| Literature DB >> 33623506 |
Elham Kashian1, Hadi Taleshi Ahangari2, Vahab Dehlaghi1, Karim Khoshgard3, Pardis Ghafarian4,5, Raheb Ghorbani6.
Abstract
The aim of this study is to simulate GE Discovery 690 VCT positron emission tomography/computed tomography (PET/CT) scanner using Geant4 Application for Tomographic Emission (GATE) simulation package (version 8). Then, we assess the performance of scanner by comparing measured and simulated parameter results. Detection system and geometry of PET scanner that consists of 13,824 LYSO crystals designed in 256 blocks and 24 ring detectors were modeled. In order to achieve a precise model, we verified scanner model. Validation was based on a comparison between simulation data and experimental results obtained with this scanner in the same situation. Parameters used for validation were sensitivity, spatial resolution, and contrast. Image quality assessment was done based on comparing the contrast recovery coefficient (CRC) of simulated and measured images. The findings demonstrate that the mean difference between simulated and measured sensitivity is <7%. The simulated spatial resolution agreed to within <5.5% of the measured values. Contrast results had a slight divergence within the range below 4%. The image quality validation study demonstrated an acceptable agreement in CRC for 8:1 and 2:1 source-to-background activity ratio. Validated performance parameters showed good agreement between experimental data and simulated results and demonstrated that GATE is a valid simulation tool for simulating this scanner model. The simulated model of this scanner can be used for future studies regarding optimization of image reconstruction algorithms and emission acquisition protocols. Copyright:Entities:
Keywords: Digital phantom; Geant4 Application for Tomographic Emission; Monte Carlo simulation; positron emission tomography scanner; validation
Year: 2020 PMID: 33623506 PMCID: PMC7875045 DOI: 10.4103/wjnm.WJNM_4_20
Source DB: PubMed Journal: World J Nucl Med ISSN: 1450-1147
Figure 1Schematic representation of the Discovery 690 positron emission tomography scanner model with its ring shields simulated using Geant4 Application for Tomographic Emission (a), including oblique view of a block (b) and a crystal (c)
Figure 2Flowchart display of digitizer module of Discovery 690 positron emission tomography scanner
Figure 3Illustration of simulated line sources placed at the center of field of view and 1, 10, and 20 cm tangentially from the center of field of view showed in the A Medical Image Data Examiner
Figure 4Graphical illustration of cylindrical validation phantom. Top (a) and oblique (b) views
Measured FWHM for sources in different positions tangentially from the center: Simulated and experimental results
| Line source position tangentially from the center (cm) | Measured values (mm) | Simulated values (mm) |
|---|---|---|
| 20 | 1.30 | 1.71 |
| 10 | 1.20 | 1.53 |
| 1 | 1.15 | 1.46 |
| 0 (center) | 1.15 | 1.45 |
FWHM: Full width at half maximum
Figure 5The full width at half maximum as a function of tangential placement of sources. Results for central slice of simulated and measured images
Figure 6Sensitivity as a function axial slice number at the center of field of view. Results for simulated and measured images
Tangential simulated and measured FWHM for sources in different positions in central and 1/4 axial field of view slice
| Source position measured axially from the center | Measured values (mm) | Simulated values (mm) | Difference (%) |
|---|---|---|---|
| Center of FOV | |||
| Tangential (right) | 6.050 | 5.16 | 14.7 |
| Tangential (left) | 4.465 | 4.08 | 8.6 |
