| Literature DB >> 34040293 |
Vikrant Kumar1, Sachin Tayal1, Abbas Ali1, Arun Gandhi1.
Abstract
PURPOSE OF STUDY: The purpose of the study was to evaluate the effect of patient characteristics and equipment-related factors on the computed tomography (CT) dose received by patients from positron emission tomography-CT (PET-CT) using system-generated dose-length product (DLP) values and also to check the effective dose (ED) received from various CT protocols at our institute.Entities:
Keywords: Computed tomography dose index; dose-length product; effective dose; positron emission tomography-computed tomography
Year: 2021 PMID: 34040293 PMCID: PMC8130704 DOI: 10.4103/ijnm.IJNM_112_20
Source DB: PubMed Journal: Indian J Nucl Med ISSN: 0974-0244
Detailed computed tomography acquisition parameters for each protocol
| Scan | Scan type | Thickness | Pitch | Tube potential (kV) | Tube current (mA) | Prep group (sec) |
|---|---|---|---|---|---|---|
| Head and neck arterial CT | Helical | 1.25 mm | 0.938:1 | 120 | 150-180 (smart mA) | 25-30 s based upon bolus tracking (arterial phase CT) |
| Brain CT | Axial | 2.5 mm | - | 120 | 250 mA | 0 s |
| Pelvis arterial CT | Helical | 1.25 mm | 0.938:1 | 120 | 150-200 (Smart mA) | 25-30 s based upon bolus tracking (arterial phase CT) |
| Thorax CT | Helical | 1.25 mm | 1.375:1 | 120 | 150-180 (smart mA) | 5.9 s |
| Whole Body CT | Helical | 3.75 mm | 1.375:1 | 120 | 150-320 (smart mA) | 40-50 s delay based upon contrast administered (venous phase CT) |
Smart mA – Modulates tube current according to patient characteristics. In Thorax CT 5.9 sec prep is to used to instruct patient to hold their breath during acquisition (to limit motion artifacts in image). For simplicity of calculation, Smart Prep CT was included in Arterial phase CT. CT: Computed tomography
Figure 1System generated dose report in case of head and neck protocol. Series 2 Represents dose report about breathhold thorax computed tomography. Series 200 Represents dose report of smart prep CT dose. Series 4- Represents dose report of arterial phase head and neck computed tomography. Series 5- Represents dose report of whole body computed tomography. Note: For simplicity of calculation, Smart Prep dose was included in arterial phase computed tomography dose
Conversion factor k-coefficient (mSv/mGy/cm)
| Region of the body | 0 year old | 1 year old | 5 year old | 10 year old | Adult |
|---|---|---|---|---|---|
| Head | 0.011 | 0.0067 | 0.0040 | 0.0032 | 0.0021 |
| Head and neck | 0.013 | 0.0085 | 0.0057 | 0.0042 | 0.0031 |
| Neck | 0.017 | 0.012 | 0.011 | 0.0079 | 0.0059 |
| Chest | 0.039 | 0.026 | 0.018 | 0.013 | 0.014 |
| Abdomen and pelvis | 0.049 | 0.030 | 0.020 | 0.015 | 0.015 |
| Trunk | 0.044 | 0.028 | 0.019 | 0.014 | 0.015 |
We assumed conversion factor for Whole body to be same as that of trunk, i.e., 0.015 (adult)
Figure 2Effective dose received in head and neck protocol
Figure 3Effective dose received in whole body with brain protocol
Effective dose delivered in various protocols
| Group | Thorax CT | Whole body CT | Head and neck arterial CT | Brain CT | Pelvis arterial CT | Ttal |
|---|---|---|---|---|---|---|
| Group A | 3.47±0.26 | 17.28±1.43 | 1.69±0.18 | _ | _ | 22.45±1.58 |
| Group B | 3.26±0.32 | 17.43±2.20 | _ | 1.7±0.16 | _ | 22.40±2.44 |
| Group C | 3.22±0.37 | 15.72±2.83 | _ | _ | 2.29±0.26 | 21.24±3.11 |
For simplicity of calculation, Smart Prep CT was included in Arterial phase CT. ED (mSv) represented as Average±SD. SD: Standard deviation, CT: Computed tomography, ED: Effective dose
Figure 4Effective dose received in pelvis protocol
Effective dose received in various scans
| Scan | Height (cm) | Weight (Kg) | Scan range (cm) | Tube current (mA) | CTDIvol (mGy) | DLP (mGy.cm) | ED (mSv) |
|---|---|---|---|---|---|---|---|
| Head and neck arterial CT | 162.57 | 52 | 24.75 | 177.14 | 19.8 | 516.81 | 1.69 |
| Brain CT | 153.54 | 55.92 | 15 | 230 | 53.28 | 811.28 | 1.7 |
| Pelvis arterial CT | 154.6 | 50.2 | 30.7 | 197.8 | 22.24 | 710.34 | 2.29 |
| Thorax CT | 158.77 | 52.97 | 36.99 | 163.28 | 5.58 | 225.31 | 3.37 |
| Whole body CT | 158.77 | 52.97 | 97.82 | 291.13 | 11.43 | 1142.07 | 17.13 |
For simplicity of calculation, Smart Prep CT was included in Arterial phase CT. CT: Computed tomography, DLP: Dose-length product, ED: Effective dose
Figure 5Effective dose received in various scans
Correlation of various parameters with effective dose
| Pearson correlations | Height (cm) | Weight (kg) | Scan range (cm) | Tube current (mA) | CTDIvol (mGy) | DLP (mGy.cm) |
|---|---|---|---|---|---|---|
| Weight (kg) | 0.493 | |||||
| Scan range (cm) | 0.104 | 0.05 | ||||
| Tube current (mA) | −0.065 | 0.056 | 0.818 | |||
| CTDIvol (mGy) | −0.119 | 0.084 | −0.4 | 0.129 | ||
| DLP (mGy.cm) | 0.036 | 0.11 | 0.763 | 0.943 | 0.244 | |
| ED (mSv) | 0.05 | 0.064 | 0.983 | 0.884 | −0.284 | 0.834 |
Scan range, Tube current and DLP are Strongly Positively correlated with ED. Patient’s Height and Weight are Weakly Positively correlated with ED. DLP: Dose-length product, ED: Effective dose
Regression analysis
| Regression Equation | |||
|---|---|---|---|
| Individual ED (mSv)=–2.35 –0.027 ˣHt (cm) +0.014 ˣWt (Kg) | |||
| +0.148 ˣScan range (cm) +0.019 ˣTube current (mA) -0.035 | |||
| ˣCTDIvol (mGy) +0.002 ˣDLP (mGy.cm) | |||
| 0.782234 | 98.79% | 98.76% | 98.66% |
| Regression | 1890.33 | 3089.33 | 0.000 |
| Height (cm) | 11.86 | 19.39 | 0.000 |
| Weight (kg) | 6.21 | 10.15 | 0.002 |
| Scan range (cm) | 213.05 | 348.18 | 0.000 |
| Tube current (mA) | 18.39 | 30.05 | 0.000 |
| CTDIvol (mGy) | 5.69 | 9.31 | 0.003 |
| DLP (mGy.cm) | 16.16 | 26.41 | 0.000 |
| Error | 0.61 | ||
| Lack-of-fit | 0.61 | ||
| Pure error | 0.00 | ||
| Total | |||
MS: Mean square, DLP: Dose-length product
Figure 6Effective dose delivered in various groups