| Literature DB >> 33674962 |
Marc-Daniel Ahrend1,2, Michael Rühle3, Fabian Springer4, Heiko Baumgartner3.
Abstract
INTRODUCTION: In absence of deformity or injury of the contralateral leg, the contralateral leg length is used to plan limb lengthening. Length variability on long-leg weight-bearing radiographs (LLR) can lead to inaccurate deformity correction. The aim of the study was to (1) examine the variability of the measured limb length on LLR and (2) to examine the influence of the position of the magnification device.Entities:
Keywords: External fixator; Lengthening; Limb length; Long-leg radiographs; TSF
Mesh:
Substances:
Year: 2021 PMID: 33674962 PMCID: PMC9217775 DOI: 10.1007/s00402-021-03831-1
Source DB: PubMed Journal: Arch Orthop Trauma Surg ISSN: 0936-8051 Impact factor: 2.928
Fig. 1Example of the variability of limb length within one patient during bone lengthening. The length of the untreated right leg varies between 747 and 774 mm. Bone lengthening protocols had to be re-calculated due to variability of limb length of both legs on long-leg radiographs leading to prolonged treatment time (approx. 3 months) (mFA-mTA mechanical tibiofemoral angle, MPTA medial proximal tibia angle, mLDFA mechanical lateral distal femur angle, mLDTA mechanical lateral distal tibia angle, mLPFA mechanical lateral proximal femur angle, JLCA joint line convergence angle, AMA anatomical and mechanical femoral angle)
Patient demographics
| Demographics | |
|---|---|
| Gender | Female: 2 (5.3%) Male: 36 (94.7%) |
| Age (years) | 46.2 ± 13.4 (21.3–80.8) |
| Body height (cm) | 176.1 ± 7.7 (160–194) |
| Body weight (kg) | 88.4 ± 19.1 (52–130) |
| Body mass index (kg/m2) | 28.5 ± 5.5 (17.9–41.5) |
| TSF treatment | Deformity correction: 38 (100%) Bone infection: 21 (55.3%) Limb lengthening 33 (86.8%) Segment transport: 8 (21.1%) |
Fig. 2Overview of the radiographic test setup. a Sagittal view: the X-ray source is 295 cm away from the validation distance. The z-position of the magnification cord (including steel ball and magnification distance) is in the middle of the validation distance. Position A is 5 cm anterior, position B 5 cm posterior of the z-position. On the same sagittal level (z-axis) as the validation distance and the z-position, position C is 45 cm proximal and position D is 45 cm distal from the z-position. b Frontal view: the position E is 5 cm medial and position F is 15 cm lateral from the z-position. Positions E, Z and F are on the same sagittal level as the validation distance. c Frontal view of the radiographic film with the magnification cord and the validation cord
The mean ± standard deviation (range) of the limb lengths measured throughout all LLR (*), the mean ± standard deviation (range) of the ranges of the measured limb length within a single patient (#), the mean ± standard deviation (range) of the standard deviations of the measured limb length within a single patient (°)
| Mean limb length of the cohort (*) | Ranges of measured limb length within each patient (#) | Deviations (SD) of measured limb length within each patient (°) | |
|---|---|---|---|
| Observer 1 | |||
| Observation 1 | 833.6 ± 48.4 (748–926) mm | 23.74 ± 8.09 (11–43) mm | 8.41 ± 2.79 (3.97–16.38) mm |
| Observation 2 | 833.9 ± 48.7 (747–924) mm | 24.24 ± 8.45 (12–50) mm | 8.57 ± 2.76 (3.92–16.34) mm |
| Observer 2 | |||
| Observation 1 | 831.6 ± 48.1 (745–924) mm | 25.47 ± 8.60 (10–42) mm | 9.07 ± 2.90 (4.20–17.17) mm |
| Observation 2 | 832.7 ± 48.0 (747–923) mm | 25.08 ± 8.96 (10–47) mm | 9.00 ± 2.81 (4.10–17.90) mm |
Intra- and interobserver reliability [ICC (95% CI)] and intra- and interobserver difference (mean ± standard deviation (minimum–maximum)) of limb length measurements in TSF patients
| Intraobserver reliability | Intraobserver differences—absolute values | Interobserver reliability | Interobserver differences—absolute values | |
|---|---|---|---|---|