| Radial (down) | 3.670 | 3.60 | 1.9 |
| Radial (up) | 3.020 | 2.85 | 5.6 |
| 1/4 of FOV | |||
| Tangential (right) | 6.040 | 5.655 | 6.4 |
| Tangential (left) | 4.400 | 4.005 | 9 |
| Radial (down) | 3.695 | 3.25 | 12 |
| Radial (up) | 3.18 | 3.1 | 2.5 |
FOV: Field of view; FWHM: Full width at half maximum
Radial simulated and measured FWHM for sources in different position in central and 1/4 axial field of view slice
| Source position measured axially from the center | Measured values (mm) | Simulated values (mm) | Difference (%) |
|---|---|---|---|
| Center of FOV | |||
| Tangential (right) | 6.060 | 6.005 | 0.9 |
| Tangential (left) | 4.515 | 4.38 | 3 |
| Radial (down) | 3.370 | 3.285 | 2.5 |
| Radial (up) | 2.930 | 2.53 | 13.6 |
| 1/4 of FOV | |||
| Tangential (right) | 6.1 | 5.975 | 2.05 |
| Tangential (left) | 4.52 | 4.31 | 4.65 |
| Radial (down) | 3.35 | 3.29 | 1.8 |
| Radial (up) | 2.885 | 2.525 | 12.5 |
FOV: Field of view; FWHM: Full width at half maximum
Figure 7Simulated and measured tangential FWHM as a function of source diameter in central (a) and ¼ axial field of views (b)
Simulated and measured sensitivity for sources in different positions in central and 1/4 axial field of view slice
| Source position measured axially from the center | Measured values (×10−3 cps/kBq) | Simulated values (×10−3 cps/kBq) | Difference (%) |
|---|---|---|---|
| Center of FOV | |||
| Tangential (right) | 1.49 | 1.39 | 7.11 |
| Tangential (left) | 1.46 | 1.37 | 6.33 |
| Radial (down) | 1.45 | 1.35 | 7.01 |
| Radial (up) | 1.23 | 1.12 | 9.21 |
| 1/4 of FOV | |||
| Tangential (right) | 1.43 | 1.33 | 6.77 |
| Tangential (left) | 1.47 | 1.35 | 8.5 |
| Radial (down) | 1.40 | 1.37 | 1.77 |
| Radial (up) | 1.21 | 1.34 | 9.99 |
FOV: Field of view
Comparison between simulated and experimental values obtained for the contrast of four sources with different size in two different source-to-background ratios
| Source position measured axially from the center | Source diameter (mm) | Measured values (%) | Simulated values (%) | Difference (%) |
|---|---|---|---|---|
| Source-to-background ratio 2:1 | ||||
| Tangential (right) | 21 | 1.73 | 1.72 | 0.42 |
| Tangential (left) | 16 | 1.67 | 1.63 | 2.2 |
| Radial (down) | 11 | 1.30 | 1.32 | 2.16 |
| Radial (up) | 8 | 1.17 | 1.16 | 1.44 |
| Source-to-background ratio 8:1 | ||||
| Tangential (right) | 21 | 4.8 | 4.88 | 1.98 |
| Tangential (left) | 16 | 4.15 | 4.53 | 5.7 |
| Radial (down) | 11 | 3.53 | 3.35 | 5.7 |
| Radial (up) | 8 | 2.43 | 2.23 | 10.85 |
Figure 8Contrast of four sources with 2:1 (a) 8:1 (b) source-to-background ratios. The plots refer to simulated and measured images
Comparison between simulated and experimental values obtained for the contrast recovery coefficient of four sources with different size in two different source-to-background ratios
| Source position measured axially from the center | Source diameter (mm) | Measured values (%) | Simulated values (%) | Difference (%) |
|---|---|---|---|---|
| Source-to-background ratio 2:1 | ||||
| Tangential (right) | 21 | 10.39 | 10.29 | 0.96 |
| Tangential (left) | 16 | 9.56 | 9.04 | 5.44 |
| Radial (down) | 11 | 4.22 | 4.62 | 9.48 |
| Radial (up) | 8 | 2.49 | 2.25 | 9.64 |
| Source-to-background ratio 8:1 | ||||
| Tangential (right) | 21 | 54.10 | 55.46 | 2.51 |
| Tangential (left) | 16 | 46.91 | 50.40 | 7.44 |
| Radial (down) | 11 | 36.39 | 33.50 | 7.94 |
| Radial (up) | 8 | 21.50 | 17.62 | 18.05 |
Figure 9Contrast recovery coefficient calculated over region of interests in central slice encompassing the four sources in cylindrical phantom as a function of source diameter. The plots refer to images with source-to-background ratio 2:1 (a) and 8:1 (b)