Validation of a Portable Blood Gas Analyzer for Use in Challenging Field Conditions at High Altitude.

Janek Nawrocki1, Michael Furian1, Aline Buergin1, Laura Mayer1, Simon Schneider1, Maamed Mademilov2, Madeleine S Bloch1, Talant M Sooronbaev2, Silvia Ulrich1, Konrad E Bloch1.   

Abstract

BACKGROUND: Novel, portable blood gas analyzers (BGAs) may serve as essential point-of-care tools in remote regions, during air travel or in ambulance services but they have not been extensively validated. RESEARCH QUESTION: We compared accuracy of a portable BGA to a validated stationary device.
METHODS: In healthy individuals and patients with chronic obstructive pulmonary disease participating in clinical field studies at different altitudes, arterial blood samples were obtained at rest and during exercise in a hospital at 760 m and in a high altitude clinic at 3100 m. Paired measurements by a portable BGA (EPOC, Siemens Healthcare) and a stationary BGA (Rapidpoint500, Siemens Healthcare) were performed to compute bias (mean difference) and limits of agreement (95% CI of bias).
RESULTS: Of 105 individuals, 248 arterial blood samples were analyzed, 108 at 760 m, 140 at 3100 m. Ranges of values measured by portable BGA were: pH 7.241-7.473, PaCO2 21.5-52.5 mmHg, and PaO2 45.5-107.1 mmHg. Bias (95% CI) between devices were: pH 0.007 (-0.029 to 0.044), PaCO2 -0.3 mmHg (-4.8 to 4.2), and PaO2 -0.2 mmHg (-9.1 to 4.7). For pH, agreement between devices was improved by the equation to correct pH by portable BGA = -1.37 + pH measured × 1.19; bias after correction -0.007 (-0.023 to 0.009). The portable BGA was easily handled and worked reliably.
INTERPRETATION: Accuracy of blood gas analysis by the portable BGA in comparison to the reference BGA was adequate for clinical use. Because of portability and ease of handling, portable BGA are valuable diagnostic tools for use in everyday practice as well as under challenging field conditions.
Copyright © 2021 Nawrocki, Furian, Buergin, Mayer, Schneider, Mademilov, Bloch, Sooronbaev, Ulrich and Bloch.

Entities:  

Keywords:  arterial blood gas (ABG) analyzers; blood gas analyzer; high altitude; hypercapnia; hypoxaemia respiratory failure; point-of-care testing

Year:  2021        PMID: 33488397      PMCID: PMC7820720          DOI: 10.3389/fphys.2020.600551

Source DB:  PubMed          Journal:  Front Physiol        ISSN: 1664-042X            Impact factor:   4.566


  12 in total

1.  Analytical evaluation of the epoc® point-of-care blood analysis system in cardiopulmonary bypass patients.

Authors:  Jianing Chen; Monique Gorman; Bill O'Reilly; Yu Chen
Journal:  Clin Biochem       Date:  2016-02-20       Impact factor: 3.281

2.  CLIA requirements for proficiency testing: the basics for laboratory professionals.

Authors:  John R Astles; Heather Stang; Todd Alspach; Graylin Mitchell; Maribeth Gagnon; Diane Bosse
Journal:  MLO Med Lab Obs       Date:  2013-09

3.  Analytical and clinical performance of the epoc blood analysis system: experience at a large tertiary academic medical center.

Authors:  Brie A Stotler; Alexander Kratz
Journal:  Am J Clin Pathol       Date:  2013-11       Impact factor: 2.493

4.  Interchangeability of blood gas, electrolyte and metabolite results measured with point-of-care, blood gas and core laboratory analyzers.

Authors:  Aila Leino; Kaisa Kurvinen
Journal:  Clin Chem Lab Med       Date:  2011-04-20       Impact factor: 3.694

5.  Emergency medical technician-performed point-of-care blood analysis using the capillary blood obtained from skin puncture.

Authors:  Changsun Kim; Hansol Kim
Journal:  Am J Emerg Med       Date:  2017-12-09       Impact factor: 2.469

6.  Comparison of Enterprise Point-of-Care and Nova Biomedical Critical Care Xpress analyzers for determination of arterial pH, blood gas, and electrolyte values in canine and equine blood.

Authors:  Taher N Elmeshreghi; Tamara L Grubb; Stephen A Greene; Claude A Ragle; Jane A Wardrop
Journal:  Vet Clin Pathol       Date:  2018-07-09       Impact factor: 1.180

7.  Evaluation of a hand-held blood gas analyzer for rapid determination of blood gases, electrolytes and metabolites in intensive care setting.

Authors:  Antti A M Luukkonen; Tiina M Lehto; Pirjo S M Hedberg; Tommy E Vaskivuo
Journal:  Clin Chem Lab Med       Date:  2016-04       Impact factor: 3.694

Review 8.  Understanding Bland Altman analysis.

Authors:  Davide Giavarina
Journal:  Biochem Med (Zagreb)       Date:  2015-06-05       Impact factor: 2.313

9.  Comparison of two portable clinical analyzers to one stationary analyzer for the determination of blood gas partial pressures and blood electrolyte concentrations in horses.

Authors:  Katharina Kirsch; Johann Detilleux; Didier Serteyn; Charlotte Sandersen
Journal:  PLoS One       Date:  2019-02-15       Impact factor: 3.240

10.  Linearity analysis and comparison study on the epoc(®) point-of-care blood analysis system in cardiopulmonary bypass patients.

Authors:  Jianing Chen; Monique Gorman; Bill O'Reilly; Yu Chen
Journal:  Data Brief       Date:  2016-01-30
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.