A phantom study comparing radial trajectories for accelerated cardiac 4D flow MRI against a particle imaging velocimetry reference.

Philip A Corrado1, Rafael Medero2, Kevin M Johnson3, Christopher J François4, Alejandro Roldán-Alzate5, Oliver Wieben3.   

Abstract

PURPOSE: Radial sampling is one method to accelerate 4D flow MRI acquisition, making feasible dual-velocity encoding (Venc) assessment of slow flow in the left ventricle (LV). Here, two radial trajectories are compared in vitro for this application: 3D radial (phase-contrast vastly undersampled isotropic projection, PC-VIPR) versus stack of stars (phase-contrast stack of stars, PC-SOS), with benchtop particle imaging velocimetry (PIV) serving as a reference standard.
METHODS: The study contained three steps: (1) Construction of an MRI- and PIV-compatible LV model from a healthy adult's CT images. (2) In vitro PIV using a pulsatile flow pump. (3) In vitro dual-Venc 4D flow MRI using PC-VIPR and PC-SOS (two repeat experiments). Each MR image set was retrospectively undersampled to five effective scan durations and compared with the PIV reference. The root-mean-square velocity vector difference (RMSE) between MRI and PIV images was compared, along with kinetic energy (KE) and wall shear stress (WSS).
RESULTS: RMSE increased as scan time decreased for both MR acquisitions. RMSE was 3% lower in PC-SOS images than PC-VIPR images in 30-min scans (3.8 vs. 3.9 cm/s) but 98% higher in 2.5-min scans (9.5 vs. 4.8 cm/s). PIV intrasession repeatability showed a RMSE of 4.4 cm/s, reflecting beat-to-beat flow variation, while MRI had intersession RMSEs of 3.8/3.5 cm/s for VIPR/SOS, respectively. Speed, KE, and WSS were overestimated voxel-wise in 30-min MRI scans relative to PIV by 0.4/0.3 cm/s, 0.2/0.1 μJ/mL, and 36/43 mPa, respectively, for VIPR/SOS.
CONCLUSIONS: PIV is feasible for application-specific 4D flow MRI protocol optimization. PC-VIPR is better-suited to dual-Venc LV imaging with short scan times.
© 2021 International Society for Magnetic Resonance in Medicine.

Entities:  

Keywords:  dual-Venc; flow; non-cartesian; phantom; radial; validation

Mesh:

Year:  2021        PMID: 33547658      PMCID: PMC8109233          DOI: 10.1002/mrm.28698

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   3.737


  35 in total

1.  Phase-contrast with interleaved undersampled projections.

Authors:  A V Barger; D C Peters; W F Block; K K Vigen; F R Korosec; T M Grist; C A Mistretta
Journal:  Magn Reson Med       Date:  2000-04       Impact factor: 4.668

2.  Time-resolved contrast-enhanced imaging with isotropic resolution and broad coverage using an undersampled 3D projection trajectory.

Authors:  Andrew V Barger; Walter F Block; Yuriy Toropov; Thomas M Grist; Charles A Mistretta
Journal:  Magn Reson Med       Date:  2002-08       Impact factor: 4.668

3.  Quantification of left and right ventricular kinetic energy using four-dimensional intracardiac magnetic resonance imaging flow measurements.

Authors:  M Carlsson; E Heiberg; J Toger; H Arheden
Journal:  Am J Physiol Heart Circ Physiol       Date:  2011-12-16       Impact factor: 4.733

4.  In Vitro Validation of 4D Flow MRI for Local Pulse Wave Velocity Estimation.

Authors:  Timothy Ruesink; Rafael Medero; David Rutkowski; Alejandro Roldán-Alzate
Journal:  Cardiovasc Eng Technol       Date:  2018-09-14       Impact factor: 2.495

5.  Independent validation of four-dimensional flow MR velocities and vortex ring volume using particle imaging velocimetry and planar laser-Induced fluorescence.

Authors:  Johannes Töger; Sebastian Bidhult; Johan Revstedt; Marcus Carlsson; Håkan Arheden; Einar Heiberg
Journal:  Magn Reson Med       Date:  2015-05-02       Impact factor: 4.668

6.  Image-based background phase error correction in 4D flow MRI revisited.

Authors:  Julia Busch; Daniel Giese; Sebastian Kozerke
Journal:  J Magn Reson Imaging       Date:  2017-02-22       Impact factor: 4.813

7.  Comparison of 4D Flow MRI and Particle Image Velocimetry Using an In Vitro Carotid Bifurcation Model.

Authors:  Rafael Medero; Carson Hoffman; Alejandro Roldán-Alzate
Journal:  Ann Biomed Eng       Date:  2018-08-15       Impact factor: 3.934

8.  Assessment of human left ventricle flow using statistical shape modelling and computational fluid dynamics.

Authors:  S S Khalafvand; J D Voorneveld; A Muralidharan; F J H Gijsen; J G Bosch; T van Walsum; A Haak; N de Jong; S Kenjeres
Journal:  J Biomech       Date:  2018-04-26       Impact factor: 2.712

9.  Quantification and visualization of cardiovascular 4D velocity mapping accelerated with parallel imaging or k-t BLAST: head to head comparison and validation at 1.5 T and 3 T.

Authors:  Marcus Carlsson; Johannes Töger; Mikael Kanski; Karin Markenroth Bloch; Freddy Ståhlberg; Einar Heiberg; Håkan Arheden
Journal:  J Cardiovasc Magn Reson       Date:  2011-10-04       Impact factor: 5.364

Review 10.  Left ventricular thrombus formation after acute myocardial infarction.

Authors:  Ronak Delewi; Felix Zijlstra; Jan J Piek
Journal:  Heart       Date:  2012-12       Impact factor: 5.994

View more
  1 in total

1.  Sildenafil administration improves right ventricular function on 4D flow MRI in young adults born premature.

Authors:  Philip A Corrado; Gregory P Barton; Christopher J Francois; Oliver Wieben; Kara N Goss
Journal:  Am J Physiol Heart Circ Physiol       Date:  2021-04-16       Impact factor: 5.125

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.