Linearity analysis and comparison study on the epoc(®) point-of-care blood analysis system in cardiopulmonary bypass patients.

Jianing Chen1, Monique Gorman2, Bill O'Reilly3, Yu Chen4.   

Abstract

The epoc(®) blood analysis system (Epocal Inc., Ottawa, Ontario, Canada) is a newly developed in vitro diagnostic hand-held analyzer for testing whole blood samples at point-of-care, which provides blood gas, electrolytes, ionized calcium, glucose, lactate, and hematocrit/calculated hemoglobin rapidly. The analytical performance of the epoc(®) system was evaluated in a tertiary hospital, see related research article "Analytical evaluation of the epoc(®) point-of-care blood analysis system in cardiopulmonary bypass patients" [1]. Data presented are the linearity analysis for 9 parameters and the comparison study in 40 cardiopulmonary bypass patients on 3 epoc(®) meters, Instrumentation Laboratory GEM4000, Abbott iSTAT, Nova CCX, and Roche Accu-Chek Inform II and Performa glucose meters.

Entities:  

Year:  2016        PMID: 26937460      PMCID: PMC4749944          DOI: 10.1016/j.dib.2016.01.040

Source DB:  PubMed          Journal:  Data Brief        ISSN: 2352-3409


Specifications table Value of the data Detailed analytical linearity analysis for 9 parameters on the epoc® meters was presented. Comparison study was conducted on 40 cardiopulmonary bypass patients. The data helps medical laboratories and point-of-care testing users to make an informed decision on blood gas analyzer selection.

Data

The data contains information on the analytical linearity performances for 9 parameters on 3 epoc® meters (Supplementary Fig. 1). It also contains information on the cardiopulmonary bypass patient sample comparison study for analytical accuracy performance for 8 parameters on 3 epoc® meters (Fig. 1, Fig. 2, Fig. 3, Fig. 4).
Fig. 1

The comparison study of the epoc® point-of-care blood analysis system with the GEM4000 in cardiopulmonary bypass patients.

Fig. 2

The comparison study of the epoc® point-of-care blood analysis system with the iSTAT in cardiopulmonary bypass patients.

Fig. 3

The comparison study of the epoc® point-of-care blood analysis system with the Nova CCX in cardiopulmonary bypass patients.

Fig. 4

The comparison study of the epoc® point-of-care blood analysis system with the Roche glucose meters in cardiopulmonary bypass patients.

Experimental design, materials and methods

The epoc® blood analysis system

The epoc® blood analysis system (Epocal Inc., Ottawa, Ontario, Canada) is a newly developed hand-held analyzer for testing whole blood samples at point-of-care, which provides blood gas, electrolytes, ionized calcium, glucose, lactate, and hematocrit/calculated hemoglobin in 30 seconds. This system contains a test card, a wireless card reader, and a host mobile computer. pH, pCO2, sodium, potassium, and ionized calcium are measure potentiometrically; pO2, glucose, and lactate are measured amperometrically, whereas hematocrit is determined conductometrically [2]. Hemoglobin is calculated from the measured hematocrit using the formula: Hemoglobin (g/L)=Hematocrit (decimal fraction)×340 [3], [4].

Linearity study

The epoc® point-of-care blood analysis system was evaluated using several Clinical and Laboratory Standards Institute (CLSI) evaluation protocols for testing the linearity (EP6) [5]. Five levels of Eurotrol epoc Calibration Verification Fluids (Eurotrol B.V., Keplerlaan, The Netherlands, lot#183-B407), and 5 levels of Eurotrol epoc Hematocrit Verification Fluids (Eurotrol B.V., Keplerlaan, The Netherlands, lot#190-B404) were measured on all three epoc® blood analysis systems. These linearity materials were analyzed in triplicate on each system respectively.

Comparison study

The epoc® point-of-care blood analysis system was evaluated using several Clinical and Laboratory Standards Institute (CLSI) evaluation protocols for testing the accuracy (EP15) [6] and bias (EP9) [7]. Remnant specimens from 40 heparinized CPB patients collected in plain 3 mL syringe (Becton Dickinson, Franklin Lakes, New Jersey) for routine clinical analysis on GEM4000 (Instrumentation Laboratory, Bedford, MA, USA) in the cardiovascular operating room of the Saint John Regional Hospital, Horizon Health Network, were used for this study. Samples collected were of arterial, mixed venous, and venous types. After being analyzed on GEM4000 and all 3 epoc meters (therefore total epoc® measurements were up to 118), samples were analyzed in Abbott iSTAT (Abbott Point of Care, Princeton, NJ, USA), Nova CCX analyzer (Nova Biomedical Corporation, Waltham, MA, USA), Accu-Chek Inform II and Performa glucose meters (Roche Diagnostics, Basel, Switzerland) side by side, with all measurements performed within 5 min. All testing devices were run according to manufacturers’ instructions by a medical laboratory technologist. These arrangements attempted to eliminate pre-analytical errors associated with blood analysis, such as different sample collection containers and sensitive specimen stability [8], [9].

