Meng Huang1, Carlos I Dorta-Quiñones2, Bradley A Minch3, Manfred Lindau1. 1. School of Applied & Engineering Physics, Cornell University, Ithaca, New York 14853, United States. 2. School of Electrical & Computer Engineering, Cornell University, Ithaca, New York 14853, United States. 3. Franklin W. Olin College of Engineering, Needham, Massachusetts 02492, United States.
Abstract
Complementary metal-oxide-semiconductor (CMOS) microelectrode arrays integrate amplifier arrays with on-chip electrodes, offering high-throughput platforms for electrochemical sensing with high spatial and temporal resolution. Such devices have been developed for highly parallel constant voltage amperometric detection of transmitter release from multiple cells with single-vesicle resolution. Cyclic voltammetry (CV) is an electrochemical method that applies voltage waveforms, which provides additional information about electrode properties and about the nature of analytes. A 16-channel, 64-electrode-per-channel CMOS integrated circuit (IC) fabricated in a 0.5 μm CMOS process for CV is demonstrated. Each detector consists of only 11 transistors and an integration capacitor with a unit dimension of 0.0015 mm2. The device was postfabricated using Pt as the working electrode material with a shifted electrode design, which makes it possible to redefine the size and the location of working electrodes. The system incorporating cell-sized (8 μm radius) microelectrodes was validated with dopamine injection tests and CV measurements of potassium ferricyanide at a 1 V/s scanning rate. The cyclic voltammograms were in excellent agreement with theoretical predictions. The technology enables rigorous characterization of electrode performance for the application of CMOS microelectrode arrays in low-noise amperometric measurements of quantal transmitter release as well as other biosensing applications.
Complementary metal-oxide-semiconductor (CMOS) microelectrode arrays integrate amplifier arrays with on-chip electrodes, offering high-throughput platforms for electrochemical sensing with high spatial and temporal resolution. Such devices have been developed for highly parallel constant voltage amperometric detection of transmitter release from multiple cells with single-vesicle resolution. Cyclic voltammetry (CV) is an electrochemical method that applies voltage waveforms, which provides additional information about electrode properties and about the nature of analytes. A 16-channel, 64-electrode-per-channel CMOS integrated circuit (IC) fabricated in a 0.5 μm CMOS process for CV is demonstrated. Each detector consists of only 11 transistors and an integration capacitor with a unit dimension of 0.0015 mm2. The device was postfabricated using Pt as the working electrode material with a shifted electrode design, which makes it possible to redefine the size and the location of working electrodes. The system incorporating cell-sized (8 μm radius) microelectrodes was validated with dopamine injection tests and CV measurements of potassium ferricyanide at a 1 V/s scanning rate. The cyclic voltammograms were in excellent agreement with theoretical predictions. The technology enables rigorous characterization of electrode performance for the application of CMOS microelectrode arrays in low-noise amperometric measurements of quantal transmitter release as well as other biosensing applications.
Neurotransmitters
are biomolecules stored in secretory vesicles
and released to the extracellular environment as messengers in neuronal
communication.[1] When a cell such as a neuron
or neuroendocrine cell is activated or stimulated, secretory vesicles
fuse with the plasma membrane and transmitter molecules are released.
This process of vesicle release in packages or quanta is called exocytosis.[2] Dysfunction of synaptic transmission is related
to many diseases, such as Parkinson’s disease and Alzheimer’s
disease. A better understanding of exocytosis and mechanisms by which
certain drugs modulate this process is essential for a mechanistic
understanding of therapeutic effects and the development of new types
of drugs.Electrochemistry has been widely used to characterize
the exocytotic
release of compounds such as dopamine (DA), norepinephrine, epinephrine,
and serotonin.[3] In amperometry, a polarizable
working electrode is set at a constant voltage with respect to a reference
electrode, and neurotransmitters are oxidized upon reaching the electrode.
