| Literature DB >> 34055588 |
Andrey Bokov1, Svetlana Pavlova2, Anatoliy Bulkin2, Alexandr Aleynik2, Sergey Mlyavykh2.
Abstract
BACKGROUND: The majority of published data report the results of biomechanical tests of various design pedicle screw performance. The clinical relevance and relative contribution of screw design to instrumentation stability have been insufficiently studied. AIM: To estimate the contribution of screw design to rate of pedicle screw loosening in patients with degenerative diseases of the lumbar spine.Entities:
Keywords: Degenerative diseases; Lumbar spine; Pedicle screw design; Pedicle screw loosening
Year: 2021 PMID: 34055588 PMCID: PMC8152435 DOI: 10.5312/wjo.v12.i5.310
Source DB: PubMed Journal: World J Orthop ISSN: 2218-5836
Figure 1Computed tomography image of the lumbar spine in the coronal plane. Bilateral double halo sign is evident as a radiolucent zone around pedicle screws surrounded by sclerotic bone.
Characteristics of the study patients, n (%)
|
|
|
| Number of levels fused | |
| 1 level | 115 (65.7) |
| 2 levels | 47 (26.9) |
| 3 levels | 16 (9.1) |
| 4 levels | 1 (0.6) |
| 5 levels | 1 (0.6) |
| Fusion type | |
| Posterior fusion only | 25 (14.3) |
| TLIF | 150 (85.6) |
| Amount of decompression | |
| No resection of posterior structures | 18 (10.3) |
| Laminectomy | 46 (26.3) |
| Unilateral total facet joint removal | 108 (61.7) |
| Bilateral total facet joint removal on at least one level | 46 (26.3) |
| Bone characteristics | |
| Radiodensity, HU | M = 126.25 ± 3.20 SD = 42.31, range 43.17-282.07 |
HU: Hounsfield units; TLIF: Transforaminal lumbar interbody fusion.
Characteristics of the screws used in this study
|
|
|
|
|
|
|
| cyl/cyl | 6.5 | 4.0 | 3.0 | Buttress | 6 (3.4) |
| cyl/cyl | 7.5 | 4.5 | 3.0 | Buttress | 9 (5.1) |
| cyl/con | 6.5 | 4.0 | 2.5 | Buttress | 40 (22.9) |
| cyl/con | 7.5 | 5.0 | 2.5 | Buttress | 4 (2.3) |
| con/con | 6.0 | 4.0 | 3.0 | V-shape | 48 (27.4) |
| con/con | 7.0 | 4.5 | 3.0 | V-shape | 5 (2.9) |
| con/con | 6.0 | 4.5 | 2.5 | V-shape | 49 (28.0) |
| con/con | 7.0 | 5.5 | 2.5 | V-shape | 14 (8.0) |
con/con: Conical core and thread; cyl/con: Cylindrical thread and conical core; cyl/cyl: Cylindrical core and thread.
Parameters included in the general logistic regression model
|
|
|
|
| Intercept | −4.819065; | - |
| Radiodensity, HU | −0.0271586; | 0.9732 (0.9623-0.9842) |
| Number of fused levels | 1.111651; | 3.0394 (1.7566-5.2590) |
| Bilateral total facet joints removal (at least one level) | 1.124871; | 3.0798 (1.0023-9.2781) |
| Laminectomy | 0.4418158; | 1.5555 (0.5090-4.7533) |
| Posterior fusion performed without interbody fusion | −0.4247137; | 0.6540 (0.2128-2.0097) |
| Maximal external screw diameter | −5.966555; | 0.0026 (0.000009-0.7153) |
| Maximal internal screw diameter | 6.430714; | 620.61 (2.2363-172234.6) |
| Helical pitch | 4.611687; | 100.6538 (3.3668-3009.1310) |
| Thread type (buttress | −1.034237; | 0.1070 (0.0090-1.2659) |
| Design cyl/cyl | 0.1446205; | 1.1566 (0.2378-5.6156) |
| Design cyl/con | −1.731799; | 0.1769 (0.0077-4.0602) |
| Design con/con | 0.2718265; | 1.3124 (0.2282-7.5451) |
CI: Confidence interval; con/con: Conical core and thread.; cyl/con: Cylindrical thread and conical core; cyl/cyl: Cylindrical core and thread; HU: Hounsfield unit; OR: Odds ratio.