Comparison of 2Win and plusoptiX A12R refractometers with Retinomax handheld autorefractor keratometer.

Elisabetta Racano1, Salvatore Alessi2, Riccardo Pertile3.   

Abstract

PURPOSE: To test the accuracy and validity of the 2Win and the plusoptiX A12R refractometers in detecting amblyopia risk factors.
METHODS: Children were screened using both devices, using two sets of referral criteria each, and underwent complete ophthalmic examination, including cycloplegic refraction. Sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) were calculated. Median values for the pairs of refractometers were compared using the Wilcoxon signed-rank sum test for sphere, cylinder, and power vectors J0 and J45 for axis (both eyes).
RESULTS: A total of 284 eyes of 142 children (mean age, 37.9 ± 19.8 months) were included. Comparison of mean cycloplegic and manifest refractometer measurements provided statistically significant differences in both eyes. For sphere, the means were lower and for cylinder, higher for both devices, and both correlated well with the gold standard for astigmatic power vectors J0 and J45. Using referral criteria 1, the sensitivity, specificity, PPV and NPV, and inconclusive results were 67.4%, 83.7%, 87.9%, 59.4%, and 4.9%, respectively, for the 2Win, with sensitivity of 13% in the hyperopia group; 73.1%, 95.9%, 96.6%, 69.1%, and 10.6%, respectively, for the plusoptiX A12R, with a sensitivity of 33.3% in the hyperopia group. Using criteria 2, the values were 98.8%, 38.8%, 73.9%, and 95.0% (2Win) and 94.9%, 65.3%, 81.3%, and 88.9% (plusoptiX A12R).
CONCLUSIONS: In manifest conditions, the accuracy of the 2Win and plusoptiX A12R refractometers is low in hyperopia and astigmatism, but the devices are well correlated with each other, and both have high specificity; sensitivity is low in hyperopia, resulting in the underestimation of hyperopic refractive error. The optional Plusoptix sensitive referral criteria seems to be appropriate for the A12R. The 2Win provided fewer inconclusive results and was preferred for use with younger and developmentally delayed children.
Copyright © 2019 American Association for Pediatric Ophthalmology and Strabismus. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2019        PMID: 31539614     DOI: 10.1016/j.jaapos.2019.05.017

Source DB:  PubMed          Journal:  J AAPOS        ISSN: 1091-8531            Impact factor:   1.220


  9 in total

1.  Artificial intelligence improves accuracy, efficiency, and reliability of a handheld infrared eccentric autorefractor for adult refractometry.

Authors:  Yi-Ting Cao; Dan-Yang Che; Yi-Lei Pan; Yun-Li Lu; Chong-Yang Wang; Xiao-Li Zhang; Yun-Fei Yang; Ke-Ke Zhao; Ji-Bo Zhou
Journal:  Int J Ophthalmol       Date:  2022-04-18       Impact factor: 1.779

2.  Performance of Glow Fixation GoCheck Kids and 2WIN Photoscreeners and Retinomax to Uncover Hyperopia.

Authors:  Alexa H Levitt; Samuel J Martin; Robert W Arnold
Journal:  Clin Ophthalmol       Date:  2020-08-10

3.  Drive-by Photoscreening: Plusoptix, 2WIN and Blinq Amblyopia Detection During the COVID-19 Pandemic.

Authors:  Mason Keffalos; Samuel Martin; Robert Arnold
Journal:  Clin Ophthalmol       Date:  2021-02-23

4.  Ellipsoid Spectacle Comparison of Plusoptix, Retinomax and 2WIN Autorefractors.

Authors:  Robert W Arnold; Samuel J Martin; Joshua R Beveridge; Andrew W Arnold; Stephanie L Arnold; Nathanael R Beveridge; Kyle A Smith
Journal:  Clin Ophthalmol       Date:  2021-08-30

5.  Instrument Referral Criteria for PlusoptiX, SPOT and 2WIN Targeting 2021 AAPOS Guidelines.

Authors:  Robert Arnold; David Silbert; Heather Modjesky
Journal:  Clin Ophthalmol       Date:  2022-02-25

6.  Multifaceted Amblyopia Screening with blinq, 2WIN, and PDI Check.

Authors:  Robert Arnold; Mario Angi
Journal:  Clin Ophthalmol       Date:  2022-02-15

7.  Comparative Validation of PlusoptiX and AI-Optic Photoscreeners in Children with High Amblyopia Risk Factor Prevalence.

Authors:  Robert W Arnold
Journal:  Clin Ophthalmol       Date:  2022-08-16

8.  Comparing School-Aged Refraction Measurements Using the 2WIN-S Portable Refractor in Relation to Cycloplegic Retinoscopy: A Cross-Sectional Study.

Authors:  Ziming Liu; Emmanuel Eric Pazo; Hong Ye; Cui Yu; Ling Xu; Wei He
Journal:  J Ophthalmol       Date:  2021-05-21       Impact factor: 1.909

9.  Performance of Two Photoscreeners Enhanced by Protective Cases.

Authors:  Samuel Joseph Martin; Hser Eh Htoo; Nay Hser; Robert W Arnold
Journal:  Clin Ophthalmol       Date:  2020-05-25
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.