| Literature DB >> 33977330 |
Sebastian Zensen1, Sumitha Selvaretnam2, Marcel Opitz2, Denise Bos2, Johannes Haubold2, Jens Theysohn2, Michael Forsting2, Nika Guberina3, Axel Wetter2,4.
Abstract
PURPOSE: Apart from the commonly applied manual needle biopsy, CT-guided percutaneous biopsies of bone lesions can be performed with battery-powered drill biopsy systems. Due to assumably different radiation doses and procedural durations, the aim of this study is to examine radiation exposure and establish local diagnostic reference levels (DRLs) of CT-guided bone biopsies of different anatomical regions.Entities:
Keywords: bone biopsy; computed tomography; diagnostic reference level; radiation exposure
Mesh:
Year: 2021 PMID: 33977330 PMCID: PMC8382636 DOI: 10.1007/s00270-021-02851-z
Source DB: PubMed Journal: Cardiovasc Intervent Radiol ISSN: 0174-1551 Impact factor: 2.740
Technical parameters of CT-guided percutaneous bone biopsy at three different multi-slice Siemens CT scanners
| CT scanner* | Femur | Ilium | Sacrum | Thoracic vertebrae | Lumbar vertebrae | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| AS + | Flash | Force | AS + | Flash | Force | AS + | Flash | Force | AS + | Flash | Force | AS + | Flash | Force | |
| 14 | 11 | 2 | 16 | 45 | 13 | 4 | 17 | 6 | 11 | 15 | 2 | 8 | 18 | 5 | |
| Tube current–time product (mAs) | 50 | 50 | 150 | 50 | 50 | 150 | 50 | 50 | 150 | 50 | 50 | 165 | 50 | 50 | 150 |
| Slice thickness (mm) | 1.2 | 1.2 | 5.0 | 1.2 | 1.2 | 5.0 | 1.2 | 1.2 | 5.0 | 1.2 | 1.2 | 1.0/5.0 | 1.2 | 1.2 | 2.5 |
*AS + : SOMATOM Definition AS + ; Flash: SOMATOM Definition Flash; Force: SOMATOM Force (all: Siemens Healthineers, Erlangen, Germany)
Fig. 1Manual and powered drill bone biopsy systems. Shown in A is a typical commercially available manual bone biopsy kit consisting of the biopsy needle (right) and an obturator (left) for pushing the specimen out of the needle. Shown in B is the reusable, non-sterile battery-powered drill (Arrow OnControl, Teleflex, Wayne, USA) (right), which was wrapped in a sterile bag prior to biopsy, and an associated disposable 11-gauge 4 inch (102 mm) biopsy needle (center) and an obturator (left) for pushing the specimen out of the needle
Volume-weighted CT dose index (CTDIvol) of CT-guided percutaneous bone biopsy with manual and battery-powered drill system for different anatomical regions
| Anatomical location | n | CTDIvol [mGy] | ||||
|---|---|---|---|---|---|---|
| 25th percentile | Median | 75th percentile | Mean ± SD | |||
| Femur | Total | 27 | 42.4 | 65.4 | 90.9 | 71.0 ± 43.1 |
| Manual | 13 | 41.4 | 56.9 | 128.5 | 80.9 ± 47.7 | |
| Powered drill | 14 | 37.6 | 66.7 | 76.2 | 61.8 ± 37.8 | |
| Ilium | Total | 74 | 50.4 | 71.5 | 90.3 | 80.0 ± 41.7 |
| Manual | 27 | 62.1 | 83.5 | 128.5 | 101.3 ± 49.2 | |
| Powered drill | 47 | 46.2 | 59.4 | 79.8 | 67.8 ± 31.1 | |
| Sacrum | Total | 27 | 58.4 | 71.6 | 115.5 | 92.3 ± 59.2 |
| Manual | 11 | 71.6 | 97.5 | 149.2 | 124.9 ± 76.6 | |
| Powered drill | 16 | 49.1 | 63.1 | 83.7 | 69.9 ± 29.0 | |
