This
is undesirable as it impacts the overall limit of detection, yet it
can very possibly be addressed through straightforward assay development,
such as optimization of the buffer and blocking solutions. Overall,
the response curves shown in Figure C,D agreed with previous studies with the same PSA
sandwich immunoassay[3] and demonstrated
that adequate sample incubation time needs to be combined with suitable
antibody surface coverage for minimizing the matrix effect in microcapillary
assays. These findings suggest that the matrix interference is time-dependent
and very probably linked to a competition for binding sites between
low-affinity interfering factor(s) with high-affinity antigens and/or
the detection antibody/complex.
Figure 3
Effect of sample incubation time on IL-1β
sandwich MCF immunoassays.
(A) IL-1β sandwich assay in buffer, whole blood, and serum,
considering 5 min sample incubation time and 0.125 ng/mL IL-1β.
(B) IL-1β sandwich assay in buffer, whole blood, and serum,
considering 30 min sample incubation time and 0.125 ng/mL IL-1β.
(C) MCF IL-1β full response curve using buffer to 50 and 100%
human serum as sample diluents. The sample was incubated for 5 min.
(D) MCF IL-1β full response curve using buffer and 100% of human
serum as sample diluents. The sample was incubated for 30 min. All
MCF assays were performed using 40 μg/mL CapAb, which promotes
approximately half of the immobilized antibody monolayer with antibodies
oriented “end-on” with F(ab) in line.[31] Note that *P ≤ 0.05; **P ≤ 0.01; ***P ≤ 0.001 in
Tukey’s multiple comparisons test.
Effect of sample incubation time on IL-1β
sandwich MCF immunoassays.
(A) IL-1β sandwich assay in buffer, whole blood, and serum,
considering 5 min sample incubation time and 0.125 ng/mL IL-1β.
(B) IL-1β sandwich assay in buffer, whole blood, and serum,
considering 30 min sample incubation time and 0.125 ng/mL IL-1β.
(C) MCF IL-1β full response curve using buffer to 50 and 100%
human serum as sample diluents. The sample was incubated for 5 min.
(D) MCF IL-1β full response curve using buffer and 100% of human
serum as sample diluents. The sample was incubated for 30 min. All
MCF assays were performed using 40 μg/mL CapAb, which promotes
approximately half of the immobilized antibody monolayer with antibodies
oriented “end-on” with F(ab) in line.[31] Note that *P ≤ 0.05; **P ≤ 0.01; ***P ≤ 0.001 in
Tukey’s multiple comparisons test.A conventional strategy for reducing sample matrix interference
in high-sensitivity immunoassays involves diluting the sample, which
can be effective depending on the sample dilution factor.[37,38]Figure C shows for
short incubation time a good overlap of full IL-1β response
curves in buffer and 50% human serum, confirming that sample dilution
is also effective in capillary immunoassays, agreeing again with our
previous results for PSA sandwich immunoassays with human serum and
whole blood samples for both colorimetric and fluorescent detection.[3,26] Yet, with respect to POC applications, sample dilution adds another
complex step, which requires automation or precise pipetting and can
compromise the clinical value of the test by reducing the limit of
detection of the immunoassay. In the case of the IL-1β sandwich
immunoassay, we noticed that sample dilution resulted in a similar
lower limit of detection (LLoD) in the 50% human serum matrix compared
to 100% human serum (Table ). This supports rapid, high-performance quantitation is also
possible in capillary immunoassays with sample dilution. Table also shows that longer
incubation times increase LLoDs. LLoDs are determined by the blank
value and its standard deviation; therefore, their variation is related
to the development of background noise, which can be caused by nonspecific
binding of serum/blood components. Although longer periods increase
the probability of desorption of low-affinity components in the presence
of the analyte eliminating the matrix effect, they increase the probability
of nonspecific binding in the absence of the analyte, negatively impacting
the LLoD of assays. Consequently, it is important to find a suitable
sample incubation time that can simultaneously enable the management
of the matrix effect and maintain the desired LLoD performance.
Table 1
IL-1β Sandwich Assay Sensitivity
Considerations in Buffer and in Human Serum after 4PL Model Fitting
and Analyses (Figure C,D)
sample incubation time (min)
sample matrix
lower limit
of detection (LLoD) (ng/mL)
precision (with 0.125 ng/mL IL-1β) (%)
R2 (with 4PL model)
5
buffer
0.021
9
0.9992
5
100% serum
0.014
6
0.9989
5
50% serum
0.006
19
0.9991
30
buffer
0.084
20
0.9929
30
100% serum
0.051
10
0.9965
Diameter Dependence of the Matrix Effect
We have recently
reported that surface coverage of an antibody by passive adsorption
in Teflon FEP microfluidic strips is dependent on the capillary diameter[31] (Figure A). From a theoretical perspective, the capillary diameter
is known to affect the total surface area (SA) available for antibody
immobilization, as well as the sample/reagent volume (V), which in turn affects the mass and density of the immobilized
capture antibody. On the other hand, the surface-area-to-volume ratio
(SAV) becomes an important parameter as it can govern the antigen–antibody
equilibrium and the rate of the reaction rate; overall the choice
about the capillary diameter can be seen as a balance between the
total SA and the SAV (Figure B). Although surface density (ng/cm2) is independent
of the diameter of the capillary, due to the smaller sample volume
loaded, small diameter capillaries yield a much lower surface density
of the immobilized antibody compared to larger diameter capillaries.
