Katharina Kirsch1, Johann Detilleux2, Didier Serteyn1, Charlotte Sandersen1.
Abstract
Portable blood gas analyzers are used to facilitate diagnosis and treatment of disorders related to disturbances of acid-base and electrolyte balance in the ambulatory care of equine patients. The aim of this study was to determine whether 2 portable analyzers produce results in agreement with a stationary analyzer. Blood samples from 23 horses hospitalized for various medical reasons were included in this prospective study. Blood gas analysis and electrolyte concentrations measured by the portable analyzers VetStat and epoc were compared to those produced by the cobas b 123 analyzer via concordance analysis, Passing-Bablok regression and Bland-Altman analysis. Limits of agreement indicated relevant bias between the VetStat and cobas b 123 for partial pressure of oxygen (pO2; 27.5-33.8 mmHg), sodium ([Na+]; 4.3-21.6 mmol/L) and chloride concentration ([Cl-]; 0.3-7.9 mmol/L) and between the epoc and cobas b 123 for pH (0.070-0.022), partial pressure of carbon dioxide (pCO2; 3.6-7.3 mmHg), pO2 (36.2-32.7 mmHg) and [Na+] (0.38.1 mmol/L). The VetStat analyzer yielded results that were in agreement with the cobas b 123 analyzer for determination of pH, pCO2, bicarbonate ([HCO3-]) and potassium concentration [K+], while the epoc analyzer achieved acceptable agreement for [HCO3-] and [K+]. The VetStat analyzer may be useful in performing blood gas analysis in equine samples but analysis of [Na+], [Cl-] and pO2 should be interpreted with caution. The epoc delivered reliable results for [HCO3-] and [K+], while results for pH, pCO2, pO2 and [Na+] should be interpreted with caution.Entities:
Mesh:
Substances:
Year: 2019 PMID: 30768603 PMCID: PMC6377089 DOI: 10.1371/journal.pone.0211104
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Concordance correlation analysis between the portable analyzers VetStat (n = 39; 30 arterial, 9 venous) and epoc (n = 33; 24 arterial, 9 venous) and the cobas b 123 analyzer.
Lin’s concordance correlation coefficients (rccc) with 95% confidence intervals (95% CI), precision (Pearson ρ) and accuracy (bias correction factor, Cb) are reported.
| Measured variables | n | Concordance correlation analysis | |||
|---|---|---|---|---|---|
| Lin’s rccc (95% CI) | Precision ρ | Accuracy Cb | |||
| pH | cobas/VetStat | 39 | 0.954 (0.920 to 0.974) | 0.973 | 0.981 |
| cobas/epoc | 33 | 0.870 (0.782 to 0.924) | 0.962 | 0.920 | |
| pCO2 (mmHg) | cobas/VetStat | 39 | 0.968 (0.939 to 0.983) | 0.954 | 0.912 |
| cobas/epoc | 33 | 0.934 (0.878 to 0.965) | 0.959 | 0.974 | |
| pO2 (mmHg) | cobas/VetStat | 39 | 0.991 (0.984 to 0.995) | 0.992 | 0.999 |
| cobas/epoc | 33 | 0.985 (0.972 to 0.993) | 0.987 | 0.998 | |
| HCO3-(mmol/L) | cobas/VetStat | 39 | 0.846 (0.736 to 0.913) | 0.892 | 0.949 |
| cobas/epoc | 33 | 0.862 (0.743 to 0.928) | 0.867 | 0.994 | |
| Na+ (mmol/L) | cobas/VetStat | 39 | 0.083 (0.022 to 0.143) | 0.471 | 0.175 |
| cobas/epoc | 33 | 0.738 (0.602 to 0.833) | 0.941 | 0.785 | |
| K+ (mmol/L) | cobas/VetStat | 39 | 0.965 (0.935 to 0.981) | 0.974 | 0.990 |
| cobas/epoc | 30 | 0.951 (0.905 to 0.976) | 0.970 | 0.981 | |
| Cl- (mmol/L) | cobas/VetStat | 39 | 0.363 (0.220 to 0.491) | 0.772 | 0.470 |
Intercept, slope and residual standard deviation of Passing-Bablok regression as well as bias, lower and upper limit of agreement from the Bland-Altman plots and their 95% confidence intervals (95% CI) from the comparisons between the cobas b 123 and VetStat and the cobas b 123 and epoc analyzers.
