| Literature DB >> 34094595 |
Ziming Liu1,2, Emmanuel Eric Pazo1, Hong Ye1, Cui Yu3, Ling Xu1, Wei He1.
Abstract
PURPOSE: To assess the repeatability and agreement of refractive measurements using 2WIN-S photoscreening with the gold-standard cycloplegic retinoscope refraction.Entities:
Year: 2021 PMID: 34094595 PMCID: PMC8163555 DOI: 10.1155/2021/6612476
Source DB: PubMed Journal: J Ophthalmol ISSN: 2090-004X Impact factor: 1.909
Figure 1The 2WIN-S includes a photorefractometer (2WIN) and an occluder tube.
Comparison of measurements between cycloplegic retinoscopy and 2WIN-S.
| Cycloplegic retinoscopy | 2WIN-S |
| 95% CI | |||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Mean (SD) | Min | Max | Mean (SD) | Min | Max | Δ | Lower | Upper | ||
| Spherical (D) | −1.23 (1.78) | −5.75 | 7.00 | −1.62 (1.37) | −5.50 | 2.75 | −0.39 | <0.01 | −0.52 | −0.27 |
| Cylindrical (D) | −0.30 (0.67) | −3.25 | 3.00 | −0.42 (0.68) | −3.50 | 3.00 | −0.12 | <0.01 | −0.16 | −0.07 |
| Spherical equivalent (D) | −1.38 (1.90) | −6.00 | 7.25 | −1.83 (1.48) | −6.00 | 3.25 | −0.45 | <0.01 | −0.60 | −0.33 |
| J0 (D) | 0.17 (0.32) | −0.47 | 1.60 | 0.14 (0.35) | −1.49 | -1.73 | −0.30 | <0.01 | −0.77 | 0.02 |
| J45 (D) | −0.01 (0.11) | −0.50 | 0.40 | −0.01 (0.11) | −0.44 | 0.57 | 0.01 | <0.01 | −0.02 | 0.01 |
P value <0.05; CI, confidence interval; SD, standard deviation; Min, minimum; Max, maximum; D, diopter; J0, (cylinder/2) cos (2 axis); J45, (cylinder/2) sin (2 axis). ΔComparison between cycloplegic retinoscopy and 2Win-S (2WIN-S reading minus cycloplegic retinoscopy reading).
Figure 2(a) Frequency distribution of spherical equivalent with cycloplegic retinoscopy measurement. (b) Frequency distribution of spherical equivalent with 2WIN-S photorefractometer measurement.
Correlation between cycloplegic retinoscopy and 2WIN-S measurements.
| Parameters | Pearson's | 95% CI |
| |
| Lower | Upper | |||
|
| ||||
| Sphere (D) | 0.876 | 0.804 | 0.926 | <0.01 |
| Cylinder (D) | 0.887 | 0.750 | 0.976 | <0.01 |
| Spherical equivalent (D) | 0.875 | 0.795 | 0.928 | <0.01 |
P value <0.05; CI, confidence interval; D, diopter.
Figure 3(a) Bland–Altman plot for spherical equivalent. (b) Bland–Altman plot for J0. (c) Bland–Altman plot for J45.
Reliability and repeatability of 2WIN-S.
| Parameters | ICC | 95% CI |
| Sw | |
| Lower | Upper | ||||
|
| |||||
| Spherical equivalent (D) | 0.900 | 0.816 | 0.939 | <0.01 | 0.61 |
| J0 | 0.666 | 0.557 | 0.748 | <0.01 | 0.30 |
| J45 | 0.639 | 0.52 | 0.728 | <0.01 | 0.31 |
P value <0.05. CI, confidence interval; D, diopter; ICC, intraclass correlation coefficient; J0, (cylinder/2) cos (2 axis); J45, (cylinder/2) sin (2 axis); Sw,within-subject standard deviation.