In such a case, the number of adsorbed molecules is limited by the
number of molecules in solution. Barbosa et al.[31] reported half of the amount of immobilized antibodies on
the 109 μm diameter MCF compared to the amount immobilized on
MCF strips with mean internal diameter 212 μm. On the other
hand, the 375 μm diameter MCF also presented a significantly
higher maximum surface density compared to the 212 μm MCF (867.8
and 609.5 ng/cm2, respectively). Capillary diameter also
affects the maximum diffusion distance that molecules have to travel
in a heterogeneous immunoassay (with the capture antibody or first
member of the binding pair immobilized on the inner wall of the capillaries)
with the time of diffusion increasing to the square of the distance
according to Einstein’s law of diffusion.[39] Diffusion can be affected by the viscosity of the sample;
however, immunoassay signals with serum samples were not significantly
different than signals with the buffer and we have not detected any
significant variations in both assay kinetics and equilibrium, as
shown and discussed in the Supporting Information (Figure S1 and Table S1). Consequently, we have studied the
effect of capillary diameter in MCF assays with the mIgG/anti-mIgG
assay. We noticed a decrease in capillary diameter from 212 to 109
μm in buffer resulted in 2 min faster antibody–antigen
equilibrium, while an increase in capillary diameter from 212 to 375
μm delayed the equilibrium by 2 min, confirming that capillary
immunoassays are also diffusion-limited (Figure C and Table ). Note that kinetic constants obtained reflect the
overall strength and stability of the antibody monolayer, which depends
on structural rearrangements of mIgG antibodies and the fact that
anti-mIgG can bind in a bivalent way to the immobilized antibodies
if properly oriented. Also, the low-affinity antigen, like anti-mIgG,
is strongly affected by the mIgG density and therefore structural
orientation.[40] Further experiments with
undiluted human serum showed that the equilibrium is surprisingly
changed for small capillaries (Figure D and Table ) in the presence of the biologic human serum matrix, revealing
a level of interference of the matrix that could not be detected by
comparing other microcapillary diameters tested. This can be explained
by the different antibody surface coverages in the different capillary
geometries due to the adsorption equilibrium and on/off rates of the
immobilized antibody onto Teflon FEP surfaces being dependent on capillary
geometry, as explained previously. Overall, this confirmed a correlation
between antibody surface coverage, microcapillary diameter, and matrix
interference, with reduced antibody coverages being more prone to
matrix interference.
Figure 4
Relationship between the human serum matrix effect and
capillary
geometry in MCF assays. (A) (i) Photograph of three different MCFs
with 109 ± 12, 212 ± 16, and 375 ± 29 μm mean
diameter bore. (ii) Microscopy photograph of a cross section from
the MCFs with 109, 212, and 375 μm diameter bore. (B) Effect
of diameter size on the total surface area (SA) and on the surface-area-to-volume
ratio (SAV). (C) mIgG–anti-mIgG binding kinetics on different
diameter MCFs in the buffer matrix. (D) mIgG–anti-mIgG binding
kinetics in different diameter MCFs in the undiluted human serum matrix.
Table 2
Kinetic Constants of Anti-mIgG Binding
in Buffer and Human Serum in Different Capillary Diameter MCFs (Figure C,D)
buffer
human
serum
109 μm MCF
212 μm MCF
375 μm MCF
109 μm MCF
212 μm MCF
375 μm MCF
Kon (M s–1)
6.39 × 106
4.24 × 106
2.65 × 106
5.38 × 106
3.35 × 106
9.62 × 107
Koff (s–1)
1.42 × 10–3
1.71 × 10–3
1.55 × 10–3
8.52 × 10–3
1.91 × 10–3
4.13 × 10–2
Kd (M–1)
2.23 × 10–10
4.04 × 10–10
5.87 × 10–10
1.58 × 10–9
5.70 × 10–10
4.29 × 10–10
Relationship between the human serum matrix effect and
capillary
geometry in MCF assays. (A) (i) Photograph of three different MCFs
with 109 ± 12, 212 ± 16, and 375 ± 29 μm mean
diameter bore. (ii) Microscopy photograph of a cross section from
the MCFs with 109, 212, and 375 μm diameter bore. (B) Effect
of diameter size on the total surface area (SA) and on the surface-area-to-volume
ratio (SAV). (C) mIgG–anti-mIgG binding kinetics on different
diameter MCFs in the buffer matrix. (D) mIgG–anti-mIgG binding
kinetics in different diameter MCFs in the undiluted human serum matrix.
