| Literature DB >> 33806531 |
Do Jung Kim1, Sak Lee2, Hyun-Chel Joo2, Young-Nam Youn2, Kyung-Jong Yoo2, Seung Hyun Lee2.
Abstract
The Trifecta valve has externally mounted leaflets; it differs from classic internally mounted valves (e.g., Carpentier-Edwards [C-E]). We evaluated post-implantation hemodynamics and clinical outcomes of these bioprostheses in small aortic annuli. From January 2015 to April 2019, 490 patients who underwent aortic valve replacement (AVR) were reviewed retrospectively. Altogether, 183 patients received 19 or 21 mm diameter C-E (n = 121) or Trifecta (n = 62) prostheses. To minimize confounding variables, we performed propensity-score matching analysis. The mean transvalvular pressure gradient (TVPG) was significantly lower in the Trifecta than in the C-E group at discharge (12.9 ± 4.8 vs. 15.0 ± 5.3 mmHg, p = 0.044). TVPG change over time was not significantly different between groups (p = 0.357). Left ventricular mass index decreased postoperatively (reduction: C-E, 28.1%; Trifecta, 30.1%, p = 0.879). No late mortality, severe patient-prosthesis mismatch, moderate-to-severe paravalvular leakage, structural valve degeneration, or valve thromboses were observed. Freedom from valve-related events at 3 years were similar for C-E (97.9% ± 2.1%) and Trifecta (97.7% ± 2.2%) patients (log-rank p = 0.993). Bioprosthesis design for small annuli significantly affected TVPG immediately after AVR. However, hemodynamics over time and clinical outcomes did not differ between the two designs.Entities:
Keywords: heart valve prosthesis implantation; hemodynamic monitoring; prosthesis design; small aortic annulus
Year: 2021 PMID: 33806531 PMCID: PMC7961368 DOI: 10.3390/jcm10051063
Source DB: PubMed Journal: J Clin Med ISSN: 2077-0383 Impact factor: 4.241