This method provides high sensitivity and high time resolution but
does not offer the capability of distinguishing between different
electroactive species. On the other hand, CV can distinguish different
electroactive species at the cost of reduced temporal resolution.[4] In CV, a voltage ramp is typically applied at
the working electrode while positive and negative currents are measured.
With the background current subtracted and the resulting current plotted
vs the applied ramp voltage as a difference voltammogram, oxidation
and reduction peaks provide a unique fingerprint for different species
and can be used to identify the measured compounds.[5]For the study of transmitter release from single
cells and individual
vesicles, amperometry and CV are conventionally performed with carbon-fiber
microelectrodes, recording from a single cell at a time. This laborious
procedure requires sequential measurements from many cells and thus
is very time-consuming. Complementary metal–oxide–semiconductor
integrated circuit (CMOS IC) technology provides the possibility to
integrate hundreds or thousands of working electrodes and amplifiers
on a single silicon die for parallel recordings from multiple cells.
CMOS IC devices specifically optimized for on-chip, high-throughput,
single-cell amperometry have been described.[6−10] These devices feature high resolution but can only
measure positive current and are not suitable for CV. Incorporating
the ability to perform CV on the same chip enables not only the distinction
of different analytes but, importantly, also a means for electrode
characterization. For CV recordings, bidirectional-current measurements
are required. A few bidirectional-current measurement CMOS devices
for CV or fast-scan cyclic voltammetry (FSCV) have recently been developed.[10−13] However, they either do not provide sufficient sensitivity for characterizing
individual vesicle release events, typically on the order of a few
pA, or are not so area efficient in terms of the circuit footprint,
a key requirement for large high-throughput arrays. A peripheral cascode
bias circuit which is shared by an entire row of electrodes can be
used to significantly reduce the number of transistors per detector
as compared to a more conventional design[12] while retaining a relatively low gain error.Another challenge
for on-chip CMOS IC electrochemistry detectors
is the postfabrication process. As semiconductor foundries do not
allow the use of polarizable electrode materials, such as Pt or Au,
original electrodes are covered with Al/Cu metal contacts, which are
not suitable for electrochemical measurements. A conventional approach
is to deposit Pt or Au electrodes directly on Al/Cu metal contacts,
but working electrodes may be potentially damaged during multiple
uses and expose Al/Cu metal contacts to the test solution, resulting
in excessive currents and damage to the chip. A surface-modified structure
has been developed to solve this issue and it also offers the possibility
to redefine the size and the location of working electrodes.[9]This work presents a novel high-throughput,
high-precision, and
area-efficient CMOS circuit with only 11 transistors per detection
pixel for CV measurements plus 2 calibration transistors. The pixel
dimension was 42 μm × 36 μm and the total area of
the 1024 detector array was 1.34 mm × 1.15 mm. The device was
postfabricated with a shifted electrode surface-modified structure[9] and validated with dopamine injection and potassium
ferricyanide CV measurement.
Experimental Section
Circuit Implementation
Figure shows the schematic diagram
of the bidirectional-current potentiostat implemented in this work.
The individual detector circuit was modified and improved based on
the schematic described in ref (12). In ref (12), to ensure that the current mirror transistors M2 and M3 have the same drain
voltage, a second half-shared amplifier was used in the detector circuit
with the output connected to VC in Figure .[12] Although this topology provided a near-perfect match for
the bias points for M2 and M3, the second half-shared amplifier consumes a large area.
To reduce the number of transistors per detector pixel, the second
half-shared amplifier was removed from the circuit and replaced by
a peripheral cascode bias circuit shared by an entire row of electrodes,
thus not consuming individual detector area. The peripheral cascode
bias circuit consists of a half-shared amplifier AP, a current source I0, and
an NMOS transistor M0. The current source I0 is set to be running at a current level close
to the normal operating current of M4 during
negative current measurement in CV. With VC connected to the gate of M4, the drain
voltages of M2 and M3 are very close and thus a low gain error is achieved.
Figure 1
Schematic diagram
of the bidirectional-current potentiostat implemented
in this work. The gray color indicates components in a single detector
pixel.