Statistical method

Statistical analysis was carried out using Microsoft Excel. The best fit line by linear regression was used to evaluate assay linearity (Supplementary Fig. 1). Regression analysis was used to evaluate method comparisons (Fig. 1, Fig. 2, Fig. 3, Fig. 4). Bland–Altman analysis was constructed to assess systematic bias between methods (see Ref. [1]). Comparison studies on hemoglobin measurements see Fig. 1 in Ref. [1].
Subject areaChemistry, Biology
More specific subject areaPoint-of-care testing
Type of dataFigure
How data was acquiredThe epoc® blood analysis system (Epocal Inc., Ottawa, Ontario, Canada)
Data formatAnalyze data
Experimental factorsPatients under cardiopulmonary bypass were all heparinized as routine
Experimental featuresLinearity was evaluated using 5 levels of Eurotrol epoc Calibration Verification Fluids and 5 levels of Eurotrol epoc Hematocrit Verification Fluids (Eurotrol B.V., Keplerlaan, The Netherlands) on 3 epoc® blood analysis systems. Linearity materials were analyzed in triplicate on each system. Remnant specimens from cardiopulmonary bypass patients collected in plain 3 mL syringe for routine clinical analysis on GEM4000 in the cardiovascular operating room were used. After being analyzed on GEM4000 and all 3 epoc meters, samples were analyzed in Abbott iSTAT, Nova CCX analyzer, Roche Accu-Chek Inform II and Performa glucose meters side by side, with all measurements performed within 5 min.
Data source locationSaint John, New Brunswick, Canada
Data accessibilityData are within this article
  3 in total

1.  Analytical evaluation of the epoc® point-of-care blood analysis system in cardiopulmonary bypass patients.

Authors:  Jianing Chen; Monique Gorman; Bill O'Reilly; Yu Chen
Journal:  Clin Biochem       Date:  2016-02-20       Impact factor: 3.281

2.  Comparison of Becton Dickinson Vacutainer rapid serum tube with the serum separator tube for routine chemistry and immunoassay tests.

Authors:  Ronald Yan; Amy Lou; Gail Watts; Heather Tarr; Hilary Smith; Lois Kinney; Yu Chen
Journal:  J Clin Pathol       Date:  2014-03-31       Impact factor: 3.411

3.  Suitability of Becton Dickinson Vacutainer rapid serum tube for collecting and storing blood samples for antibiotic and anticonvulsant drug monitoring.

Authors:  Ronald Yan; David Colantonio; Pui-Yuen Wong; Yu Chen
Journal:  J Clin Pathol       Date:  2014-09       Impact factor: 3.411

  3 in total
  4 in total

Review 1.  Point-of-Care Diagnostics: Recent Developments in a Connected Age.

Authors:  Samiksha Nayak; Nicole R Blumenfeld; Tassaneewan Laksanasopin; Samuel K Sia
Journal:  Anal Chem       Date:  2016-12-13       Impact factor: 6.986

2.  Community paramedic point of care testing: validity and usability of two commercially available devices.

Authors:  Ian E Blanchard; Ryan Kozicky; Dana Dalgarno; Justin Simms; Stacy Goulder; Tyler S Williamson; Susan Biesbroek; Lenore Page; Karen Leaman; Suzanne Snozyk; Lyle Redman; Keith Spackman; Christopher J Doig; Eddy S Lang; Gerald Lazarenko
Journal:  BMC Emerg Med       Date:  2019-05-02

3.  Validation of a Portable Blood Gas Analyzer for Use in Challenging Field Conditions at High Altitude.

Authors:  Janek Nawrocki; Michael Furian; Aline Buergin; Laura Mayer; Simon Schneider; Maamed Mademilov; Madeleine S Bloch; Talant M Sooronbaev; Silvia Ulrich; Konrad E Bloch
Journal:  Front Physiol       Date:  2021-01-08       Impact factor: 4.566

4.  Performance evaluation of all analytes on the epoc® Blood Analysis System for use in hospital surgical and intensive care units.

Authors:  Zahraa Mohammed-Ali; Sousan Bagherpoor; Pauline Diker; Thuy Hoang; Ivana Vidovic; Christine Cursio; Felix Leung; Davor Brinc
Journal:  Pract Lab Med       Date:  2020-11-25
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.