PURPOSE: To evaluate the reliability of attenuation values in virtual non-contrast images (VNC) reconstructed from contrast-enhanced, dual-energy scans performed on a second-generation dual-energy CT scanner, compared to single-energy, non-contrast images (TNC).
MATERIALS AND METHODS: Sixteen phantoms containing a mixture of contrast agent and water at different attenuations (0-1400 HU) were investigated on a Definition Flash-CT scanner using a single-energy scan at 120 kV and a DE-CT protocol (100 kV/SN140 kV). For clinical assessment, 86 patients who received a dual-phase CT, containing an unenhanced single-energy scan at 120 kV and a contrast enhanced (110 ml Iomeron 400 mg/ml; 4 ml/s) DE-CT (100 kV/SN140 kV) in an arterial (n=43) or a venous phase, were retrospectively analyzed. Mean attenuation was measured within regions of interest of the phantoms and in different tissue types of the patients within the corresponding VNC and TNC images. Paired t-tests and Pearson correlation were used for statistical analysis.
RESULTS: For all phantoms, mean attenuation in VNC was 5.3±18.4 HU, with respect to water. In 86 patients overall, 2637 regions were measured in TNC and VNC images, with a mean difference between TNC and VNC of -3.6±8.3 HU. In 91.5% (n=2412) of all cases, absolute differences between TNC and VNC were under 15HU, and, in 75.3% (n=1986), differences were under 10 HU.
CONCLUSIONS: Second-generation dual-energy CT based VNC images provide attenuation values close to those of TNC. To avoid possible outliers multiple measurements are recommended especially for measurements in the spleen, the mesenteric fat, and the aorta.
Copyright © 2012 Elsevier Ireland Ltd. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22236702     DOI: 10.1016/j.ejrad.2011.12.011

Source DB:  PubMed          Journal:  Eur J Radiol        ISSN: 0720-048X            Impact factor:   3.528


  36 in total

1.  Virtual versus true non-contrast dual-energy CT imaging for the diagnosis of aortic intramural hematoma.

Authors:  Salim Si-Mohamed; Nicolas Dupuis; Valérie Tatard-Leitman; David Rotzinger; Sara Boccalini; Matthias Dion; Alain Vlassenbroek; Philippe Coulon; Yoad Yagil; Nadav Shapira; Philippe Douek; Loic Boussel
Journal:  Eur Radiol       Date:  2019-07-01       Impact factor: 5.315

2.  Dual-energy CT angiography-derived virtual non-contrast images for follow-up of patients with surgically clipped aneurysms: a retrospective study.

Authors:  Su Young Yun; Young Jin Heo; Hae Woong Jeong; Jin Wook Baek; Hye Jung Choo; Gi Won Shin; Sung Tae Kim; Young Gyun Jeong; Ji Young Lee; Hyun Seok Jung
Journal:  Neuroradiology       Date:  2019-01-25       Impact factor: 2.804

3.  Photon-counting CT for simultaneous imaging of multiple contrast agents in the abdomen: An in vivo study.

Authors:  Rolf Symons; Bernhard Krauss; Pooyan Sahbaee; Tyler E Cork; Manu N Lakshmanan; David A Bluemke; Amir Pourmorteza
Journal:  Med Phys       Date:  2017-08-20       Impact factor: 4.071

4.  Can virtual non-enhanced CT be used to replace true non-enhanced CT for the detection of palpable cervical lymph nodes? A preliminary study.

Authors:  Yaying Yang; Xiaoxia Jia; Yamin Deng; Jiuhong Chen; Long Jiang Zhang
Journal:  Jpn J Radiol       Date:  2014-04-19       Impact factor: 2.374

5.  Longitudinal reproducibility of attenuation measurements on virtual unenhanced images: multivendor dual-energy CT evaluation.

Authors:  Simon Lennartz; Anushri Parakh; Jinjin Cao; Avinash Kambadakone
Journal:  Eur Radiol       Date:  2021-06-10       Impact factor: 5.315

6.  Accuracy of spectral curves at different phantom sizes and iodine concentrations using dual-source dual-energy computed tomography.

Authors:  Kazuhiro Sato; Ryota Kageyama; Yuta Sawatani; Hirokazu Takano; Shingo Kayano; Yumi Takane; Haruo Saito
Journal:  Phys Eng Sci Med       Date:  2021-02-02

7.  Compatibility of true and virtual unenhanced attenuation in rapid kV-switching dual energy CT.

Authors:  İlkay Çamlıdağ
Journal:  Diagn Interv Radiol       Date:  2020-03       Impact factor: 2.630

8.  Dual-energy CT of the brain: Comparison between DECT angiography-derived virtual unenhanced images and true unenhanced images in the detection of intracranial haemorrhage.

Authors:  Matteo Bonatti; Fabio Lombardo; Giulia A Zamboni; Patrizia Pernter; Roberto Pozzi Mucelli; Giampietro Bonatti
Journal:  Eur Radiol       Date:  2016-11-23       Impact factor: 5.315

9.  Can single-phase dual-energy CT reliably identify adrenal adenomas?

Authors:  A Helck; N Hummel; F G Meinel; T Johnson; K Nikolaou; A Graser
Journal:  Eur Radiol       Date:  2014-05-08       Impact factor: 5.315

10.  CT of suspected thoracic acute aortic injury in the emergency department: is routine abdominopelvic imaging worth the additional collective radiation dose?

Authors:  Shawn Haji-Momenian; Jonathan Rischall; Neil Okey; Myles Taffel; Nadia Khati; Robert Zeman
Journal:  Emerg Radiol       Date:  2016-08-27