A comparison of the Goldmann applanation and non-contact (Keeler Pulsair EasyEye) tonometers and the effect of central corneal thickness in indigenous African eyes.

O E Babalola1, A V Kehinde, A C Iloegbunam, T Akinbinu, C Moghalu, I Onuoha.   

Abstract

UNLABELLED: The Keeler Pulsair EasyEye non-contact tonometer (NCT) was introduced into practice at Rachel Eye Center Abuja, Nigeria, where the patients are indigenous Africans. This was compared to the 'gold standard' Goldmann applanation tonometer (GAT) to determine if the instrument was accurate in Africans, with particular reference to the influence of central corneal thickness (CCT). PATIENTS AND METHODS: 174 eyes of 88 patients were analysed. Pachymetry was performed using Sonomed PacScan AP300, and GAT with the Haag Streit R-900. Pachymetric corrections of NCT (NCTc) and GAT (GATc) were carried out with the Sonomed algorithm. Pearson's correlation r, linear regression analysis, Student t-test and Bland-Altman analysis were used to compare the instruments.
RESULTS: Mean NCT readings were similar (17.36 mmHg) to mean GAT (17.42 mmHg; p = 0.769). GAT/NCT correlation coefficient, r, was 0.883 as compared with 0.868 for GATc/NCTc. The linear regression equation was GAT = 2.79 + 0.84*NCT (r(2) = 0.78). Forty-five per cent of differences were within 1 mmHg, while 79% were within 3 mmHg. This was similar to findings in some studies on Caucasians. Bland-Altman analysis however suggested that the spread of differences was wider than in those studies. Outliers (differences more than 5 mmHg) sometimes reflected difficulties encountered with GAT in routine practice. Mean CCT was 537.9 microm, (S.D. 38.4, 95% confidence interval 532.1-543.7 microm) and CCT appeared lower than in Caucasians. Both GAT and NCT IOP tended to rise with increasing CCT but NCT had a greater tendency to do so. Regression analysis suggested that NCT IOP increased by 0.6 mmHg for every 10 mum increase in CCT, compared to 0.4 mmHg for GAT. Thirty-eight per cent of the patients preferred NCT as opposed to 25% GAT.
CONCLUSIONS: The Keeler Pulsair EasyEye gives reliable measurements of IOP in African eyes but is significantly affected by CCT. Particularly in borderline cases where management decisions have to be taken, it may be necessary to have pachymetric corrections based on an NCT algorithm, which appears steeper than the GAT algorithm.

Entities:  

Mesh:

Year:  2009        PMID: 19236588     DOI: 10.1111/j.1475-1313.2008.00621.x

Source DB:  PubMed          Journal:  Ophthalmic Physiol Opt        ISSN: 0275-5408            Impact factor:   3.117


  10 in total

1.  Central corneal thickness in southern Egypt.

Authors:  Engy Mohamed Mostafa
Journal:  Int Ophthalmol       Date:  2013-11-22       Impact factor: 2.031

2.  Clinical comparison of Pulsair non-contact tonometer and Goldmann applanation tonometer in Indian population.

Authors:  Shalini Mohan; Satyaprakash Tiwari; Arvind Jain; Jaya Gupta; Surendra Kumar Sachan
Journal:  J Optom       Date:  2013-07-26

3.  Purinergic dysregulation causes hypertensive glaucoma-like optic neuropathy.

Authors:  Youichi Shinozaki; Kenji Kashiwagi; Kazuhiko Namekata; Akiko Takeda; Nobuhiko Ohno; Bernard Robaye; Takayuki Harada; Takeshi Iwata; Schuichi Koizumi
Journal:  JCI Insight       Date:  2017-10-05

4.  Comparison of intraocular pressure as measured by three different non-contact tonometers and goldmann applanation tonometer for non-glaucomatous subjects.

Authors:  Seung Pil Bang; Chong Eun Lee; Yu Cheol Kim
Journal:  BMC Ophthalmol       Date:  2017-11-02       Impact factor: 2.209

5.  Comparison of current tonometry techniques in measurement of intraocular pressure.

Authors:  Behrooz Kouchaki; Hassan Hashemi; Abbasali Yekta; Mehdi Khabazkhoob
Journal:  J Curr Ophthalmol       Date:  2016-11-28

6.  Comparative Assessment of The Goldmann Applanation and Noncontact Tonometers in Intraocular Pressure Measurements in a Sample of Glaucoma Patients in the Cape Coast Metropolis, Ghana.

Authors:  Samuel Kyei; Cynthia Pakyennu Gboglu; Michael Agyemang Kwarteng; Frank Assiamah
Journal:  Niger Med J       Date:  2020-12-19

7.  Comparison of Non-contact Tonometry and Goldmann Applanation Tonometry Measurements in Non-pathologic High Myopia.

Authors:  Peiyuan Wang; Yunhe Song; Fengbin Lin; Zhenyu Wang; Xinbo Gao; Weijing Cheng; Meiling Chen; Yuying Peng; Yuhong Liu; Xiulan Zhang; Shida Chen
Journal:  Front Med (Lausanne)       Date:  2022-03-03

8.  Association of intraocular pressure and postoperative nausea and vomiting after microvascular decompression - a prospective cohort study.

Authors:  Yuantao Hou; Hansheng Liang; Cungang Fan; Ruen Liu; Yi Feng
Journal:  BMC Anesthesiol       Date:  2022-04-30       Impact factor: 2.217

9.  Micropulse diode laser trabeculoplasty in Nigerian patients.

Authors:  Olufemi Emmanuel Babalola
Journal:  Clin Ophthalmol       Date:  2015-07-20

10.  Comparison of the corrected intraocular pressure by tonopachy with that of Goldmann applanation tonometry in normal and glaucomatous patients.

Authors:  Girish Velis; Srinivasan Kavitha; Nazlee Zebardast; Sabyasachi Sengupta; Rengaraj Venkatesh
Journal:  Indian J Ophthalmol       Date:  2020-04       Impact factor: 1.848

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.