Schematic diagram
of the bidirectional-current potentiostat implemented
in this work. The gray color indicates components in a single detector
pixel.Although the original circuit
in ref (12) enabled
bidirectional-current measurement, both
the positive current and the negative current were measured as positive
output voltage with the same integration capacitor. Therefore, the
output voltage indicated only the absolute value but not the polarity
of the input current. In the design presented here, a sign binary
output circuit was added to compare two bias points in the circuit
indicating if the output current was positive or negative. The sign
binary output circuit is implemented with two transistors M6 and M7 (Figure ) added to each detector
pixel. When the current is positive, the sign binary output will generate
a high voltage. When the current is negative, the sign binary output
will be low. The detector pixel also includes 1:1 current mirrors BP and BN, for the
calibration of negative current (IP) and
positive current (IN), respectively.With the removal of the second half-shared amplifier and the addition
of the sign binary output circuit, the number of transistors reduced
by three compared with ref (12) to a total number of 11 transistors per detector pixel
for the bidirectional-current potentiostat plus two calibration transistors.
CMOS Sensor Array Fabrication
The
CMOS sensor array IC devices were fabricated through MOSIS in the
ON Semiconductor C5F/N process. A micrograph of the sensor array is
presented in Figure . To improve the area efficiency, the physical layout of the 16-channel,
64-electrode-per-channel sensor array was arranged in 32 × 32
format (Figure ).
Each channel was composed of two 32-electrode columns. The 64 electrodes
(2 × 32) of a given channel shared one output buffer. The half-shared
amplifier A0 and the peripheral cascode
bias circuit in Figure were shared by the 32 electrodes in the same row. As original electrodes
were covered with Al/Cu metal contacts, postfabrication processes
were used to deposit Pt as the polarizable working electrode material.
A holder with the exact dimension of the CMOS die was fabricated to
mount the CMOS chip inside for better handling. The holder was etched
using a deep silicon etcher (Unaxis 770). The chip was bonded to the
holder with epoxy (Hardman DOUBLE/BUBBLE Extra Fast Epoxy).
Figure 2
Micrograph
of the 1024-electrode bidirectional-current measurement
CMOS IC.
Micrograph
of the 1024-electrode bidirectional-current measurement
CMOS IC.For the postfabrication, a modified
shifted electrode design based
on the method described in ref (9) was used. The chip was spin-coated with LOR 5A (Microchem)
at 2000 rpm for 45 s and then soft-baked at 180 °C for 4.5 min.
It was then spin-coated with Shipley S1813 at 4000 rpm for 30 s and
soft-baked at 90 °C for 1 min. After UV exposure, the chip was
baked at 90 °C for 1 min to remove the air bubbles generated
during the exposure. Then, the chip was soaked in 726 MIF developer
(Microchem) for 1 min 10 s. Descum was performed with YES Asher. A
sputtering system (AJA International Inc.) was used for Ti (60 s,
10 nm) and Pt (500 s, 250 nm) deposition using 400 W power. The chip
was immersed in Remover PG (Microchem) overnight for metal lift-off.
The insulation layer was fabricated by first spin-coating AZ nLOF
2020 (3000 rpm, 30 s) on the CMOS chip followed by a soft-bake at
115 °C for 1 min, resulting in a 2.4 μm thick resist layer.
The chip was then aligned with the photomask and exposed with UV (365/405
nm). The postexposure bake was at 115 °C for 1 min. 726 MIF was
used to develop the photoresist for 1 min. A final descum was performed
with YES Asher. The postfabricated electrodes are shown in Figure . The chip was attached
on a side-braze dual in-line package (Addison Engineering, San Jose,
CA) and wire bonded. Silicone (RTV 615, GE) was applied to insulate
the entire device periphery, exposing only the sensor array area.
Figure 3
Microscope
image of the patterned shifted electrode array with
microwells. The diameter of the well was 16 μm and depth 2.4
μm.