| Measured variables | Passing-Bablok regression | Bland-Altman plot | |||||
|---|---|---|---|---|---|---|---|
| Intercept | Slope | Residual standard deviation | Bias | Lower limit of agreement | Upper limit of agreement | ||
| pH | cobas/ VetStat | -0.89 | 1.12 | 0.010 | 0.009 | -0.022 | 0.040 |
| cobas/ epoc | -1.65 | 1.22 | 0.014 | -0.024 | -0.070 | 0.022 | |
| pCO2 (mmHg) | cobas/ VetStat | -0.34 | 1.02 | 1.450 | 0.5 | -3.4 | 4.5 |
| cobas/ epoc | -6.65 | 1.16 | 1.885 | 1.9 | -3.6 | 7.3 | |
| pO2 (mmHg) | cobas/ VetStat | 8.78 | 0.94 | 11.425 | 3.2 | -27.5 | 33.8 |
| cobas/epoc | 3.17 | 0.95 | 12.019 | -1.7 | -36.2 | 32.7 | |
| HCO3- (mmol/L) | cobas/ VetStat | 4.99 | 0.81 | 0.787 | -0.7 | -2.8 | 1.3 |
| cobas/ epoc | -2.28 | 1.08 | 0.918 | 0.0 | -2.5 | 2.6 | |
| Na+ (mmol/L) | cobas/ VetStat | 43.52 | 0.78 | 3.277 | 12.9 | 4.3 | 21.6 |
| cobas/ epoc | 42.07 | 0.71 | 1.921 | 4.2 | 0.3 | 8.1 | |
| K+ (mmol/L) | cobas/ VetStat | -0.41 | 0.097 | 0.07 | -0.18 | 0.33 | |
| cobas/ epoc | -0.13 | 1.14 | 0.110 | 0.12 | -0.19 | 0.42 | |
| Cl- (mmol/L) | cobas/ VetStat | 49.50 | 0.56 | 1.242 | 3.8 | -0.3 | 7.9 |
Fig 1Passing-Bablok regression analysis and Bland-Altman plot analysis for comparison of different analyzers for blood pH, pCO2, and pO2.
Fig 2Passing-Bablok regression analysis and Bland-Altman plot analysis for comparison of different analyzers for blood [HCO3-], [Na+], [K+] and [Cl-].
Passing-Bablok regression analysis for comparison of VetStat versus the cobas b 123 analyzer for blood [HCO3-] (A), [Na+] (B), [K+] (C), and [Na+] (M), and for comparison of epoc versus cobas b 123 analyzer for blood [HCO3-] (G), [Na+] (H) and [K+] (I). The dashed line is the line of identity, the solid line is the best fit. Bland-Altman plot for comparison of VetStat versus cobas b 123 analyzer for blood [HCO3-] (D), [Na+] (E), [K+] (F) and [Cl-] (N) and for comparison of epoc versus cobas b 123 analyzer for blood [HCO3-] (J), [Na+] (K), and [K+] (L). The dashed lines are upper and lower limit of agreement, the solid line is the mean difference (bias).
Mean, within-run standard deviation (SDw-run) and within-run coefficient of variation (CVw-run) in percent from one blood sample analyzed for each of the 3 analyzers (cobas b 123, VetStat and epoc) and compared to published precision targets [21].
| Measured variables | cobas b 123 | VetStat | epoc | ||||
|---|---|---|---|---|---|---|---|
| Mean ± SDw-run | CVw-run (%) | Mean ± SDw-run | CVw-run (%) | Mean ± SDw-run | CVw-run (%) | Precision target (CV%) | |
| pH | 7.386 ± 0.005 | 0.07 | 7.436 ± 0.011 | 0.15 | 7.341 ± 0.004 | 0.06 | 0.1[ |
| pCO2 (mmHg) | 42.71 ± 1.13 | 50.29 ± 0.49 | 0.97 | 52.11 ± 1.68 | 2.4[ | ||
| pO2 (mmHg) | 95.19 ± 3.88 | 4.08 | 314.39 ± 18.71 | 4.8[ | |||
| HCO3- (mmol/L) | 25.00 ± 0.37 | 1.50 | 31.19 ± 0.48 | 1.53 | 28.19 ± 0.76 | 2.69 | <5[ |
| Na+ (mmol/L) | 133.1 ± 1.0 | 0.70 | 149.4 ± 1.1 | 0.80 | 135.6 ± 1.4 | 1.00 | 0.3[ |
| K+ (mmol/L) | 4.040 ± 0.125 | 4.286 ± 0.038 | 0.88 | 4.114 ± 0.267 | 2.3[ | ||
| Cl- (mmol/L) | 100.8 ± 0.2 | 0.20 | 111.0 ± 1.4 | 0.6[ | |||
pCO2, partial pressure of carbon dioxide; pO2, partial pressure of oxygen; HCO3-, bicarbonate; Na+, sodium; K+, potassium; Cl-, chloride. Values not meeting the precision targets are in bold.