| Thoracic vertebrae | Total | 28 | 34.1 | 48.3 | 86.6 | 64.6 ± 49.6 |
| Manual | 13 | 40.3 | 67.0 | 86.6 | 75.7 ± 61.4 | |
| Powered drill | 15 | 29.9 | 39.7 | 89.0 | 54.9 ± 35.9 | |
| Lumbar vertebrae | Total | 31 | 52.0 | 70.9 | 115.0 | 91.1 ± 68.2 |
| Manual | 10 | 66.5 | 100.1 | 162.6 | 133.0 ± 97.7 | |
| Powered drill | 21 | 48.0 | 62.5 | 90.0 | 71.2 ± 37.3 | |
Dose–length product (DLP) of CT-guided percutaneous bone biopsy with manual and battery-powered drill system for different anatomical regions
| Anatomical location | n | DLP [mGy·cm] | ||||
|---|---|---|---|---|---|---|
| 25th percentile | Median | 75th percentile | Mean ± SD | |||
| Femur | Total | 27 | 179.3 | 332.0 | 522.0 | 366.3 ± 212.6 |
| Manual | 13 | 203.1 | 410.0 | 682.8 | 442.5 ± 250.3 | |
| Powered drill | 14 | 127.7 | 302.8 | 403.1 | 295.5 ± 146.3 | |
| Ilium | Total | 74 | 327.7 | 419.8 | 529.8 | 458.9 ± 211.1 |
| Manual | 27 | 322.0 | 489.0 | 471.8 | 524.7 ± 255.0 | |
| Powered drill | 47 | 328.5 | 399.7 | 695.0 | 421.0 ± 173.2 | |
| Sacrum | Total | 27 | 341.0 | 492.6 | 740.0 | 537.0 ± 248.2 |
| Manual | 11 | 496.0 | 626.6 | 740.0 | 659.3 ± 287.6 | |
| Powered drill | 16 | 316.1 | 403.7 | 515.3 | 453.0 ± 182.0 | |
| Thoracic vertebrae | Total | 28 | 193.5 | 315.7 | 577.7 | 418.3 ± 331.6 |
| Manual | 13 | 267.0 | 392.0 | 596.0 | 473.3 ± 316.3 | |
| Powered drill | 15 | 165.8 | 206.9 | 402.1 | 370.6 ± 347.9 | |
| Lumbar vertebrae | Total | 31 | 263.4 | 402.0 | 545.8 | 437.2 ± 259.2 |
| Manual | 10 | 384.8 | 520.5 | 619.0 | 559.0 ± 279.8 | |
| Powered drill | 21 | 240.4 | 314.5 | 452.1 | 379.2 ± 233.7 | |
Fig. 2Distribution of volume-weighted CT dose index (CTDIvol) and dose–length product (DLP) of CT-guided percutaneous bone biopsy with manual and battery-powered drill system for different anatomical locations. Highlighting radiation exposure indices of biopsy-guiding scans with exclusion of radiation exposure of topogram, pre- and post-bioptic scans to emphasize radiation exposure differences between manual and battery-powered drill bone biopsy approaches. Whiskers represent min to max. Asterisks indicate significant difference (p < 0.05)
Duration of CT-guided percutaneous bone biopsy with manual versus battery-powered drill system at different anatomical regions
| Region | n | Duration [min] | ||
|---|---|---|---|---|
| Median | IQR | |||
| Femur | Total | 27 | 22.7 | 16.4–27.9 |
| Manual | 13 | 24.7 | 18.6–40.5 | |
| Powered drill | 14 | 22.5 | 16.3–26.3 | |
| Ilium | Total | 74 | 21.0 | 16.6–27.7 |
| Manual | 27 | 23.5 | 16.8–31.3 | |
| Powered drill | 47 | 19.2 | 15.0–26.8 | |
| Sacrum | Total | 27 | 21.9 | 15.7–26.4 |
| Manual | 11 | 25.1 | 15.7–29.2 | |
| Powered drill | 16 | 20.6 | 15.9–24.2 | |
| Thoracic vertebrae | Total | 28 | 22.0 | 11.5–31.7 |
| Manual | 13 | 23.4 | 14.4–31.7 | |
| Powered drill | 15 | 18.7 | 9.7–32.7 | |
| Lumbar vertebrae | Total | 31 | 23.3 | 17.5–32.1 |
| Manual | 10 | 28.7 | 16.9–39.7 | |
| Powered drill | 21 | 22.1 | 18.1–30.3 | |