Microscope
image of the patterned shifted electrode array with
microwells. The diameter of the well was 16 μm and depth 2.4
μm.
Testing
Solutions
Physiological buffer
solution contained (in mM) 140 NaCl, 5 KCl, 1 MgCl2, 10
HEPES/NaOH with pH adjusted to 7.3. DA solution was supplemented with
200 μM DA in the physiological buffer solution. Ferricyanide
solutions contained 25/50/75/100 μM potassium ferricyanide in
0.1 M KCl (pH = 3).
Data Acquisition and Analysis
A grounded
Ag|AgCl reference electrode was placed in the buffer solution. For
the DA injection test, working electrodes on chip were set at +700
mV against the reference electrode. For the ferricyanide CV measurement,
the working electrode potential was ramped between +0.6 and −0.1
V with the holding potential at +0.6 V against the reference electrode.
The voltage ramp scan rate was 1 V/s.A customized breadboard
was used for the wiring of the power and clock inputs and the device
outputs (Figure a).
To exchange solutions during the CV measurement, two pipettes were
used, one injecting the new solution and one aspirating the currently
present solution at the same time (Figure b).
Figure 4
(a) Customized breadboard used for the data
acquisition of the
CMOS device. (b) Protocol for solution exchange. Two pipettes were
used to exchange the solution during the recording, simultaneously
injecting the ferricyanide solution and aspirating the currently present
KCl solution.
(a) Customized breadboard used for the data
acquisition of the
CMOS device. (b) Protocol for solution exchange. Two pipettes were
used to exchange the solution during the recording, simultaneously
injecting the ferricyanide solution and aspirating the currently present
KCl solution.The analogue outputs were connected
to a multichannel module NIDAQ
PXIe-6368 (National Instruments, Austin, TX) and recorded by a computer
using Igor Pro 6 (Wavemetrics). The sampling rates were 2 and 10 kS/s
for DA injection tests and for ferricyanide CV measurements, respectively.
The higher sampling rate was used for CV to decrease the gain from
10 to 2 mV/pA, increasing the full-scale range to ±1.5 nA. The
raw data was demultiplexed by a custom program and filtered with five-point
binomial smoothing in Igor Pro 6.
Results
Sign Binary Output Validation
Figure a presents the results
from a current injection test feeding a prerecorded CV current (gray
solid trace) with known polarity measured with a carbon-fiber microelectrode
in 1 μM DA solution through the calibration current mirrors
(BP and BN) on the CMOS chip. The sign binary output (Figure a, blue trace) dropped from high to low at
the positive–negative transition point. At near-zero current,
the sign binary output switched between high and low a few times because
at very low current levels the difference between the two bias points
used for the sign binary output circuit is too small to accurately
generate the sign output. However, this sign uncertainty was limited
to a small range very close to zero current levels (±35 pA) (Figure b) and does not affect
higher current levels, where the oxidation and reduction peaks are
located. It is worth noting that both the positive and the negative
current matched very well with the input currents, indicating that
the modified negative current measurement path with the peripheral
cascode bias circuit has a very low gain error.
Figure 5
(a) Output current was
measured while input prerecorded current
was injected through the calibration current mirrors (BP and BN). The gray trace
indicates the input current. The black trace indicates the output
current from the CMOS chip. The blue trace indicates the sign binary
output. The black dotted trace indicates the corrected negative current
based on the sign binary output. (b) Zoomed-in graph showing the sign
uncertainty region. The horizontal scale at the top shows the applied
electrode voltage. The sign uncertainty was limited to the current
range of ±35 pA, and does not affect current at the oxidation
and reduction peaks.
(a) Output current was
measured while input prerecorded current
was injected through the calibration current mirrors (BP and BN). The gray trace
indicates the input current. The black trace indicates the output
current from the CMOS chip. The blue trace indicates the sign binary
output. The black dotted trace indicates the corrected negative current
based on the sign binary output. (b) Zoomed-in graph showing the sign
uncertainty region. The horizontal scale at the top shows the applied
electrode voltage. The sign uncertainty was limited to the current
range of ±35 pA, and does not affect current at the oxidation
and reduction peaks.
