Literature DB >> 26147451

Displacement or Force Control Knee Simulators? Variations in Kinematics and in Wear.

Saverio Affatato1.   

Abstract

The problems associated with prosthetic failure and revision surgery still constitute the main clinical problem of prosthetic surgery. The objective of wear evaluation is to determine the wear rate and its dependence on the test conditions. To obtain realistic results, a wear test can be performed to reproduce in vivo working conditions and compare the wear characteristics of various total knee prostheses designs. At the state of the art, two simulation concepts are available and defined in ISO 14243 standards series. In both these guidelines, level walking is the sole activity of daily living that is represented for testing. With so many variables and so many sources of error and the sensitivity of the output to these errors, can the motion determined in the simulator be representative of the in vivo motion? This article goes beyond the current status of these knee simulations comparing literature results.
Copyright © 2015 International Center for Artificial Organs and Transplantation and Wiley Periodicals, Inc.

Entities:  

Keywords:  Anterior/posterior translation; Axial force; Displacement control; Force control; Internal/external flexion-extension rotations; Knee simulator

Mesh:

Year:  2015        PMID: 26147451     DOI: 10.1111/aor.12508

Source DB:  PubMed          Journal:  Artif Organs        ISSN: 0160-564X            Impact factor:   3.094


  5 in total

1.  Preliminary In Vitro Wear Assessment of Ceramic Cemented Femoral Components Coupled with Polyethylene Menisci.

Authors:  Saverio Affatato; Paolo Erani; Maurizio Fersini; Vincenzo Contaldi; Anna Rita Terrizzi; Antonio Licciulli
Journal:  Materials (Basel)       Date:  2021-04-22       Impact factor: 3.623

2.  Influence of conformity on the wear of total knee replacement: An experimental study.

Authors:  Claire L Brockett; Silvia Carbone; John Fisher; Louise M Jennings
Journal:  Proc Inst Mech Eng H       Date:  2017-12-17       Impact factor: 1.617

3.  Influence of tibiofemoral congruency design on the wear of patient-specific unicompartmental knee arthroplasty using finite element analysis.

Authors:  Y-G Koh; K-M Park; H-Y Lee; K-T Kang
Journal:  Bone Joint Res       Date:  2019-04-02       Impact factor: 5.853

4.  Wear Behavior Characterization of Hydrogels Constructs for Cartilage Tissue Replacement.

Authors:  Saverio Affatato; Diego Trucco; Paola Taddei; Lorenzo Vannozzi; Leonardo Ricotti; Gilbert Daniel Nessim; Gina Lisignoli
Journal:  Materials (Basel)       Date:  2021-01-16       Impact factor: 3.623

5.  The impact of variations in input directions according to ISO 14243 on wearing of knee prostheses.

Authors:  Xiao-Hong Wang; Wei Zhang; Da-Yong Song; Hui Li; Xiang Dong; Min Zhang; Feng Zhao; Zhong-Min Jin; Cheng-Kung Cheng
Journal:  PLoS One       Date:  2018-10-29       Impact factor: 3.240