| Literature DB >> 21637125 |
Metin Kucukkaya1, Ozgur Karakoyun, Raffi Armagan, Unal Kuzgun.
Abstract
The Taylor Spatial Frame uses a computer program-based six-axis deformity analysis. However, there is often a residual deformity after the initial correction, especially in deformities with a rotational component. This problem can be resolved by recalculating the parameters and inputting all new deformity and mounting parameters. However, this may necessitate repeated x-rays and delay treatment. We believe that error in the mounting parameters is the main reason for most residual deformities. To prevent these problems, we describe a new calculation technique for determining the mounting parameters that uses computed tomography. This technique is especially advantageous for deformities with a rotational component. Using this technique, exact calculation of the mounting parameters is possible and the residual deformity and number of repeated x-rays can be minimized. This new technique is an alternative method to accurately calculating the mounting parameters.Entities:
Mesh:
Year: 2011 PMID: 21637125 DOI: 10.1097/BOT.0b013e3181ee40c5
Source DB: PubMed Journal: J Orthop Trauma ISSN: 0890-5339 Impact factor: 2.512