DA Injection
Test
Although the CMOS
IC in this work was designed and optimized for CV measurements, it
can also operate in low-noise constant voltage amperometry mode where
only a positive current is present. When a positive current enters
through the electrode into the chip, the circuit works the same way
as described in refs (7, 8).
DA is an electroactive neurotransmitter, which is efficiently oxidized
at an electrode set at +700 mV and was injected to validate the CMOS
chip.Twenty microliters of buffer solution was placed on the
chip surface with the reference electrode inserted into the buffer.
After the recording started, 40 μL of DA solution was added
to the device, resulting in a final DA concentration of 133 μM. Figure shows the electrode
response of DA injection. Immediately after DA injection, the current
detected increased to near saturation of the measurement circuit,
and slowly decayed after the maximum point, which is expected in such
an experiment. The root-mean-square (RMS) noise of the amperometric
recording at 2 kS/s was 217 fA before the addition of DA.
Figure 6
DA injection
test with a final DA concentration of 133 μM.
Working electrode potential +700 mV.
DA injection
test with a final DA concentration of 133 μM.
Working electrode potential +700 mV.
Ferricyanide CV Measurement
Potassium
ferricyanide is commonly used as a test analyte for the characterization
of the electrochemical properties of electrodes, such as electrochemical
sensitivity, effective electrode area, stability, etc.[14] In potassium ferricyanide CV, a negative scan
is first applied at the electrode followed by the positive scan (Figure a, black solid line)
because the ferricyanide is originally in its oxidized form. Figure a shows a typical
example of the background current for an on-chip working electrode
in 0.1 M KCl solution with a scan rate of 1 V/s. The holding potential
was set at +0.6 V. A negative scan was first performed from +0.6 to
−0.1 V and a positive scan from −0.1 V back to +0.6
V (Figure a, black
line).
Figure 7
Measured background current of 0.1 M KCl solution vs electrode
potential with a scan rate of 1 V/s. (a) Time course of background
current (blue) in response to applied ramp voltage (black). (b) Time
course around the point of ramp reversal (box in panel a) on an expanded
scale. (c) Circuit simulation of ramp voltage showing the time course
of Vref (black line), electrode voltage VE (black dashed line), and output of the half-shared
amplifier (A1 in Figure ) VOA (red dashed
line). (d) Time course around the point of ramp reversal (box in panel
c) on an expanded scale. (e) Superimposed current time courses from
three subsequent scans. (f) Cyclic voltammograms plotting currents
from panel e against Vref.
Measured background current of 0.1 M KCl solution vs electrode
potential with a scan rate of 1 V/s. (a) Time course of background
current (blue) in response to applied ramp voltage (black). (b) Time
course around the point of ramp reversal (box in panel a) on an expanded
scale. (c) Circuit simulation of ramp voltage showing the time course
of Vref (black line), electrode voltage VE (black dashed line), and output of the half-shared
amplifier (A1 in Figure ) VOA (red dashed
line). (d) Time course around the point of ramp reversal (box in panel
c) on an expanded scale. (e) Superimposed current time courses from
three subsequent scans. (f) Cyclic voltammograms plotting currents
from panel e against Vref.During the negative scan, ferricyanide is reduced to ferrocyanideand during the positive
scan, ferrocyanide
is oxidized back to ferricyanideFigure a (blue
trace) shows the background current for an on-chip working electrode
in 0.1 M KCl solution in the absence of ferricyanide with a scan rate
of 1 V/s. As expected, the current was negative during the negative
slope of the scan and turned positive during the positive slope of
the ramp. When the electrode potential returned to the holding potential
of +0.6 V at the end of the positive ramp, the current decayed back
to zero. However, close inspection of the current around −0.1
V, where the applied ramp slope reverses its sign, revealed a slight
delay in the onset of the positive current slope (Figure b). A possible cause of this
phenomenon could be that near this point the actual electrode voltage
(VE) may differ from the externally applied Vref. To address this question, we performed
simulations of the circuit revealing the time course of VE relative to Vref (Figure c,d). As shown in Figure d, the electrode
voltage shows an ∼20 ms long plateau starting at the time where
the slope of Vref switches from negative
to positive. The delayed rise of the current seen in Figure b is therefore explained by
the distorted time course of the electrode voltage. The cause of the
distortion is the limited slew rate of the output voltage VOA of the half-shared amplifier (A1 in Figure ), as shown in Figure c,d (dashed red lines). To switch the slope of Vref from negative to positive, VOA has to increase by ∼450 mV, which takes ∼20
ms, causing the plateau of the electrode voltage. However, since this
distortion of VE is confined to a very
narrow range near the slope transition point, the relevant parts of
the Faradaic currents will not be affected and the measured reduction
and oxidation peaks in difference voltammograms should not be distorted.The stability of the working electrode is essential for CV measurements.