| Limb length measurement in TSF patients | Observer 1: 0.998 (0.997–0.998) | Observer 1: 2.78 ± 2.01 (0–10) mm | First observation: 0.995 (0.994–0.996) | First observation: 4.17 ± 3.16 (0–16) mm |
Observer 2: 0.995 (0.993–0.996) | Observer 2: 4.07 ± 3.28 (0–16) mm | Second observation: 0.996 (0.994–0.996) | Second observation: 3.75 ± 2.97 (0–13) mm |
Fig. 3The measured length of the validation distance differed regarding the position of the magnification device. Automatic and manual calibration with the 25-mm magnification ball and manual calibration with the 30 cm magnification distance were presented seperately. Manual calibration with the 25-mm magnification ball showed the largest variation between measurements (dot: observer 1, x: observer 2, blue colour indicates first and red colour indicates second observation)
Measured object length at different positions of the magnification device
| Positions | Mean ± SD (min.–max.) | Mean differences between positions |
|---|---|---|
| (A) 5 cm anterior | 886.3 ± 2.2 (883–889) | Center vs. anterior: − 14.6 mm |
| (Z) Center | 900.8 ± 2.5 (894–903) | Center vs. posterior: 16.2 mm |
| (B) 5 cm posterior | 917.0 ± 1.7 (915–920) | Anterior vs. posterior: 30.8 mm |
| (C) 45 cm proximal | 904.0 ± 3.6 (897–907) | Center vs. proximal: 3.2 mm |
| (Z) Center | 900.8 ± 2.5 (894–903) | Center vs. distal: − 1.7 mm |
| (D) 45 cm distal | 899.2 ± 3.2 (895–904) | Proximal vs. distal: − 4.8 mm |
| (E) 5 cm medial | 901.9 ± 3.8 (897–907) | Center vs. medial: 1.1 mm |
| (Z) Center | 900.8 ± 2.5 (894–903) | Center vs. lateral: − 2.4 mm |
| (F) 15 cm lateral | 898.4 ± 4.1 (892–903) | Medial vs. lateral: − 3.5 mm |
All measurements regardless of observer, observation or calibration method were combined and compared to the other postions
Intra- and interobserver reliability (ICC (95%CI)) and intra- and interobserver difference [mean ± standard deviation (minimum–maximum)] of limb length measurements in the test setup with automatic and manual calibration with the 25 mm magnification ball and manual calibration with the 30 cm magnification distance
| Intraobserver reliability | Intraobserver differences—absolute values | Interobserver reliability | Interobserver differences—absolute values | |
|---|---|---|---|---|
| Test setup | ||||
| Automatic calibration with the magnification ball | Observer 1: 0.999 (0.995–1.000) | Observer 1: 0.1 ± 0.4 (0–1) mm | First observation: 0.998 (0.988–1.000) | First observation: 0.3 ± 0.5 (0–1) mm |
Observer 2: 0.997 (0.983–0.999) | Observer 2: 0.4 ± 0.5 (0–1) mm | Second observation: 0.999 (0.991–1.000) | Second observation: 0.3 ± 0.5 (0–1) mm | |
| Manual calibration with the magnification ball | Observer 1: 0.884 (0.472–0.980) | Observer 1: 3.7 ± 2.6 (0–7) mm | First observation: 0.954 (0.759–0.992) | First observation: 1.9 ± 2.1 (0–5) mm |
Observer 2: 0.952 (0.782–0.992) | Observer 2: 2.3 ± 2.0 (0–5) mm | Second observation: 0.921 (0.616–0.986) | Second observation: 2.4 ± 2.9 (0–7) mm | |
| Manual calibration with the 30 cm magnification distance | Observer 1 0.999 (0.995–1.000) | Observer 1 0.1 ± 0.4 (0–1) mm | First observation: 0.999 (0.995–1.000); | First observation: 0.1 ± 0.4 (0–1) mm |
Observer 2: 0.998 (0.988–1.000) | Observer 2: 0.3 ± 0.5 (0–1) mm | Second observation: 0.998 (0.988–1.000); | Second observation: 0.3 ± 0.5 (0–1) mm | |
Fig. 4a Overview of all distances in the test setup. b Focused on the relevant lines for calculating the depicted magnification device on the radiograph (a1). c Mathematical explanation, why correct magnification device positioning is important to avoid magnification errors. The final formula consists of the distance from the X-ray tube to the correct position/same level of the limb (l1) and the distance from X-ray tube to position of the magnification device (l1–l4). The size of the magnification device (l2), the level of the device in the frontal plane (h1, l3) and the distance from leg to radiographic film (l5) can be neglected and do not influence magnification errors mathematically