Novel Analytical Approach for Managing Matrix Interference in
Miocrocapillary Protein Immunoassays
We have combined the
new finding in sample matrix interference with our several years’
experience in high-performance microcapillary immunoassays to propose
a new analytical approach (Figure A), which we believe will help the effective management
of the sample matrix effect in miniaturized immunoassays. This approach
can easily be integrated into routine assay development helping to
deliver more robust microfluidic immunoassays, especially in the MCF
platform, enabling rapid, sensitive, accurate, and decentralized quantitative
protein immunoassay testing (Figure B). The first stage in the development of, e.g., a
protein sandwich immunoassay should be the choice of optimal antibody
surface coverage for minimal matrix interference. This can be done
by changing CapAb concentration in buffer, yet it is essential that
this is also carried out in human serum. Low antigen concentrations
will favor the antibody surface coverage yielding an enhanced limit
of detection for the test (Phase A, Figure ). For the best signal-to-noise ratio (yielding
the lowest limit of detection), concentration and incubation times
should be optimized for both the detection antibody and enzyme. This
assay development stage can be performed in buffer (Phase B, Figure ) and easily translated
into whole serum. Finally, the protein immunoassay should be performed
in buffer and serum samples, with sample incubation times varied to
obtain the effective working window offering negligible or minimum
matrix interference (Phase C, Figure ). This integrated analytical approach will enable
the accurate quantitation of different proteins in the microfluidic
platforms from nondiluted serum samples, as shown in this work.
Figure 5
Analytical
approach for minimizing biological matrix interference
in MCF sandwich assays. (A) Diagram with CapAb concentration and sample
incubation time for minimizing matrix interference in MCF assays.
(B) Diagram showing MCF assay development and optimization for rapid,
sensitive, and accurate quantitative assays.
Analytical
approach for minimizing biological matrix interference
in MCF sandwich assays. (A) Diagram with CapAb concentration and sample
incubation time for minimizing matrix interference in MCF assays.
(B) Diagram showing MCF assay development and optimization for rapid,
sensitive, and accurate quantitative assays.
Conclusions
Human serum sample matrix interference was fully
eliminated in
three different miniaturized enzymatic immunoassays (a direct mIgG/anti-mIgG,
a sandwich human PSA, and a sandwich human IL-1β) by manipulating
antibody surface coverage and sample incubation time. An optimal antibody
density, with antibodies presenting optimal binding capacity, is ideal
for overcoming the matrix effect. Longer sample incubation can be
effective in minimizing sample interference for certain capillary
immunoassays, with equilibrium clearly not affected by the matrix
effect, only the kinetics of binding being slowed down, yet the strategy
revealed diameter-dependent. Our results pointed to matrix interference
being linked to a competition between low-affinity interference factor(s)
and high-affinity antigens and reagents; therefore, both sample dilution
and incubation time are effective in minimizing matrix interference
in microcapillary immunoassays. The novel simple, analytical immunoassay
development approach proposed is expected to help in speeding up the
development of robust, accurate, high-performance, and decentralized
miniaturised protein immunoassays. The results shown are perhaps specific
to fluoropolymer microcapillaries (which allow production of hundreds
or thousands of disposable test strips at a minimum cost without any
complex automation), yet the surface properties of Teflon FEP are
not so distinct from other polymers such as PDMS (both hydrophobic);
therefore, we believe that these new bioanalytical approaches can
benefit both the research and innovation communities working on immunoassay
miniaturization including conventional and modern microfluidic technologies
for POC testing.
Authors: Changchun Liu; Shih-Chuan Liao; Jinzhao Song; Michael G Mauk; Xuanwen Li; Gaoxiang Wu; Dengteng Ge; Robert M Greenberg; Shu Yang; Haim H Bau Journal: Lab Chip Date: 2016-02-07 Impact factor: 6.799
Authors: A Ahmad Tajudin; K Petersson; A Lenshof; A-M Swärd-Nilsson; L Aberg; G Marko-Varga; J Malm; H Lilja; T Laurell Journal: Lab Chip Date: 2013-05-07 Impact factor: 6.799
Authors: Marcin Drozd; Polina Ivanova; Katarzyna Tokarska; Kamil Żukowski; Aleksandra Kramarska; Adam Nowiński; Ewa Kobylska; Mariusz Pietrzak; Zbigniew Brzózka; Elżbieta Malinowska Journal: Int J Mol Sci Date: 2022-03-29 Impact factor: 5.923