The influence of soft contact lenses on the intraocular pressure measurement.

P G Firat1, C Cankaya, S Doganay, M Cavdar, S Duman, E Ozsoy, B Koc.   

Abstract

PURPOSE: To evaluate the influence of silicone hydrogel contact lenses on the intraocular pressure (IOP) measurement using Goldmann applanation tonometry (GAT), non-contact tonometry (NCT), and Pascal dynamic contour tonometry (DCT).
METHODS: We included in the study 40 eyes of 40 patients who did not have any ocular or systemic diseases or contraindications to contact lens use. We measured and recorded the IOP values of each patient using NCT without and with contact lenses (groups 1 and 2, respectively), using DCT without and with contact lenses (groups 3 and 4, respectively), and using GAT without contact lenses (group 5).
RESULTS: The mean IOP value of group 1 was 14.55 ± 2.95 mm Hg and 13.92 ± 2.58 mm Hg in group 2. We detected no statistically significant difference between group 1 and group 2 (P=0.053). The mean IOP values for group 3 and group 4 were 16.26 ± 2.33 mm Hg and 15.19 ± 2.40 mm Hg, respectively. We detected a statistically significant difference between groups 3 and 4 (P=0.005). Group 5's mean IOP value was 12.97 ± 2.65 mm Hg. IOP values measured with DCT were statistically significantly higher compared with IOP values measured with NCT and GAT (P<0.0001 and P<0.0001, respectively). Additionally, IOP values measured with NCT were statistically significantly higher compared with IOP values measured with GAT (P<0.0001).
CONCLUSION: According to the results of our study, silicone hydrogel soft contact lens use does not significantly affect IOP values measured with NCT, but it affects IOP values measured with DCT.

Entities:  

Mesh:

Year:  2011        PMID: 22079968      PMCID: PMC3272185          DOI: 10.1038/eye.2011.271

Source DB:  PubMed          Journal:  Eye (Lond)        ISSN: 0950-222X            Impact factor:   3.775


  24 in total

1.  The validity of a new noncontact tonometer and its comparison with the Goldmann tonometer.

Authors:  Andrew K C Lam; Rufina Chan; Chin-hang Lam
Journal:  Optom Vis Sci       Date:  2004-08       Impact factor: 1.973

Review 2.  The use of contact lenses in the treatment of persistent epithelial defects.

Authors:  Sally Jayne Blackmore
Journal:  Cont Lens Anterior Eye       Date:  2010-07-17       Impact factor: 3.077

3.  Validity and limitations of the Nidek NT-4000 non-contact tonometer: a clinical study.

Authors:  Federico Regine; Gian Luca Scuderi; Massimo Cesareo; Federico Ricci; Claudio Cedrone; Carlo Nucci
Journal:  Ophthalmic Physiol Opt       Date:  2006-01       Impact factor: 3.117

4.  Comparison of intraocular pressure measured by Pascal dynamic contour tonometry and Goldmann applanation tonometry.

Authors:  J Y F Ku; H V Danesh-Meyer; J P Craig; G D Gamble; C N J McGhee
Journal:  Eye (Lond)       Date:  2006-02       Impact factor: 3.775

5.  Performance of the rebound, noncontact and Goldmann applanation tonometers in routine clinical practice.

Authors:  Jose M Martinez-de-la-Casa; Maria Jimenez-Santos; Federico Saenz-Frances; Maria Matilla-Rodero; Carmen Mendez-Hernandez; Rocio Herrero-Vanrell; Julian Garcia-Feijoo
Journal:  Acta Ophthalmol       Date:  2009-11-09       Impact factor: 3.761

6.  Intraocular pressure measurement with the noncontact tonometer through soft contact lenses.

Authors:  Yi-Chun Liu; Jehn-Yu Huang; I-Jong Wang; Fung-Rong Hu; Yu-Chih Hou
Journal:  J Glaucoma       Date:  2011-03       Impact factor: 2.503

7.  Accurate intraocular pressure measurement in contact lens wearers with normal pressures.

Authors:  L Lim; T P Ng; D T Tan
Journal:  CLAO J       Date:  1997-04

8.  Accuracy of dynamic contour tonometry compared with applanation tonometry in human cadaver eyes of different hydration states.

Authors:  Christoph Kniestedt; Michelle Nee; Robert L Stamper
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2004-10-13       Impact factor: 3.117

9.  [Effect of intraocular pressure measurement through therapeutic soft contact lenses by noncontact tonometer].

Authors:  R Sugimoto-Takeuchi; R Yamamoto; Y Kuwayama; S Kinoshita
Journal:  Nippon Ganka Gakkai Zasshi       Date:  1991-09

10.  The effect of contact lens-induced corneal edema on Goldmann applanation tonometry and dynamic contour tonometry.

Authors:  Jong-Hyun Oh; Chungkwon Yoo; Yong Yeon Kim; Hyo-Myung Kim; Jong-Suk Song
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2008-10-09       Impact factor: 3.117

View more
  6 in total

1.  Comparative measurement of intraocular pressure by Icare tonometry and Airpuff tonometry in healthy subjects and patients wearing therapeutic soft contact lenses.

Authors:  Alexandra Anton; Matthias Neuburger; Daniel Böhringer; Jens F Jordan
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2013-03-28       Impact factor: 3.117

2.  Comparative evaluation of intraocular pressure with an air-puff tonometer versus a Goldmann applanation tonometer.

Authors:  Qasim K Farhood
Journal:  Clin Ophthalmol       Date:  2012-12-27

3.  Comparison of intraocular pressure measurement with Scheimpflug-based noncontact tonometer with and without hydrogel contact lenses.

Authors:  Mukesh Kumar; Rohit Shetty; Chaitra Jayadev; Debarun Dutta; Maneck D Nicolsan; Sriharsha Nagaraj; Rajesh S Kumar
Journal:  Indian J Ophthalmol       Date:  2015-04       Impact factor: 1.848

4.  Comparative Assessment of The Goldmann Applanation and Noncontact Tonometers in Intraocular Pressure Measurements in a Sample of Glaucoma Patients in the Cape Coast Metropolis, Ghana.

Authors:  Samuel Kyei; Cynthia Pakyennu Gboglu; Michael Agyemang Kwarteng; Frank Assiamah
Journal:  Niger Med J       Date:  2020-12-19

5.  Intraocular pressure readings obtained through soft contact lenses using four types of tonometer.

Authors:  Joji Takenaka; Eriko Kunihara; Ulfah Rimayanti; Junko Tanaka; Makoto Kaneko; Yoshiaki Kiuchi
Journal:  Clin Ophthalmol       Date:  2015-10-05

6.  The measurement of intraocular pressure over positive soft contact lenses by rebound tonometry.

Authors:  Fabrizio Zeri; Mario De Cusatis; Luigi Lupelli; Peter Graham Swann
Journal:  J Optom       Date:  2015-10-21
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.