Accurate Monte Carlo modeling and performance assessment of the X-PET subsystem of the FLEX triumph preclinical PET/CT scanner.

N Zeraatkar1, M R Ay, A R Kamali-Asl, H Zaidi.   

Abstract

PURPOSE: X-PET is a commercial small animal PET scanner incorporating several innovative designs to achieve improved performance. It is employed as a PET subsystem in the FLEX Triumph preclinical PET/CT scanner, the first commercial small animal PET/CT scanner worldwide. The authors report on a novel Monte Carlo (MC) model designed for the evaluation of performance parameters of the X-PET
METHODS: The Geant4 Application for Tomographic Emission (GATE) MC code was used as a simulation tool. The authors implemented more accurate modeling of the geometry of detector blocks and associated electronic chains, including dead-time and time-independent parameters, compared to previously presented MC models of the X-PET scanner. Validation of the MC model involved comparison between simulated and measured performance parameters of the X-PET, including spatial resolution, sensitivity, and noise equivalent count rate (NECR). Thereafter, various simulations were performed to assess scanner performance parameters according to NEMA NU 4-2008 standards with the aim to present a reliable Monte Carlo platform for small animal PET scanner design optimization.
RESULTS: The average differences between simulated and measured results were 11.2%, 33.3%, and 9.1% for spatial resolution, sensitivity, and NECR, respectively. The average system absolute sensitivity was 2.7%. Furthermore, the peak true count rate, peak NECR, and scatter fraction were 2050 kcps, 1520 kcps, and 4.7%, respectively, for a mouse phantom and 1017 kcps, 469 kcps, and 18.2%, respectively, for a rat phantom. Spatial resolution was also measured in ten different positions at two axial locations. The radial, tangential, and axial FWHM ranged from 1.31 to 1.96 mm, 1.17 to 2.11 mm, and 1.77 to 2.44 mm, respectively, as the radial position varied from 0 to 25 mm at the centre of the axial field-of-view.
CONCLUSIONS: The developed MC simulation platform provides a reliable tool for performance evaluation of small animal PET scanners and has the potential to be used in other applications such as detector design optimization, correction of image degrading factors such as randoms, scatter, intercrystal scatter, parallax error, and partial volume effect.

Entities:  

Mesh:

Year:  2011        PMID: 21520834     DOI: 10.1118/1.3547721

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  5 in total

1.  A cone-shaped phantom for assessment of small animal PET scatter fraction and count rate performance.

Authors:  Rameshwar Prasad; Habib Zaidi
Journal:  Mol Imaging Biol       Date:  2012-10       Impact factor: 3.488

2.  Respiratory-induced errors in tumor quantification and delineation in CT attenuation-corrected PET images: effects of tumor size, tumor location, and respiratory trace: a simulation study using the 4D XCAT phantom.

Authors:  Parham Geramifar; Mojtaba Shamsaie Zafarghandi; Pardis Ghafarian; Arman Rahmim; Mohammad Reza Ay
Journal:  Mol Imaging Biol       Date:  2013-12       Impact factor: 3.488

3.  Preclinical positron emission tomography scanner based on a monolithic annulus of scintillator: initial design study.

Authors:  Alexander V Stolin; Peter F Martone; Gangadhar Jaliparthi; Raymond R Raylman
Journal:  J Med Imaging (Bellingham)       Date:  2017-01-05

4.  Combined micro-PET/micro-CT imaging of lung tumours in SPC-raf and SPC-myc transgenic mice.

Authors:  Thomas Rodt; Matthias Luepke; Claudia Boehm; Katja Hueper; Roman Halter; Silke Glage; Ludwig Hoy; Frank Wacker; Juergen Borlak; Christian von Falck
Journal:  PLoS One       Date:  2012-09-21       Impact factor: 3.240

5.  Monte Carlo simulation and performance assessment of GE Discovery 690 VCT positron emission tomography/computed tomography scanner.

Authors:  Elham Kashian; Hadi Taleshi Ahangari; Vahab Dehlaghi; Karim Khoshgard; Pardis Ghafarian; Raheb Ghorbani
Journal:  World J Nucl Med       Date:  2020-07-22
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.