Preliminary In Vitro Wear Assessment of Ceramic Cemented Femoral Components Coupled with Polyethylene Menisci.

Saverio Affatato1, Paolo Erani1, Maurizio Fersini2, Vincenzo Contaldi2, Anna Rita Terrizzi3, Antonio Licciulli3.   

Abstract

Success of total knee replacement (TKR) depends on the prosthetic design and materials. The use of metal components is well established with the disadvantage of allergic reactions. Ceramics have been recently proposed because of high wear resistance, excellent biocompatibility, wettability, and suitable mechanical properties. This study was aimed at investigating in vitro wear resistance of Zirconia Toughened Alumina (ZTA)/Ultra-high-molecular-weight polyethylene (UHMWPE) of TKR femoral components. An in vitro protocol was designed with the application of relevant load profile, 6-degrees-of-freedom knee simulator, and 8 × 105 cycles on the ZTA/UHMWPE configuration under bovine calf serum. Before and after wear test, the femoral components were investigated by using the Scanning Electron Microscope (SEM) and the X-Ray Diffraction (XRD) analyses, and stylus surface roughness measurements. The proposed pre-clinical test yielded repeatable results. In particular, gravimetric results showed that, after 8 × 105 cycles, the mean weight loss of the polyethylene mobile components is 5.3 ± 1.1 mg. The surface roughness measurements (Ramax) performed after the wear test showed no significant variation on the UHMWPE menisci. A slight increase of roughness has been found on the ZTA (0.02 µm before wear test, 0.28 µm after the test). SEM observations did not show significant modification of the surface morphology. Tetragonal to monoclinic phase ratio was measured by XRD before and after wear test to evaluate stability of tetragonal ZrO2 phase. Minimal conversion of tetragonal to monoclinic phase was found from 5.4 to 8%. Although this study is a preliminary evaluation limited to in vitro tests, it provides novel pre-clinical indications about the potential of ceramic TKR femoral components.

Entities:  

Keywords:  SEM; X-ray diffractometry; bovine calf serum; ceramic composite; knee arthroplasty; knee simulator; roughness measurements; wear

Year:  2021        PMID: 33921972     DOI: 10.3390/ma14092112

Source DB:  PubMed          Journal:  Materials (Basel)        ISSN: 1996-1944            Impact factor:   3.623


  12 in total

1.  Mobile or fixed unicompartmental knee prostheses? In-vitro wear assessments to solve this dilemma.

Authors:  P Taddei; E Modena; Thomas M Grupp; Saverio Affatato
Journal:  J Mech Behav Biomed Mater       Date:  2011-06-26

2.  Wear simulation of total knee prostheses using load and kinematics waveforms from stair climbing.

Authors:  Sami Abdel-Jaber; Claudio Belvedere; Alberto Leardini; Saverio Affatato
Journal:  J Biomech       Date:  2015-09-21       Impact factor: 2.712

3.  Chemically driven tetragonal-to-monoclinic polymorphic transformation in retrieved ZTA femoral heads from dual mobility hip implants.

Authors:  M Boffelli; A Doimo; E Marin; L Puppulin; W Zhu; N Sugano; I C Clarke; G Pezzotti
Journal:  J Mech Behav Biomed Mater       Date:  2015-12-05

4.  Investigation on wear of knee prostheses under fixed kinematic conditions.

Authors:  Saverio Affatato; Walther Leardini; Mirko Rocchi; Aldo Toni; Marco Viceconti
Journal:  Artif Organs       Date:  2008-01       Impact factor: 3.094

5.  Implant fixation in knee replacement: preliminary in vitro comparison of ceramic and metal cemented femoral components.

Authors:  Luca Cristofolini; Saverio Affatato; Paolo Erani; Domenico Tigani; Marco Viceconti
Journal:  Knee       Date:  2008-12-06       Impact factor: 2.199

6.  Displacement or Force Control Knee Simulators? Variations in Kinematics and in Wear.

Authors:  Saverio Affatato
Journal:  Artif Organs       Date:  2015-07-06       Impact factor: 3.094

7.  Third-body abrasive wear of tibial polyethylene inserts combined with metallic and ceramic femoral components in a knee simulator study.

Authors:  Carmen Zietz; Philipp Bergschmidt; Regina Lange; Wolfram Mittelmeier; Rainer Bader
Journal:  Int J Artif Organs       Date:  2013-01       Impact factor: 1.595

8.  Metal hypersensitivity in total knee arthroplasty: revision surgery using a ceramic femoral component - a case report.

Authors:  Philipp Bergschmidt; Rainer Bader; Wolfram Mittelmeier
Journal:  Knee       Date:  2011-02-02       Impact factor: 2.199

9.  In vitro effects on mobile polyethylene insert under highly demanding daily activities: stair climbing.

Authors:  Sami Abdel Jaber; Paola Taddei; Silvia Tozzi; Alessandra Sudanese; Saverio Affatato
Journal:  Int Orthop       Date:  2014-12-14       Impact factor: 3.075

10.  Ceramic femoral components in total knee arthroplasty - two year follow-up results of an international prospective multi-centre study.

Authors:  Philipp Bergschmidt; Rainer Bader; Dirk Ganzer; Christian Hauzeur; Christoph Lohmann; Wolfgang Rüther; Domenico Tigani; Nicola Rani; Fernando Lopez Prats; Claudio Zorzi; Vincenzo Madonna; Stefano Rigotti; Francesco Benazzo; Stefano Marco Paolo Rossi; Guenther Kundt; Hans Rudolf Bloch; Wolfram Mittelmeier
Journal:  Open Orthop J       Date:  2012-04-20
View more
  1 in total

1.  Evaluation of temperature of a full ceramic total knee arthroplasty during MRI examinations.

Authors:  Klemens Trieb; Andreas Artmann; Michael Krupa; Sasch Senck; Franz Landauer
Journal:  Medicine (Baltimore)       Date:  2022-09-30       Impact factor: 1.817

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.