The superimposed three cycles of background currents (Figure e) and resulting cyclic voltammograms
(Figure f) show very
little variation, indicating that capacitive background currents were
stable over the electrode potential range from −0.1 to 0.6
V.To perform CV measurements with various concentrations of
ferricyanide,
25/50/75/100 μM potassium ferricyanide in 0.1 M KCl (pH = 3)
solutions were applied. The chip was first loaded with 40 μL
of 0.1 M KCl solution without any analytes and the background CV current
was recorded for 3 cycles. While the CV recording was continued, the
KCl solution was replaced with 40 μL of 25 μM ferricyanide
solution using two pipettes, as illustrated in Figure b. The same solution exchange procedure was
repeated for 50, 75, and 100 μM potassium ferricyanide solutions.
The measured currents shown in Figure a were averaged over 3 cycles for the background current
and over 4 cycles for each of the ferricyanide concentrations and
plotted as a function of Vref as cyclic
difference voltammograms after background subtraction in Figure b. The cyclic voltammograms
(Figure b) clearly
demonstrate the oxidation and reduction peaks, which occurred at 0.22
and 0.14 V, respectively, consistent with previously reported data.[15]
Figure 8
CV measurement of various concentrations of potassium
ferricyanide.
(a) Measured background current and CV current with 25/50/75/100 μM
potassium ferricyanide with time as the horizontal axis. The applied
voltage vs Ag|AgCl reference electrode (black trace) is demonstrated
to the right vertical axis. (b) Cyclic difference voltammograms of
25/50/75/100 μM potassium ferricyanide with the background current
subtracted.
CV measurement of various concentrations of potassium
ferricyanide.
(a) Measured background current and CV current with 25/50/75/100 μM
potassium ferricyanide with time as the horizontal axis. The applied
voltage vs Ag|AgCl reference electrode (black trace) is demonstrated
to the right vertical axis. (b) Cyclic difference voltammograms of
25/50/75/100 μM potassium ferricyanide with the background current
subtracted.Figure presents
the time course of the reduction peak current amplitude (Figure b) of each cycle
during the recording. Each dot represents the currents averaged over
10 ms at the peak position (∼0.138–0.148 V). The blue
dots indicate the cycles that were used to generate current averages
for the corresponding ferricyanide concentrations in Figure . Upon solution exchange, the
first few cycles showed deviating data points due to the mixing of
the solutions and mechanical vibrations. Subsequently, peak currents
settled to a stable level for the remaining cycles. Figure clearly shows the step increases
with increasing concentrations of ferricyanide.
Figure 9
Time course of peak currents
during the forward scans of the recording.
Each dot indicates the reduction peak current of each cycle. The blue
dots indicate cycles that were used to generate the average current
for the corresponding ferricyanide concentration in Figure . The horizontal axis indicates
the start time of each cycle.
Time course of peak currents
during the forward scans of the recording.
Each dot indicates the reduction peak current of each cycle. The blue
dots indicate cycles that were used to generate the average current
for the corresponding ferricyanide concentration in Figure . The horizontal axis indicates
the start time of each cycle.The reduction/oxidation peak values are plotted in Figure (black dots) as a function
of ferricyanide concentration together with linear regression lines
(black lines). The linear fits were performed with the intercepts
fixed at zero. Both oxidation and reduction peaks showed good linearity
against concentrations of the analyte.
Figure 10
Reduction (a) and oxidation
(b) peak values (black dots) and linear
regression lines (black). The red dashed line in (a) shows the theoretical
line determined by eq .
Reduction (a) and oxidation
(b) peak values (black dots) and linear
regression lines (black). The red dashed line in (a) shows the theoretical
line determined by eq .Theoretically, the peak current
for the forward scan (reduction)
described by the Randles–Sevcik equation at 25 °C is[16]where n is the number of
electrons transferred in the redox reaction, A is
the area of the working electrode, D is the diffusion
coefficient, C is the ferricyanide concentration,
and ν is the scan rate. With n for potassium
ferricyanide equal to 1, the radius of the working electrode equal
to 8 μm, D of ferricyanide equal to 7.2 ×
10–6 cm2/s, and v at
1 V/s, the dashed red line with a slope of −1.448 pA/μM,
is calculated (Figure a). The measured peak current for 100 μM potassium ferricyanide
was −135.2 pA, a 6.6% deviation from the theoretically expected
value (−144.8 pA). The slope of the linear fit of the reduction
peak values (Figure a, black line) was −1.318 ± 0.066 pA/μM, a 9.0%
deviation compared to the theoretical value. The small deviation may
originate from incomplete solution exchange resulting in slightly
reduced actual ferricyanide concentrations or slightly different electrode
areas.One of the advantages of the CMOS-based detector array
is that
a large number of electrodes can be integrated on one single chip
and operated in parallel. Figure shows results from one channel on the chip with 64
electrodes. Five out of 64 electrodes showed a different response
(Figure S1) and were therefore not included
in this figure. Two of the excluded electrodes showed no response,
while the other 3 showed a rather normal reduction peak but a distorted
current during the oxidation sweep. These distortions were presumably
due to contamination of these three electrodes, which were located
adjacent to each other. The overlapped voltammograms in Figure show very close
positions and amplitudes of the oxidation and reduction peaks. The
oxidation and reduction peak values are (mean ± standard deviation)
127.10 ± 4.03 and 130.28 ± 4.61 pA, respectively.
Figure 11
Superimposed
voltammograms for 100 μM potassium ferricyanide
for a single channel with 64 electrodes. Five out of 64 electrodes
were not giving reasonable results, and are not included in the graph
(Figure S1). The black trace indicates
the voltammogram for 100 μM potassium ferricyanide in Figure b.
Superimposed
voltammograms for 100 μM potassium ferricyanide
for a single channel with 64 electrodes. Five out of 64 electrodes
were not giving reasonable results, and are not included in the graph
(Figure S1). The black trace indicates
the voltammogram for 100 μM potassium ferricyanide in Figure b.
Discussion
In amperometry experiments,
it has been observed that different
electrode materials exhibit different detection sensitivities. For
example, gold electrodes have been found to have higher detection
efficiency compared with indium tin oxide (ITO) electrodes.[17] A conducting polymer, PDOT:PSS, used as electrode
material shows a 2-fold increment in the quantal size detected than
the planar Pt electrode.[18] Cyclic voltammetry
(CV) can be used to characterize electrode properties, such as electrochemical
sensitivity, effective electrode area, stability, and noise.[14] The CMOS device with CV measurement capability
described here is an ideal platform for on-chip electrode characterization
as well as amperometric in vitro recording of single vesicle release
events from live cells as previously reported.[9] With 1064 working electrodes integrated on one chip, these measurements
can be performed with high throughput.The small distortion
of the rapid current change around the transition
from negative scan to positive scan (Figure b) is likely due to a distortion of the actual
electrode voltage VE. To further reduce
the delayed response in Figure b and improve the slew rate of VOA in Figure c,d, the
half-shared amplifiers, A0 and A1 in Figure , could be redesigned and improved for better performance.
However, this distortion is restricted to a very narrow range around
the transition point such that the shapes of difference voltammograms
and reduction and oxidation peaks (Figure b) are not affected.The dependence
of reduction peaks on ferricyanide concentration
is very close to the theoretically predicted values (Figure ). The average RMS noise for
the baseline current with buffer added was 217 fA at 2 kS/s, which
is very similar to 200 fA at 2 kS/s RMS noise level for the amperometry
version of the chip, which measures only positive currents.[9]Postfabrication is required to enable electrochemical
measurement
by CMOS devices as polarizable electrodes are not presented on chip
upon delivery. In this work, platinum electrodes were deposited on
devices with surface-modified structures to increase the reliability
of the chips. Using clean-room technologies, it is also possible to
deposit various other materials as working electrodes, such as Au,
diamond-like carbon,[17,19] indium tin oxide,[17,20] boron-doped diamond,[21,22] graphitic electrodes embedded
in a diamond substrate,[21,23,24] and PEDOT:PSS,[18] to study the electrochemical
properties of these materials as working electrodes.CV measurements
were successfully performed on the CMOS sensor
array using the ferricyanide/ferrocyanide system, which is generally
used as the gold standard for electrode characterization. Calibration
measurements using DA suffer from strong adsorption of DA to Pt electrodes,
which could only be removed when 0.18 M H2SO4 was added and the electrode potential was scanned at a rate of 100
mV/s between −500 and +975 mV for up to 1 h.[25]The ferricyanide CV results deviated <10% from
the theoretically
expected values. Microfluidic devices could be potentially fabricated
on the surface of the CMOS chip, enabling more precise measurement,
high throughput, and improved reproducibility.[26] A two-layer structure for the microfluidic device was developed,[27] which makes it possible to fabricate the microfluidic
devices on surface-modified structures.CV measurements can
also be used to identify faulty electrodes.
Five out of 64 electrodes were identified that did not show proper
responses (Figure S1) and were excluded
from Figure . Variation
in the semiconductor fabrication processes or postfabrication processes
may result in faulty electrodes that cannot measure current precisely.
The capability to perform simultaneous CV characterization of all
electrodes in an array allows for the identification of any faulty
electrodes such that in subsequent amperometry recordings data from
these electrodes will be excluded from the analysis.For FSCV
recordings of quantal release events, the higher scan
rate generates much larger currents and the gain of the circuit needs
to be reduced. This can be achieved by increasing the size of the
integration capacitor. Using the more advanced On Semi C18 (0.18 μm)
technology would allow a 10-fold increase of the integration capacitance
without significantly increasing the pixel size.In this work,
a CMOS device with bidirectional-current measurement
capability has been demonstrated. The compact circuit of each detector
pixel consists of only 11 transistors, significantly decreasing the
required area for each detector. The sign binary output implemented
provides an accurate knowledge of the polarity of the measured current.
A dopamine injection test revealed that the chip can be used to detect
electroactive species and can potentially be used for single-cell
amperometry. A CV experiment with potassium ferricyanide validated
the CMOS chip for performing CV measurement and generating voltammograms.
The oxidation and reduction peak values showed a linear relationship
with analyte concentrations. Five faulty electrodes were identified
and excluded from the analysis.
Authors: Yogesh S Singh; Lauren E Sawarynski; Heather M Michael; Robert E Ferrell; Michael A Murphey-Corb; Greg M Swain; Bhavik A Patel; Anne M Andrews Journal: ACS Chem Neurosci Date: 2010-01-20 Impact factor: 4.418
Authors: Kevin A White; Geoffrey Mulberry; Jonhoi Smith; Manfred Lindau; Bradley A Minch; Kiminobu Sugaya; Brian N Kim Journal: IEEE Trans Biomed Circuits Syst Date: 2018-07-30 Impact factor: 3.833